 "schema_version": 1,
 "id": "ncsu_sma_crews2012",
 "name": "SMA ablation cath.",
 "full_name": "NC State SMA-actuated robotic ablation catheter (Buckner lab)",
 "group": "North Carolina State University, Mechanical and Aerospace Engineering, Electro-Mechanics Research Laboratory (G. D. Buckner; J. H. Crews, J. H. Wiest); NIH SBIR 5R44HL095227 (no company named)",
 "group_short": "NC State",
 "family": "tip",
 "category": "actuated-device",
 "first_year": 2008,
 "latest_year": 2015,
 "versions": "2008 Veeramani, Buckner et al.: modelling (Smart Mater. Struct.) and design and control (ASME DSCC) of an SMA-actuated robotic catheter, cited by crews2012 and wiest2015 (not read); 2012 JIMSS design optimisation of single-tendon and antagonistic SMA segments in COMSOL with a genetic algorithm, tested on planar wire rigs (crews2012_design, full); 2015 RAS path optimisation and control of a two-segment, eight-tendon catheter (wiest2015_path, full); the lab's later bedside robot is robots_A3 portable_kantu2025 (NC State/NYU), predecessor of LegoBot",
 "target_anatomy": "endocardial RF ablation (left atrium)",
 "devices": [
  "catheter"
 ],
 "devices_driven": "own 2.7 mm catheter with two bending segments, each actuated by four offset SMA tendons (wiest2015 Sec. 2)",
 "n_dof": 4,
 "n_dof_detail": "4 bending (two segments x bend angle and orientation)",
 "translation_mechanism": "tendon/tip-actuation",
 "translation_detail": "tendon/tip-actuation (SMA tendons heated by PWM); no robotic insertion stage",
 "rotation_mechanism": "none",
 "rotation_detail": "none",
 "device_range": "90 and 180 degrees of omnidirectional bending for the proximal and distal segments (wiest2015_path p. 93); own 2.7 mm catheter",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "none",
 "haptics_leader": "joystick on the catheter handle",
 "imaging_compat": "modified EnSite NavX (St. Jude) electrode localisation (wiest2015_path p. 94); trakSTAR magnetic tracking on the 2012 test rigs (crews2012_design p. 552)",
 "sterility": "N.R.",
 "highest_validation": "bench",
 "validation_detail": "bench: catheter submerged in a saline tank with EnSite patch electrodes; path optimisation repeated on four point sets; closed-loop control on one optimised 16-point path (wiest2015_path pp. 95-96); 2012: single-tendon rig (4 tendon offsets x 11 prestrains, 5 tests each) and four antagonistic rigs (crews2012_design p. 552-553)",
 "key_numbers": "path optimisation about 20 s; step response about 2 s with < 10 deg overshoot; mean steady-state tip error 5.5 mm, called unacceptably large by the authors; NavX uncertainty 7.1 +- 3.5 mm, imprecision +- 0.8 mm (wiest2015_path); 2012: optimal bend radius 7.0 mm (single-crystal model) and 7.6 mm (polycrystalline); experimental optimum prestrain about 3.5-4.5% and tendon offset about 1 mm (crews2012_design p. 555-556)",
 "sim_model": "yes: COMSOL finite-element models of SMA-actuated segments coupled to a genetic algorithm in MATLAB (single-crystal and polycrystalline free-energy SMA models), checked on wire rigs (crews2012_design p. 546-557); MATLAB simulations to tune the 2015 path optimiser (wiest2015_path Sec. 2.6)",
 "has_sim_model": true,
 "sim2real": "none-found",
 "sim2real_detail": "none found (simulation used for design optimisation, not for a controller)",
 "picture": {
  "status": "not-cleared",
  "grade": "C",
  "kind": "photo",
  "source": "wiest2015_path Fig. 1 (SMA catheter with handle and joystick); https://cdn.ncbi.nlm.nih.gov/pmc/blobs/efac/4326024/83bb6e389e21/nihms648507f1.jpg (PMC4326024)",
  "source_url": "https://doi.org/10.1016/j.robot.2014.10.019",
