Catheter Robot Atlas

Magnetic navigation · millispinner_wu2024

Milli-spinner

Magnetic milli-spinner (untethered)
Zhao, Stanford Mechanical Engineering · 2024–2026
Milli-spinner: main picture1 / 6
Not cleared for distribution. Stanford Robotics Center demo page (src.stanford.edu/demo/magnetic-milli-spinner-for-robotic-endovascular-surgery-launch), image "magnetic millispinner (2).jpg" at 2500w (UR arm with rotating magnet over the head phantom). Licence: web image (Stanford Robotics Center), terms not stated. Picture grade A.
Family
Magnetic navigation (magnetic-navigation)
Devices
untethered 2.5 mm OD hollow spinner with helical fins and slits, in 3.5 mm ID tubes
Degrees of freedom
rotating magnetic field from Helmholtz coils; six-axis robot arm holding a permanent magnet for navigation
Push
Magnetic
Twist
Field direction
Tested up to
Phantom
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Phantom
What was tested
vascular phantoms; in-situ embolisation of an aneurysm model filled with anticoagulated porcine blood; no animal
Key numbers
23 cm/s in tubular environment (abstract); stable spinning needs 15-20 mT; 56.4 cm/s predicted by CFD at 9.6k rpm

Mechanism

Push (translation)
magnetic
Twist (rotation)
spinning in rotating field (up to 8.4k rpm tested)
Device range
untethered
Target anatomy
cerebral and general vasculature; phantoms, porcine blood aneurysm model

Sensing and interface

Force sensing
none
Leader and haptics
no
Imaging
fluoroscopy-guided in phantom
Sterility
N.R.

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
CFD of swimming speed compared with experiment (wu2024_millispinner, full)
Sim-trained controller
none found

Versions

Lineage
arXiv 2410.21112; optimisation follow-up Adv. Robotics Res. 2026 (arXiv 2601.01319, abstract not read)

Sources

  1. Magnetic Milli-spinner for Robotic Endovascular Surgery · link [wu2024_millispinner]

Notes

Notes
Untethered: no bedside drive and no tip actuation, kept as a magnetic-navigation comparison point.

This record

Read level
full — full (abstract and results by grep)
Picture
Stanford Robotics Center demo page (src.stanford.edu/demo/magnetic-milli-spinner-for-robotic-endovascular-surgery-launch), image "magnetic millispinner (2).jpg" at 2500w (UR arm with rotating magnet over the head phantom) — licence: web image (Stanford Robotics Center), terms not stated; grade A
Picture changes
WebP converted to PNG; no crop, no edits; not upscaled
Found by
actuated-tip: Search of actuated-tip and magnetic devices.
Data
JSON · YAML: records/robots/millispinner_wu2024.yaml
Correction