Catheter Robot Atlas

Academic bedside robots · tmdu_harada2020

TMDU pneumatic

TMDU pneumatically driven catheter-gripping robot for hot-balloon ablation
Tokyo Medical and Dental University, Institute of Biomaterials and Bioengineering (Kenji Kawashima, Tetsuro Miyazaki, Takahiro Kanno) with Tokyo Institute of Technology (Toshihiro Kawase) · 2020
TMDU pneumatic: main picture1 / 3
Not cleared for distribution. harada2020_development Fig. 2 (catheter gripping robot photo), SII 2020, doi:10.1109/SII46433.2020.9026176. Licence: publisher copyright (IEEE). Picture grade C.
Family
Academic bedside robots (academic)
Devices
SATAKE HotBalloon catheter (Toray) and its sheath
Degrees of freedom
1 (push/pull)
Push
Other drives
Twist
None
Tested up to
Animal
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Animal
What was tested
pulmonary-vein blocking with the balloon in a live pig (one animal described); anomaly detection tested on an excised pig heart (harada2020_development Sec. IV-V)
Key numbers
push force about 10 N; gripper magnets 11.8 N (9 mm) and 4.3 N (5 mm) (harada2020_development Sec. III)

Mechanism

Push (translation)
other: pneumatic air cylinder (100 mm stroke) moving a magnet-closed catheter gripper
Twist (rotation)
none
Device range
N.R.
Target anatomy
left atrium / pulmonary veins (hot-balloon catheter ablation)

Sensing and interface

Force sensing
contact force estimated from cylinder air pressure; magnetic distance sensor on the gripper
Leader and haptics
haptic display of the pressure-estimated force (prototype)
Imaging
X-ray fluoroscopy
Sterility
N.R.

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
single prototype (harada2020_development)

Sources

  1. Development of Pneumatically Driven Surgical Robot for Catheter Ablation · 2020 IEEE/SICE International Symposium on System Integration (SII), 2020 · doi:10.1109/sii46433.2020.9026176 [harada2020_development]

Notes

Notes
Back-drivable pneumatic actuation (Kawashima's surgical-robot line).

This record

Read level
full
Picture
harada2020_development Fig. 2 (catheter gripping robot photo), SII 2020, doi:10.1109/SII46433.2020.9026176 ↗ — licence: publisher copyright (IEEE); grade C
Picture changes
native extract of the embedded photo (132 x 237, stored rotated) and rotated 90 deg to the page orientation; no crop, no erase; not upscaled
Found by
gap: Gap searches for robots the other searches missed.
Data
JSON · YAML: records/robots/tmdu_harada2020.yaml
Correction
Suggest a correction [repository not created yet]