 "schema_version": 1,
 "id": "tmdu_harada2020",
 "name": "TMDU pneumatic",
 "full_name": "TMDU pneumatically driven catheter-gripping robot for hot-balloon ablation",
 "group": "Tokyo Medical and Dental University, Institute of Biomaterials and Bioengineering (Kenji Kawashima, Tetsuro Miyazaki, Takahiro Kanno) with Tokyo Institute of Technology (Toshihiro Kawase)",
 "group_short": "TMDU",
 "family": "academic",
 "category": "academic",
 "first_year": 2020,
 "latest_year": 2020,
 "versions": "single prototype (harada2020_development)",
 "target_anatomy": "left atrium / pulmonary veins (hot-balloon catheter ablation)",
 "devices": [
  "sheath",
  "therapeutic-device"
 ],
 "devices_driven": "SATAKE HotBalloon catheter (Toray) and its sheath",
 "n_dof": 1,
 "n_dof_detail": "1 (push/pull)",
 "translation_mechanism": "other",
 "translation_detail": "other: pneumatic air cylinder (100 mm stroke) moving a magnet-closed catheter gripper",
 "rotation_mechanism": "none",
 "rotation_detail": "none",
 "device_range": "N.R.",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "contact force estimated from cylinder air pressure; magnetic distance sensor on the gripper",
 "haptics_leader": "haptic display of the pressure-estimated force (prototype)",
 "imaging_compat": "X-ray fluoroscopy",
 "sterility": "N.R.",
 "highest_validation": "animal",
 "validation_detail": "pulmonary-vein blocking with the balloon in a live pig (one animal described); anomaly detection tested on an excised pig heart (harada2020_development Sec. IV-V)",
 "key_numbers": "push force about 10 N; gripper magnets 11.8 N (9 mm) and 4.3 N (5 mm) (harada2020_development Sec. III)",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "C",
  "kind": "photo",
  "source": "harada2020_development Fig. 2 (catheter gripping robot photo), SII 2020, doi:10.1109/SII46433.2020.9026176",
  "source_url": "https://doi.org/10.1109/SII46433.2020.9026176",
  "licence": "publisher copyright (IEEE)",
  "credit": "Harada et al., 2020 IEEE/SICE International Symposium on System Integration (SII) (2020), Fig. 2. publisher copyright (IEEE).",
  "edits": "native extract of the embedded photo (132 x 237, stored rotated) and rotated 90 deg to the page orientation; no crop, no erase; not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "tmdu_harada2020_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Catheter gripping robot at the table during the swine pulmonary-vein blocking test",
   "source_key": "harada2020_development",
   "source": "Harada et al., 2020 IEEE/SICE International Symposium on System Integration (SII) (2020), Fig. 7, p. 4",
   "source_url": "https://doi.org/10.1109/sii46433.2020.9026176",
   "where": "Fig. 7, p. 4",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "tmdu_harada2020_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Sheath gripper part with the lid open, silicone seat removed",
   "source_key": "harada2020_development",
   "source": "Harada et al., 2020 IEEE/SICE International Symposium on System Integration (SII) (2020), Fig. 3(b), p. 2",
   "source_url": "https://doi.org/10.1109/sii46433.2020.9026176",
   "where": "Fig. 3(b), p. 2",
   "licence": "publisher copyright (IEEE)"
  }
 ],
 "sources": [
  {
   "key": "harada2020_development",
   "doi": "10.1109/sii46433.2020.9026176",
   "url": null,
   "title": "Development of Pneumatically Driven Surgical Robot for Catheter Ablation",
   "year": 2020,
   "venue": "2020 IEEE/SICE International Symposium on System Integration (SII)"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full",
 "notes": "Back-drivable pneumatic actuation (Kawashima's surgical-robot line).",
 "maintainers_own": false,
 "found_by": "gap"
}