Academic bedside robots · sjtumr_huang2022

SJTU MR wire drive

SJTU pneumatic MR-conditional guidewire delivery mechanism with decoupled actuation
Shanghai Jiao Tong University, Institute of Medical Robotics (Anzhu Gao, Guang-Zhong Yang; Shaoping Huang) with Empa · 2022
SJTU MR wire drive: main picture1 / 4
Not cleared for distribution. huang2022_pneumatic Fig. 1(a) (native raster p. 2). Licence: publisher copyright. Picture grade B.
Family
Academic bedside robots (academic)
Devices
guidewire
Degrees of freedom
2 (translation + rotation, decoupled)
Push
Friction rollers
Twist
Rotating drive unit
Tested up to
Phantom
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Phantom
What was tested
real-time MRI-guided guidewire delivery in an aorta phantom (TRANDOMED SJK001F) (huang2022_pneumatic Sec. IV-C)
Key numbers
resolution 0.238 mm/step and 0.4 deg/step; mean translational error 0.37 mm; rotary motor torque 15.75 N·m/MPa as printed; pneumatic clamp 9 mm bore, 63 N/MPa; SNR loss < 2% (huang2022_pneumatic Abstract, Sec. II-III)

Mechanism

Push (translation)
friction-roller
Twist (rotation)
other: pneumatic slip ring rotates the whole wheel unit, driven by a rotary pneumatic step motor through 1:5 spur gears
Device range
N.R.
Target anatomy
coronary (PCI) under MRI; aorta phantom

Sensing and interface

Force sensing
none
Leader and haptics
handheld control (huang2022_pneumatic Fig. 4); haptics N.R.
Imaging
MR-conditional: pneumatic step motors, 3D-printed parts; tested in a Siemens MAGNETOM Aera 1.5 T
Sterility
N.R.

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
single version (huang2022_pneumatic)

Sources

  1. A Pneumatic MR-conditional Guidewire Delivery Mechanism with Decoupled Actuations for Endovascular Intervention · 2022 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS), 2022 · doi:10.1109/iros47612.2022.9981596 [huang2022_pneumatic]

Notes

Notes
Third SJTU robot in the catalogue (A2 has sjtu_xie_shen2019 and sjtu_coronary_meng2025); this one is MR-conditional. Photo in the MR room (Fig. 4) is 573 px.

This record

Read level
full
Picture
huang2022_pneumatic Fig. 1(a) (native raster p. 2) ↗ — licence: publisher copyright; grade B
Picture changes
native extract; 50 px white margin added; not upscaled
Found by
gap: Gap searches for robots the other searches missed.
Data
JSON · YAML: records/robots/sjtumr_huang2022.yaml
Correction
Suggest a correction [repository not created yet]