 "schema_version": 1,
 "id": "millispinner_wu2024",
 "name": "Milli-spinner",
 "full_name": "Magnetic milli-spinner (untethered)",
 "group": "Zhao, Stanford Mechanical Engineering",
 "group_short": "Stanford",
 "family": "magnetic",
 "category": "magnetic-navigation",
 "first_year": 2024,
 "latest_year": 2026,
 "versions": "arXiv 2410.21112; optimisation follow-up Adv. Robotics Res. 2026 (arXiv 2601.01319, abstract not read)",
 "target_anatomy": "cerebral and general vasculature; phantoms, porcine blood aneurysm model",
 "devices": [
  "other"
 ],
 "devices_driven": "untethered 2.5 mm OD hollow spinner with helical fins and slits, in 3.5 mm ID tubes",
 "n_dof": null,
 "n_dof_detail": "rotating magnetic field from Helmholtz coils; six-axis robot arm holding a permanent magnet for navigation",
 "translation_mechanism": "magnetic",
 "translation_detail": "magnetic",
 "rotation_mechanism": "magnetic-field",
 "rotation_detail": "spinning in rotating field (up to 8.4k rpm tested)",
 "device_range": "untethered",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "none",
 "haptics_leader": "no",
 "imaging_compat": "fluoroscopy-guided in phantom",
 "sterility": "N.R.",
 "highest_validation": "phantom",
 "validation_detail": "vascular phantoms; in-situ embolisation of an aneurysm model filled with anticoagulated porcine blood; no animal",
 "key_numbers": "23 cm/s in tubular environment (abstract); stable spinning needs 15-20 mT; 56.4 cm/s predicted by CFD at 9.6k rpm",
 "sim_model": "CFD of swimming speed compared with experiment (wu2024_millispinner, full)",
 "has_sim_model": true,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "A",
  "kind": "photo",
  "source": "Stanford Robotics Center demo page (src.stanford.edu/demo/magnetic-milli-spinner-for-robotic-endovascular-surgery-launch), image \"magnetic millispinner (2).jpg\" at 2500w (UR arm with rotating magnet over the head phantom)",
  "source_url": null,
  "licence": "web image (Stanford Robotics Center), terms not stated",
  "credit": "Stanford Robotics Center demo page (src.stanford.edu/demo/magnetic-milli-spinner-for-robotic-endovascular-surgery-launch), image \"magnetic millispinner (2).jpg\" at 2500w (UR arm with rotating magnet over the head phantom). Licence: web image (Stanford Robotics Center), terms not stated.",
  "edits": "WebP converted to PNG; no crop, no edits; not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "millispinner_wu2024_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Robot arm with rotating magnet beside the head phantom on a light table",
   "source_key": "wu2024_millispinner",
   "source": "Wu et al. (2024), Stanford Robotics Center demo page image 'magnetic millispinner (1).jpg', src.stanford.edu/demo/magnetic-milli-spinner-for-robotic-endovascular-surgery-launch",
   "source_url": "https://arxiv.org/abs/2410.21112",
   "where": "Stanford Robotics Center demo page image 'magnetic millispinner (1).jpg', src.stanford.edu/demo/magnetic-milli-spinner-for-robotic-endovascular-surgery-launch",
   "licence": "press image (Stanford Robotics Center demo page)"
  },
  {
   "id": "millispinner_wu2024_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Rotating magnet close to the transparent head phantom with cerebral arteries",
   "source_key": "wu2024_millispinner",
   "source": "Wu et al. (2024), Stanford Robotics Center demo page image 'Magnetic Milli-spinner.png'",
   "source_url": "https://arxiv.org/abs/2410.21112",
   "where": "Stanford Robotics Center demo page image 'Magnetic Milli-spinner.png'",
   "licence": "press image (Stanford Robotics Center demo page)"
  },
  {
   "id": "millispinner_wu2024_03",
   "status": "cleared",
   "file": "pictures/extra/millispinner_wu2024_03.jpg",
   "thumb": "pictures/thumbs/millispinner_wu2024_03.jpg",
   "kind": "in-use",
   "caption": "Rotating magnet on a robot arm driving the spinner through a head-model flow phantom",
   "source_key": "wu2025_millispinner",
   "source": "Wu et al., Advanced Materials (2025), Fig. 2(h-j), p. 3",
   "source_url": "https://doi.org/10.1002/adma.202508180",
   "where": "Fig. 2(h-j), p. 3",
   "licence": "CC BY (Creative Commons Attribution License)"
  },
  {
   "id": "millispinner_wu2024_04",
   "status": "cleared",
   "file": "pictures/extra/millispinner_wu2024_04.jpg",
   "thumb": "pictures/thumbs/millispinner_wu2024_04.jpg",
   "kind": "in-use",
   "caption": "Fluoroscopy-guided navigation setup in an angiography suite with the flow model",
   "source_key": "wu2025_millispinner",
   "source": "Wu et al., Advanced Materials (2025), Fig. 2(e-f), p. 3",
   "source_url": "https://doi.org/10.1002/adma.202508180",
   "where": "Fig. 2(e-f), p. 3",
   "licence": "CC BY (Creative Commons Attribution License)"
  },
  {
   "id": "millispinner_wu2024_05",
   "status": "cleared",
   "file": "pictures/extra/millispinner_wu2024_05.jpg",
   "thumb": "pictures/thumbs/millispinner_wu2024_05.jpg",
   "kind": "close-up",
   "caption": "Optimized milli-spinner design drawing with photos of the printed spinner",
   "source_key": "lu2026_millispinner",
   "source": "Lu et al., Advanced Robotics Research (2026), Fig. 8(a), p. 9",
   "source_url": "https://doi.org/10.1002/adrr.70121",
   "where": "Fig. 8(a), p. 9",
   "licence": "CC BY (Creative Commons Attribution License)"
  }
 ],
 "sources": [
  {
   "key": "wu2024_millispinner",
   "doi": null,
   "url": "https://arxiv.org/abs/2410.21112",
   "title": "Magnetic Milli-spinner for Robotic Endovascular Surgery",
   "year": 2024,
   "venue": null
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (abstract and results by grep)",
 "notes": "Untethered: no bedside drive and no tip actuation, kept as a magnetic-navigation comparison point.",
 "maintainers_own": false,
 "found_by": "actuated-tip"
}