Catheter Robot Atlas

Academic bedside robots · unitedimaging_yao2025

United Imaging

United Imaging vascular interventional robot (VascularPilot3D platform)
United Imaging and United Imaging Research Institute of Intelligent Imaging, with U. Michigan (Ghaffari); also used by Tongji University (Tianliang Yao, Peng Qi), Soochow University (Bo Lu), Siemens Healthineers · 2024–2025
United Imaging: main picture1 / 3
Song et al., 2025 IEEE International Conference on Robotics and Automation (ICRA) (2025), Fig. 5. CC BY 4.0. Changes: native extract; cropped to the top panel; nothing erased; none upscaled. Source ↗ Picture grade A.
Family
Academic bedside robots (academic)
Devices
guidewire (simulated 0.9 mm radius, 1.55 g/cm3)
Degrees of freedom
translation + rotation of the guidewire (Fig. 3 labels); count N.R.
Push
Not reported
Twist
Not reported
Tested up to
Phantom
Sim-to-real
Yes — A controller or perception model trained in simulation ran on this robot.

Testing

Highest level
Phantom
What was tested
Physical robot, Target B of a vascular phantom; 5 trials per RL method, 3 trials per human operator (yao2025_sim2real); autonomous navigation to 5 destinations in a transparent phantom under fluoroscopy, 5 trials each, which the paper calls ex-vivo (song2025_vascularpilot3d Sec. III, Fig. 5)
Key numbers
PPO+GAE+DR: 100 % success, 53.0 +/- 5.1 s, error 2.92 +/- 0.41 mm (A) and 2.06 +/- 0.31 mm (B); plain PPO 60 % in 61.0 +/- 4.8 s; SAC 75 % (45.0 s); DDPG 75 % (40.0 s); humans 100 % in 35.0 and 39.0 s (yao2025_sim2real); VascularPilot3D 25 of 25 autonomous navigations, 18.38 % fewer control loops than the human operator, one control step about 2 s (song2025_vascularpilot3d abstract, Sec. III)

Mechanism

Push (translation)
N.R. (robot only described as a 'specialized vascular intervention robot provided by United Imaging')
Twist (rotation)
N.R.
Device range
N.R.
Target anatomy
simplified vascular phantom and complex coronary phantom

Sensing and interface

Force sensing
N.R.
Leader and haptics
N.R.
Imaging
United Imaging uAngio 960 fluoroscopy, mean delay 0.2 s from the C-arm to the navigation software; pre-operative CT on uCT 860 (song2025_vascularpilot3d Sec. III)
Sterility
N.R.

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
SOFA with BeamAdapter discrete-point guidewire model; anatomy from CT, built with Blender and Onshape; Sim4EndoR platform (yao2025_sim4endor)
Sim-trained controller
yes: policy trained only in simulation (PPO with GAE and domain randomisation over friction and tissue elasticity) deployed on this robot with no real-time position feedback; sim 90 % (9/10, Target B) -> robot 100 % (5/5) (yao2025_sim2real, full). yao2025_sim4endor: robot 5/5 and 3/5, open-loop replay (S7b evidence)
Studies on this robot

Versions

Lineage
song2025_vascularpilot3d (arXiv 2405.09375 v1 2024-05-15; ICRA 2025): United Imaging's own robot carrying the VascularPilot3D navigation stack ('Our own robot was built', Sec. III, Fig. 5) -> yao2025_sim4endor (ICRA 2025, open-loop replay) -> yao2025_sim2real (T-ASE 2025)

Sources

  1. VascularPilot3D: Toward a 3D Fully Autonomous Navigation for Endovascular Robotics · 2025 IEEE International Conference on Robotics and Automation (ICRA), 2025 · doi:10.1109/icra55743.2025.11127370 [song2025_vascularpilot3d]
  2. Sim2Real Learning With Domain Randomization for Autonomous Guidewire Navigation in Robotic-Assisted Endovascular Procedures · IEEE Transactions on Automation Science and Engineering, 2025 · doi:10.1109/tase.2025.3555559 [yao2025_sim2real]
  3. Sim4EndoR: A Reinforcement Learning Centered Simulation Platform for Task Automation of Endovascular Robotics · 2025 IEEE International Conference on Robotics and Automation (ICRA), 2025 · doi:10.1109/icra55743.2025.11127627 [yao2025_sim4endor]

Notes

Notes
United Imaging built the robot itself ('Our own robot was built', song2025_vascularpilot3d Sec. III) and runs its navigation software on ROS on a laptop; no paper describes the drive mechanism. Same white two-cassette bedside unit on an arm as in yao2025_sim2real Fig. 3. Found 2026-11-07 by the user (arXiv 2405.09375), missed by the searches because the title names the navigation system, not the robot.

This record

Read level
full — full (song2025_vascularpilot3d; yao2025_sim2real); S7b evidence (sim4endor)
Picture
song2025_vascularpilot3d Fig. 5, top panel (arXiv 2405.09375v2, p. 5, native raster 1052 x 1462) ↗ — licence: CC BY 4.0; grade A
Picture changes
native extract; cropped to the top panel; nothing erased; none upscaled
Found by
academic: Search of academic bedside robots.
Data
JSON · YAML: records/robots/unitedimaging_yao2025.yaml
Correction
Suggest a correction [repository not created yet]