 "schema_version": 1,
 "id": "roses_danieli2022",
 "name": "ROSES",
 "full_name": "ROSES (RObotic System for Endovascular Surgery)",
 "group": "Calabrian High Tech (CHT S.r.l.; G. Danieli) with Magna Graecia University (C. Indolfi, A. Polimeni, S. De Rosa)",
 "group_short": "Calabrian High Tech",
 "family": "commercial",
 "category": "clinical-stage",
 "first_year": 2018,
 "latest_year": 2023,
 "versions": "2018 PCT patent (ref. 6 of indolfi2023_first); robot actuator v1 (motor axes parallel to the gear train), v2 (motors perpendicular, thinner), v3 (36 mm central passage for haemostasis valve and large catheters) (danieli2022_development Fig. 12); disposables for angioplasty (DA), big catheters (DEBC), movable-core guidewires and a tip-curvature catheter; cart with rail and force transducer for endovascular surgery (Fig. 13); first patient 22 Jan 2019; 2023 case report",
 "target_anatomy": "coronary PCI; endovascular surgery, large catheters for aneurysm repair or TAVI (cart version)",
 "devices": [
  "guidewire",
  "therapeutic-device"
 ],
 "devices_driven": "guidewire plus balloon or stent catheter per slave; first-in-human: 6 Fr JR4 guide, 0.014 in BMW wire, 2.0 x 12 mm balloon, 2.5 x 15 mm DES (indolfi2023_first)",
 "n_dof": 3,
 "n_dof_detail": "3 per slave (catheter advance/retract, guidewire advance/retract, common rotation); one console drives up to three slaves (techbriefs 2019 entry)",
 "translation_mechanism": "friction-roller",
 "translation_detail": "friction-roller (3-D-printed disposables convert gear rotation into advance/retract through friction wheels)",
 "rotation_mechanism": "rotating-drive-unit",
 "rotation_detail": "the disposable is rotated by a concentric gear train in the robot actuator (danieli2022_development Sec. 2)",
 "device_range": "0.014 in guidewires to large catheters (36 mm passage in v3); exact Fr range N.R.",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "resistance to advancement shown on the console; cart version measures forces opposed by the patient with a force transducer (danieli2022_development abstract, Sec. 3)",
 "haptics_leader": "passive master console with joysticks per track (one for the guidewire, one for devices); no force rendering reported",
 "imaging_compat": "fluoroscopy; cockpit inside or outside the cath lab",
 "sterility": "sterile disposables in contact with the devices, split upper/lower so they can be removed without withdrawing the device (danieli2022_development abstract)",
 "highest_validation": "human",
 "validation_detail": "in vitro tubing and X-ray tests; first patient 22 Jan 2019 (two stents), trials 'still in progress' in 2022 (danieli2022_development Sec. 2); 2023 case report of one 64-year-old patient (R-PCI of a PL branch)",
 "key_numbers": "case: procedure 35 min, contrast 110 mL, stent result satisfactory at 2.5-year follow-up (indolfi2023_first); fluoroscopy time N.R.",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "cleared",
  "grade": "B",
  "kind": "photo",
  "source": "danieli2022_development Fig. 12, centre unit (robot actuator, second edition); https://mdpi-res.com/applmech/applmech-03-00054/article_deploy/html/images/applmech-03-00054-g012.png (article https://doi.org/10.3390/applmech3030054)",
  "source_url": "https://doi.org/10.3390/applmech3030054",
  "licence": "CC BY 4.0",
  "credit": "Danieli et al., Applied Mechanics (2022), Fig. 12. CC BY 4.0.",
  "edits": "full-size MDPI figure image; cropped to the centre unit; not upscaled",
  "file": "pictures/roses_danieli2022.jpg"
 },
 "pictures": [
  {
   "id": "roses_danieli2022_01",
   "status": "cleared",
   "file": "pictures/extra/roses_danieli2022_01.jpg",
   "thumb": "pictures/thumbs/roses_danieli2022_01.jpg",
   "kind": "in-use",
   "caption": "First patient case: operator at the ROSES console, robot actuator on the table",
   "source_key": "danieli2022_development",
   "source": "Danieli et al., Applied Mechanics (2022), Fig. 9(a), p. 6",
   "source_url": "https://doi.org/10.3390/applmech3030054",
   "where": "Fig. 9(a), p. 6",
   "licence": "CC BY 4.0"
  },
  {
   "id": "roses_danieli2022_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "photo",
