TMVR delivery

Georgia Tech telerobotic transcatheter delivery system for a mitral valve implant
Georgia Institute of Technology, RoboMed Laboratory (J. P. Desai; R. Qi, N. U. Nayar) · 2020–2023
TMVR delivery: main picture1 / 9
Not cleared for distribution. qi2023_telerobotic Fig. 2 (system CAD: actuation system on the stabiliser with the guide tube), RA-L 2023, doi:10.1109/LRA.2023.3265592; native embedded raster p. 2 (3166 x 1136), which carries none of the vector labels. Licence: publisher copyright (IEEE). Picture grade B.
Family
Actuated-tip devices (actuated-device)
Devices
own robotic catheter (5.7 mm OD) on a 1.11 m reinforced guide tube inside a customised steerable outer sheath (VERGO, Duke Extrusion; 6.6 mm ID, 8.3 mm OD, 990.6 mm) (qi2023 Abstract, Sec. IV-A)
Degrees of freedom
4 (proximal bending +-90 deg, distal bending +-120 deg, prismatic 70 mm actuatable and restricted to 60 mm, distal torsion +-75 deg; qi2023 Sec. II)
Push
Tip-actuated
Twist
Tendon bending only
Tested up to
Phantom
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Phantom
What was tested
implant delivery in a realistic human cardiovascular phantom (United Biologics) with a ViVitro pulsatile pump (63.1 ml/stroke, 70 beats/min), one demonstration; tube validation with the implant free and against the leaflet; no animal study yet (qi2023 Sec. IV-A, IV-B, V)
Key numbers
joint tracking RMSE 0.20, 0.74, 0.02 and 0.17 mm (qi2023 Sec. IV-A); Maxon 0.5 W 8 mm DC motors (64:1 bending, 16:1 torsion) and Pololu 100:1 prismatic gearmotor, tendon force up to about 20 N (Sec. II); implant advanced 40 mm under the leaflet (Sec. IV-A); tip deflection about 2 mm during a 15 s hold (Sec. IV-B)

Mechanism

Push (translation)
tendon/tip-actuation (distal prismatic joint: the quadlumen tube is advanced by a Pololu 100:1 gearmotor through a lead screw on a linear guide; tendon-driven bending and torsion joints; actuation unit on a stabiliser with a linear guide rail; qi2023 Sec. II)
Twist (rotation)
tendon-driven distal torsion joint: two antagonistic tendons on a lead-screw holder driven by a DC motor with a 16:1 gearbox (qi2023 Sec. II)
Device range
own catheter
Target anatomy
mitral valve (transseptal implant delivery)

Sensing and interface

Force sensing
none reported
Leader and haptics
wireless Xbox 360 controller, no haptics
Imaging
N.R.
Sterility
N.R.; waterproofed with two three-way stopcocks for saline introduction (qi2023 Sec. II)

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
analytical static deflection model of the reinforced quadlumen tube (eqs. 1-10), checked against ANSYS 18.2 FEM and a camera bench test in 12 configurations (qi2023 Sec. III-C, Table I); not a model of the whole robot
Sim-trained controller
none found

Versions

Lineage
2020 IROS first robotic transcatheter delivery system (title); ICRA 2021, T-MRB 2022 and ICRA 2023 papers of Nayar, Qi and Desai (qi2023 refs. [18]-[20], not read); 2023 RAL telerobotic system: reinforced guide tube in a 24 Fr commercial steerable sheath, stronger actuation, waterproofing, stabiliser, wireless joystick (qi2023, full)

Sources

  1. Telerobotic Transcatheter Delivery System for Mitral Valve Implant · IEEE Robotics and Automation Letters, 2023 · doi:10.1109/lra.2023.3265592 [qi2023_telerobotic]
  2. Towards the Development of a Robotic Transcatheter Delivery System for Mitral Valve Implant · 2020 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS), 2020 · doi:10.1109/iros45743.2020.9341551 [nayar2020_towards]

Notes

Notes
Tip-actuated device, so category actuated-device; Desai's COAST and balloon guidewires are in robots_A5. F4 2026-10-14: no clean photo of the whole robot in the paper (Figs. 1, 7(a), 8 show the unit small or with pointers).

This record

Read level
full — full (IEEE PDF, 2026-10-14)
Picture
qi2023_telerobotic Fig. 2 (system CAD: actuation system on the stabiliser with the guide tube), RA-L 2023, doi:10.1109/LRA.2023.3265592; native embedded raster p. 2 (3166 x 1136), which carries none of the vector labels ↗ — licence: publisher copyright (IEEE); grade B
Picture changes
native extract; cropped to the actuation system and stabiliser (guide tube cut at the right edge so the detail insets stay out); 40 px of white added at the top as margin (render background is pure white); no other edits; not upscaled
Found by
gap: Gap searches for robots the other searches missed.
Data
JSON · YAML: records/robots/gatech_tmvr_qi2023.yaml
Correction
Suggest a correction [repository not created yet]