  "licence": "publisher copyright (Elsevier; NIH author manuscript in PMC)",
  "credit": "Wiest et al., Robotics and Autonomous Systems (2015), Fig. 1. publisher copyright (Elsevier; NIH author manuscript in PMC).",
  "edits": "PMC figure image as is; not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "ncsu_sma_crews2012_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "photo",
   "caption": "Catheter segment prototype: solid model and photo of SMA actuators, collets and beam",
   "source_key": "crews2012_design",
   "source": "Crews et al., Journal of Intelligent Material Systems and Structures (2012), Fig. 2(a-b), p. 546",
   "source_url": "https://doi.org/10.1177/1045389x12436738",
   "where": "Fig. 2(a-b), p. 546",
   "licence": "publisher copyright"
  },
  {
   "id": "ncsu_sma_crews2012_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Rendering of the two-segment catheter bent to both sides",
   "source_key": "wiest2015_path",
   "source": "Wiest et al., Robotics and Autonomous Systems (2015), Fig. 3, p. 89",
   "source_url": "https://doi.org/10.1016/j.robot.2014.10.019",
   "where": "Fig. 3, p. 89",
   "licence": "publisher copyright"
  },
  {
   "id": "ncsu_sma_crews2012_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Single-tendon test rig with trakSTAR sensors and the actuator-offset block",
   "source_key": "crews2012_design",
   "source": "Crews et al., Journal of Intelligent Material Systems and Structures (2012), Fig. 8(a-b), p. 552",
   "source_url": "https://doi.org/10.1177/1045389x12436738",
   "where": "Fig. 8(a-b), p. 552",
   "licence": "publisher copyright"
  },
  {
   "id": "ncsu_sma_crews2012_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Antagonistic test rig: flexible beam with two SMA tendons",
   "source_key": "crews2012_design",
   "source": "Crews et al., Journal of Intelligent Material Systems and Structures (2012), Fig. 9, p. 553",
   "source_url": "https://doi.org/10.1177/1045389x12436738",
   "where": "Fig. 9, p. 553",
   "licence": "publisher copyright"
  },
  {
   "id": "ncsu_sma_crews2012_05",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Catheter submerged in a saline tank with EnSite patch electrodes",
   "source_key": "wiest2015_path",
   "source": "Wiest et al., Robotics and Autonomous Systems (2015), Fig. 11(a), p. 95",
   "source_url": "https://doi.org/10.1016/j.robot.2014.10.019",
   "where": "Fig. 11(a), p. 95",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "crews2012_design",
   "doi": "10.1177/1045389x12436738",
   "url": null,
   "title": "Design optimization of a shape memory alloy–actuated robotic catheter",
   "year": 2012,
   "venue": "Journal of Intelligent Material Systems and Structures"
  },
  {
   "key": "wiest2015_path",
   "doi": "10.1016/j.robot.2014.10.019",
   "url": null,
   "title": "Path optimization and control of a shape memory alloy actuated catheter for endocardial radiofrequency ablation",
   "year": 2015,
   "venue": "Robotics and Autonomous Systems"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (crews2012_design, wiest2015_path journal PDFs)",
 "notes": "Seed-lineage row requested by the coordinator: the BIRO Lab robots (Kantu, LegoBot) descend from Buckner's NC State catheter work. | X2 2026-11-08: first_year 2012 -> 2008 from the lab's earlier SMA robotic-catheter papers cited in both full texts (not read). wiest2015 journal PDF matches the PMC manuscript. | X4 2026-11-08: first_year marked (inferred): Veeramani, Buckner et al. 2008 SMA robotic-catheter papers are cited by crews2012 and wiest2015 but not read",
 "maintainers_own": false,
 "found_by": "gap"
}