   "caption": "Bedside power unit on its stand and close-up of the functional unit",
   "source_key": "indolfi2023_first",
   "source": "Indolfi et al., European Heart Journal - Case Reports (2023), Fig. 1(B-C), p. 3",
   "source_url": "https://doi.org/10.1093/ehjcr/ytad506",
   "where": "Fig. 1(B-C), p. 3",
   "licence": "CC BY-NC 4.0"
  },
  {
   "id": "roses_danieli2022_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "leader",
   "caption": "Interventional cardiologist at the ROSES console with joystick and touch screen",
   "source_key": "indolfi2023_first",
   "source": "Indolfi et al., European Heart Journal - Case Reports (2023), graphical abstract, p. 2",
   "source_url": "https://doi.org/10.1093/ehjcr/ytad506",
   "where": "graphical abstract, p. 2",
   "licence": "CC BY-NC 4.0"
  },
  {
   "id": "roses_danieli2022_04",
   "status": "cleared",
   "file": "pictures/extra/roses_danieli2022_04.jpg",
   "thumb": "pictures/thumbs/roses_danieli2022_04.jpg",
   "kind": "photo",
   "caption": "First ROSES prototype ever built, on a wooden bench",
   "source_key": "danieli2022_development",
   "source": "Danieli et al., Applied Mechanics (2022), Fig. 1, p. 2",
   "source_url": "https://doi.org/10.3390/applmech3030054",
   "where": "Fig. 1, p. 2",
   "licence": "CC BY 4.0"
  },
  {
   "id": "roses_danieli2022_05",
   "status": "cleared",
   "file": "pictures/extra/roses_danieli2022_05.jpg",
   "thumb": "pictures/thumbs/roses_danieli2022_05.jpg",
   "kind": "close-up",
   "caption": "Disposable actuator closed and open, showing four friction wheels and spur gears",
   "source_key": "danieli2022_development",
   "source": "Danieli et al., Applied Mechanics (2022), Fig. 6, p. 5",
   "source_url": "https://doi.org/10.3390/applmech3030054",
   "where": "Fig. 6, p. 5",
   "licence": "CC BY 4.0"
  },
  {
   "id": "roses_danieli2022_06",
   "status": "cleared",
   "file": "pictures/extra/roses_danieli2022_06.jpg",
   "thumb": "pictures/thumbs/roses_danieli2022_06.jpg",
   "kind": "in-use",
   "caption": "Early in vitro trial under X-ray with PVC tubing on the angiography table",
   "source_key": "danieli2022_development",
   "source": "Danieli et al., Applied Mechanics (2022), Fig. 8(a), p. 6",
   "source_url": "https://doi.org/10.3390/applmech3030054",
   "where": "Fig. 8(a), p. 6",
   "licence": "CC BY 4.0"
  },
  {
   "id": "roses_danieli2022_07",
   "status": "cleared",
   "file": "pictures/extra/roses_danieli2022_07.jpg",
   "thumb": "pictures/thumbs/roses_danieli2022_07.jpg",
   "kind": "in-use",
   "caption": "First in vitro experiment in water-filled transparent tubing at Catanzaro",
   "source_key": "danieli2022_development",
   "source": "Danieli et al., Applied Mechanics (2022), Fig. 7(a), p. 5",
   "source_url": "https://doi.org/10.3390/applmech3030054",
   "where": "Fig. 7(a), p. 5",
   "licence": "CC BY 4.0"
  },
  {
   "id": "roses_danieli2022_08",
   "status": "cleared",
   "file": "pictures/extra/roses_danieli2022_08.jpg",
   "thumb": "pictures/thumbs/roses_danieli2022_08.jpg",
   "kind": "render",
   "caption": "CAD of the cart arm that holds the robot actuator",
   "source_key": "danieli2022_development",
   "source": "Danieli et al., Applied Mechanics (2022), Fig. 13, p. 9",
   "source_url": "https://doi.org/10.3390/applmech3030054",
   "where": "Fig. 13, p. 9",
   "licence": "CC BY 4.0"
  }
 ],
 "sources": [
  {
   "key": "danieli2022_development",
   "doi": "10.3390/applmech3030054",
   "url": null,
   "title": "Development of Disposables and Accessories for ROSES and Their In Vitro Experimentation",
   "year": 2022,
   "venue": "Applied Mechanics"
  },
  {
   "key": "indolfi2023_first",
   "doi": "10.1093/ehjcr/ytad506",
   "url": null,
   "title": "First-in-human robotic percutaneous coronary intervention by the ROSES robotic system: a case report",
   "year": 2023,
   "venue": "European Heart Journal - Case Reports"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (danieli2022_development, indolfi2023_first); web (techbriefs 2019 contest entry)",
 "notes": "In S8_systems.csv but in no catalogue file before A9. A larger multicentre pivotal trial is described as ongoing in the case report (status not found).",
 "maintainers_own": false,