Catheter Robot Atlas

Commercial and clinical-stage systems · roses_danieli2022

ROSES

ROSES (RObotic System for Endovascular Surgery)
Calabrian High Tech (CHT S.r.l.; G. Danieli) with Magna Graecia University (C. Indolfi, A. Polimeni, S. De Rosa) · 2018–2023
ROSES: main picture1 / 9
Danieli et al., Applied Mechanics (2022), Fig. 12. CC BY 4.0. Changes: full-size MDPI figure image; cropped to the centre unit; not upscaled. Source ↗ Picture grade B.
Family
Commercial and clinical-stage systems (clinical-stage)
Devices
guidewire plus balloon or stent catheter per slave; first-in-human: 6 Fr JR4 guide, 0.014 in BMW wire, 2.0 x 12 mm balloon, 2.5 x 15 mm DES (indolfi2023_first)
Degrees of freedom
3 per slave (catheter advance/retract, guidewire advance/retract, common rotation); one console drives up to three slaves (techbriefs 2019 entry)
Push
Friction rollers
Twist
Rotating drive unit
Tested up to
Human
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Human
What was tested
in vitro tubing and X-ray tests; first patient 22 Jan 2019 (two stents), trials 'still in progress' in 2022 (danieli2022_development Sec. 2); 2023 case report of one 64-year-old patient (R-PCI of a PL branch)
Key numbers
case: procedure 35 min, contrast 110 mL, stent result satisfactory at 2.5-year follow-up (indolfi2023_first); fluoroscopy time N.R.

Mechanism

Push (translation)
friction-roller (3-D-printed disposables convert gear rotation into advance/retract through friction wheels)
Twist (rotation)
the disposable is rotated by a concentric gear train in the robot actuator (danieli2022_development Sec. 2)
Device range
0.014 in guidewires to large catheters (36 mm passage in v3); exact Fr range N.R.
Target anatomy
coronary PCI; endovascular surgery, large catheters for aneurysm repair or TAVI (cart version)

Sensing and interface

Force sensing
resistance to advancement shown on the console; cart version measures forces opposed by the patient with a force transducer (danieli2022_development abstract, Sec. 3)
Leader and haptics
passive master console with joysticks per track (one for the guidewire, one for devices); no force rendering reported
Imaging
fluoroscopy; cockpit inside or outside the cath lab
Sterility
sterile disposables in contact with the devices, split upper/lower so they can be removed without withdrawing the device (danieli2022_development abstract)

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
2018 PCT patent (ref. 6 of indolfi2023_first); robot actuator v1 (motor axes parallel to the gear train), v2 (motors perpendicular, thinner), v3 (36 mm central passage for haemostasis valve and large catheters) (danieli2022_development Fig. 12); disposables for angioplasty (DA), big catheters (DEBC), movable-core guidewires and a tip-curvature catheter; cart with rail and force transducer for endovascular surgery (Fig. 13); first patient 22 Jan 2019; 2023 case report

Sources

  1. Development of Disposables and Accessories for ROSES and Their In Vitro Experimentation · Applied Mechanics, 2022 · doi:10.3390/applmech3030054 [danieli2022_development]
  2. First-in-human robotic percutaneous coronary intervention by the ROSES robotic system: a case report · European Heart Journal - Case Reports, 2023 · doi:10.1093/ehjcr/ytad506 [indolfi2023_first]

Notes

Notes
In S8_systems.csv but in no catalogue file before A9. A larger multicentre pivotal trial is described as ongoing in the case report (status not found).

This record

Read level
full — full (danieli2022_development, indolfi2023_first); web (techbriefs 2019 contest entry)
Picture
danieli2022_development Fig. 12, centre unit (robot actuator, second edition); https://mdpi-res.com/applmech/applmech-03-00054/article_deploy/html/images/applmech-03-00054-g012.png (article https://doi.org/10.3390/applmech3030054) ↗ — licence: CC BY 4.0; grade B
Picture changes
full-size MDPI figure image; cropped to the centre unit; not upscaled
Found by
gap: Gap searches for robots the other searches missed.
Data
JSON · YAML: records/robots/roses_danieli2022.yaml
Correction