Catheter Robot Atlas

Academic bedside robots · tju_he2018

TJU linear-stepping

Tianjin University linear-stepping endovascular robot with variable-stiffness SMP catheter and FBG tip-force sensing
Tianjin University, Key Laboratory of Mechanism Theory and Equipment Design (Chengbin He, Shuxin Wang, Siyang Zuo) · 2018
TJU linear-stepping: main picture1 / 4
Not cleared for distribution. he2018_linear Fig. 2(b) prototype photo (native raster p. 4). Licence: publisher copyright. Picture grade A.
Family
Academic bedside robots (academic)
Devices
custom catheter: shape-memory-polymer tube 2.5 mm OD / 1.0 mm ID with water cooling, silicone outer sheath 5 mm x 90 mm, four FBG fibres at the tip (he2018_linear Sec. Methods)
Degrees of freedom
2 (linear + rotation)
Push
Clamp relay
Twist
Rotating clamp
Tested up to
Phantom
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Phantom
What was tested
vascular phantom insertion in rigid vs flexible state with tip-force monitoring (he2018_linear Fig. 11)
Key numbers
two electromagnet clutches (front + travel); Faulhaber LM1247-100-11 linear motor, max 100 mm per step; insertion 10 mm/s in the phantom test; SMP Tg 35 deg C; rigid catheter holds its shape against a 1.8 N·cm load (he2018_linear Abstract, Sec. Methods/Results)

Mechanism

Push (translation)
clamp-relay
Twist (rotation)
brushless DC servomotor (Faulhaber 1226S012B) turns the travel clutch on a gear pair through a conducting rotation ring (he2018_linear Sec. Linear stepping actuation unit)
Device range
the custom catheter above only
Target anatomy
aortic arch phantom (ascending/descending aorta)

Sensing and interface

Force sensing
distal: four FBG sensors (0.16 mm) at the catheter tip with temperature compensation
Leader and haptics
no (PC user interface)
Imaging
N.R.
Sterility
N.R.

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
single version (he2018_linear)

Sources

  1. A linear stepping endovascular intervention robot with variable stiffness and force sensing · International Journal of Computer Assisted Radiology and Surgery, 2018 · doi:10.1007/s11548-018-1722-x [he2018_linear]

Notes

Notes
Distinct from A2's tianjin_song2023 (Tianjin University of Technology). Robot and actuated catheter in one system; the bedside unit is a clamp-relay drive in the Nagoya 'propelling pencil' tradition (he2018_linear Introduction).

This record

Read level
full
Picture
he2018_linear Fig. 2(b) prototype photo (native raster p. 4) ↗ — licence: publisher copyright; grade A
Picture changes
native extract; cropped to panel (b) and above the '(b)' panel letter; not upscaled
Found by
gap: Gap searches for robots the other searches missed.
Data
JSON · YAML: records/robots/tju_he2018.yaml
Correction
Suggest a correction [repository not created yet]