 "schema_version": 1,
 "id": "tju_he2018",
 "name": "TJU linear-stepping",
 "full_name": "Tianjin University linear-stepping endovascular robot with variable-stiffness SMP catheter and FBG tip-force sensing",
 "group": "Tianjin University, Key Laboratory of Mechanism Theory and Equipment Design (Chengbin He, Shuxin Wang, Siyang Zuo)",
 "group_short": "Tianjin U.",
 "family": "academic",
 "category": "academic",
 "first_year": 2018,
 "latest_year": 2018,
 "versions": "single version (he2018_linear)",
 "target_anatomy": "aortic arch phantom (ascending/descending aorta)",
 "devices": [
  "catheter",
  "sheath"
 ],
 "devices_driven": "custom catheter: shape-memory-polymer tube 2.5 mm OD / 1.0 mm ID with water cooling, silicone outer sheath 5 mm x 90 mm, four FBG fibres at the tip (he2018_linear Sec. Methods)",
 "n_dof": 2,
 "n_dof_detail": "2 (linear + rotation)",
 "translation_mechanism": "clamp-relay",
 "translation_detail": "clamp-relay",
 "rotation_mechanism": "rotating-clamp",
 "rotation_detail": "brushless DC servomotor (Faulhaber 1226S012B) turns the travel clutch on a gear pair through a conducting rotation ring (he2018_linear Sec. Linear stepping actuation unit)",
 "device_range": "the custom catheter above only",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "distal: four FBG sensors (0.16 mm) at the catheter tip with temperature compensation",
 "haptics_leader": "no (PC user interface)",
 "imaging_compat": "N.R.",
 "sterility": "N.R.",
 "highest_validation": "phantom",
 "validation_detail": "vascular phantom insertion in rigid vs flexible state with tip-force monitoring (he2018_linear Fig. 11)",
 "key_numbers": "two electromagnet clutches (front + travel); Faulhaber LM1247-100-11 linear motor, max 100 mm per step; insertion 10 mm/s in the phantom test; SMP Tg 35 deg C; rigid catheter holds its shape against a 1.8 N·cm load (he2018_linear Abstract, Sec. Methods/Results)",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "A",
  "kind": "photo",
  "source": "he2018_linear Fig. 2(b) prototype photo (native raster p. 4)",
  "source_url": "https://doi.org/10.1007/s11548-018-1722-x",
  "licence": "publisher copyright",
  "credit": "He et al., International Journal of Computer Assisted Radiology and Surgery (2018), Fig. 2(b). publisher copyright.",
  "edits": "native extract; cropped to panel (b) and above the '(b)' panel letter; not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "tju_he2018_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Linear-stepping robot inserting the catheter into the aortic phantom",
   "source_key": "he2018_linear",
   "source": "He et al., International Journal of Computer Assisted Radiology and Surgery (2018), Fig. 11(a), p. 10",
   "source_url": "https://doi.org/10.1007/s11548-018-1722-x",
   "where": "Fig. 11(a), p. 10",
   "licence": "publisher copyright"
  },
  {
   "id": "tju_he2018_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Configuration of the linear stepping actuation unit",
   "source_key": "he2018_linear",
   "source": "He et al., International Journal of Computer Assisted Radiology and Surgery (2018), Fig. 2(a), p. 4",
   "source_url": "https://doi.org/10.1007/s11548-018-1722-x",
   "where": "Fig. 2(a), p. 4",
   "licence": "publisher copyright"
  },
  {
   "id": "tju_he2018_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Variable-stiffness SMP catheter prototype with FBG fibres",
   "source_key": "he2018_linear",
   "source": "He et al., International Journal of Computer Assisted Radiology and Surgery (2018), Fig. 5(b), p. 6",
   "source_url": "https://doi.org/10.1007/s11548-018-1722-x",
   "where": "Fig. 5(b), p. 6",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "he2018_linear",
   "doi": "10.1007/s11548-018-1722-x",
   "url": null,
   "title": "A linear stepping endovascular intervention robot with variable stiffness and force sensing",
   "year": 2018,
   "venue": "International Journal of Computer Assisted Radiology and Surgery"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full",
 "notes": "Distinct from A2's tianjin_song2023 (Tianjin University of Technology). Robot and actuated catheter in one system; the bedside unit is a clamp-relay drive in the Nagoya 'propelling pencil' tradition (he2018_linear Introduction).",
 "maintainers_own": false,
 "found_by": "gap"
}