UJN catheter drive

University of Jinan compact continuous-delivery catheter drive with friction wheels in a disposable surgical box and a constant-force clamp
School of Mechanical Engineering, University of Jinan (Tianxiao Cao, Rongyu Ge [corresponding], Yi Zhang, Wenlong Wang) (cao2024_cathdrive, affiliation lines) · 2024
UJN catheter drive: main picture1 / 6
Not cleared for distribution. cao2024_cathdrive Fig. 1 overall CAD diagram, p. 2. Licence: publisher copyright (IEEE). Picture grade C.
Family
Academic bedside robots (academic)
Devices
one catheter (Sec. II); tests with a catheter (Figs. 13-14)
Degrees of freedom
2 (axial feed, circumferential rotation) (Sec. II)
Push
Friction rollers
Twist
Rotating drive unit
Tested up to
Bench
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Bench
What was tested
bench: 10 axial moves of 100 mm measured with a vernier caliper and 10 rotations of 60 deg measured with a protractor on the experimental platform (Sec. V, Figs. 12-14); no phantom. Alekseeva 2025 Table 9 lists it as TRL3 (ex vivo phantom).
Key numbers
axial error -0.65 to +0.62 mm per 100 mm move; rotation error -2 to +2 deg per 60 deg (Sec. V); calculated: 19.72 mm per feed-motor turn, max feed speed 3845.39 mm/s, 0.05 rad per rotation-motor turn, max rotation speed printed as 6.42 rad/s in Eq. 11 and 6.24 rad/s in the text (Sec. IV); friction wheel radius 14.5 mm; feed motor 11,700 rpm with reducer 4.62; rotation motor 7,600 rpm with reducer 62.01 and gear ratio 2:1 (Sec. IV)

Mechanism

Push (translation)
friction-roller: two active and two driven grooved friction wheels in a disposable box, the active ones driven through three gear shafts by one motor; the box couples to the power unit by a triangular-prism shaft (Sec. II-A1, II-B, Figs. 2, 4, 7)
Twist (rotation)
gear pair rotates the whole axial-feed unit, the feed motor rides along, wires through a conductive slip ring (Sec. II-A2, Fig. 3)
Device range
N.R. (constant-force clamp adjustable by spring pre-compression, Sec. II-B, III)
Target anatomy
N.R. (general vascular intervention; coronary, brain, liver and kidney lesions named in the introduction)

Sensing and interface

Force sensing
none (a sawtooth constant-force clamp caps the clamping force mechanically, Sec. II-B, III)
Leader and haptics
a "master manipulator" box stands on the bench in the platform photo (Fig. 12); its design is N.R.; the tests were run from upper-computer software (Sec. V)
Imaging
N.R.
Sterility
disposable surgical box holds every part that touches the catheter; reusable power unit below (Sec. II-B, Fig. 5)

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
single version (cao2024_cathdrive, IDITR 2024); Alekseeva 2025 Table 10 lists a design by the same authors (W. Wang et al. 2024, clamping bracket block) as TRL2 concept, not catalogued

Sources

  1. Design of Catheter Drive Mechanism for Vascular Interventional Surgery Robot · 2024 3rd International Conference on Innovations and Development of Information Technologies and Robotics (IDITR), 2024 · doi:10.1109/IDITR62018.2024.10554319 [cao2024_cathdrive]
  2. The Design of Endovascular Robots for Cardiac Interventions: A Survey · Journal of Field Robotics, 2024 · doi:10.1002/rob.22480 [alekseeva2025_design]

Notes

Notes
N2 2026-11-08, from review_comparison §1.2 row 26. Main picture is CAD: the only prototype photo (Fig. 12, platform) shows the drive about 200 px wide inside a red box; that photo is gallery picture 01. Alekseeva Fig. 12(a,b) draws the generic roller principle only.

This record

Read level
full — full (cao2024_cathdrive); review-only (alekseeva2025_design Table 9 row 5)
Picture
cao2024_cathdrive Fig. 1 overall CAD diagram, p. 2 ↗ — licence: publisher copyright (IEEE); grade C
Picture changes
native extract; cropped off the label row and leader lines (box 0,84,620,480); two dashed red group boxes remain across the model; not upscaled
Found by
gap: Gap searches for robots the other searches missed.
Data
JSON · YAML: records/robots/ujn_cao2024.yaml
Correction
Suggest a correction [repository not created yet]