Catheter Robot Atlas

Actuated-tip devices · ncsu_sma_crews2012

SMA ablation cath.

NC State SMA-actuated robotic ablation catheter (Buckner lab)
North Carolina State University, Mechanical and Aerospace Engineering, Electro-Mechanics Research Laboratory (G. D. Buckner; J. H. Crews, J. H. Wiest); NIH SBIR 5R44HL095227 (no company named) · 2008–2015
SMA ablation cath.: main picture1 / 6
Not cleared for distribution. wiest2015_path Fig. 1 (SMA catheter with handle and joystick); https://cdn.ncbi.nlm.nih.gov/pmc/blobs/efac/4326024/83bb6e389e21/nihms648507f1.jpg (PMC4326024). Licence: publisher copyright (Elsevier; NIH author manuscript in PMC). Picture grade C.
Family
Actuated-tip devices (actuated-device)
Devices
own 2.7 mm catheter with two bending segments, each actuated by four offset SMA tendons (wiest2015 Sec. 2)
Degrees of freedom
4 bending (two segments x bend angle and orientation)
Push
Tip-actuated
Twist
None
Tested up to
Bench
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Bench
What was tested
bench: catheter submerged in a saline tank with EnSite patch electrodes; path optimisation repeated on four point sets; closed-loop control on one optimised 16-point path (wiest2015_path pp. 95-96); 2012: single-tendon rig (4 tendon offsets x 11 prestrains, 5 tests each) and four antagonistic rigs (crews2012_design p. 552-553)
Key numbers
path optimisation about 20 s; step response about 2 s with < 10 deg overshoot; mean steady-state tip error 5.5 mm, called unacceptably large by the authors; NavX uncertainty 7.1 +- 3.5 mm, imprecision +- 0.8 mm (wiest2015_path); 2012: optimal bend radius 7.0 mm (single-crystal model) and 7.6 mm (polycrystalline); experimental optimum prestrain about 3.5-4.5% and tendon offset about 1 mm (crews2012_design p. 555-556)

Mechanism

Push (translation)
tendon/tip-actuation (SMA tendons heated by PWM); no robotic insertion stage
Twist (rotation)
none
Device range
90 and 180 degrees of omnidirectional bending for the proximal and distal segments (wiest2015_path p. 93); own 2.7 mm catheter
Target anatomy
endocardial RF ablation (left atrium)

Sensing and interface

Force sensing
none
Leader and haptics
joystick on the catheter handle
Imaging
modified EnSite NavX (St. Jude) electrode localisation (wiest2015_path p. 94); trakSTAR magnetic tracking on the 2012 test rigs (crews2012_design p. 552)
Sterility
N.R.

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
yes: COMSOL finite-element models of SMA-actuated segments coupled to a genetic algorithm in MATLAB (single-crystal and polycrystalline free-energy SMA models), checked on wire rigs (crews2012_design p. 546-557); MATLAB simulations to tune the 2015 path optimiser (wiest2015_path Sec. 2.6)
Sim-trained controller
none found (simulation used for design optimisation, not for a controller)

Versions

Lineage
2008 Veeramani, Buckner et al.: modelling (Smart Mater. Struct.) and design and control (ASME DSCC) of an SMA-actuated robotic catheter, cited by crews2012 and wiest2015 (not read); 2012 JIMSS design optimisation of single-tendon and antagonistic SMA segments in COMSOL with a genetic algorithm, tested on planar wire rigs (crews2012_design, full); 2015 RAS path optimisation and control of a two-segment, eight-tendon catheter (wiest2015_path, full); the lab's later bedside robot is robots_A3 portable_kantu2025 (NC State/NYU), predecessor of LegoBot

Sources

  1. Design optimization of a shape memory alloy–actuated robotic catheter · Journal of Intelligent Material Systems and Structures, 2012 · doi:10.1177/1045389x12436738 [crews2012_design]
  2. Path optimization and control of a shape memory alloy actuated catheter for endocardial radiofrequency ablation · Robotics and Autonomous Systems, 2015 · doi:10.1016/j.robot.2014.10.019 [wiest2015_path]

Notes

Notes
Seed-lineage row requested by the coordinator: the BIRO Lab robots (Kantu, LegoBot) descend from Buckner's NC State catheter work. | X2 2026-11-08: first_year 2012 -> 2008 from the lab's earlier SMA robotic-catheter papers cited in both full texts (not read). wiest2015 journal PDF matches the PMC manuscript. | X4 2026-11-08: first_year marked (inferred): Veeramani, Buckner et al. 2008 SMA robotic-catheter papers are cited by crews2012 and wiest2015 but not read

This record

Read level
full — full (crews2012_design, wiest2015_path journal PDFs)
Picture
wiest2015_path Fig. 1 (SMA catheter with handle and joystick); https://cdn.ncbi.nlm.nih.gov/pmc/blobs/efac/4326024/83bb6e389e21/nihms648507f1.jpg (PMC4326024) ↗ — licence: publisher copyright (Elsevier; NIH author manuscript in PMC); grade C
Picture changes
PMC figure image as is; not upscaled
Found by
gap: Gap searches for robots the other searches missed.
Data
JSON · YAML: records/robots/ncsu_sma_crews2012.yaml
Correction
Suggest a correction [repository not created yet]