Catheter Robot Atlas

Academic bedside robots · polimi_zhang2022

MitraClip stabiliser

PoliMi motorised stabiliser for the MitraClip delivery catheter with IRL path planning
Politecnico di Milano, DEIB NEARLab (E. De Momi; X. Zhang, A. Segato) with E. Votta (cardiovascular biomechanics); Horizon 2020 funding · 2022
MitraClip stabiliser: main picture1 / 2
Not cleared for distribution. zhang2022_structural_IROS Fig. 4(A) catheter actuation plant (inside the red frame). Licence: publisher copyright (IEEE). Picture grade C.
Family
Academic bedside robots (academic)
Devices
Abbott MitraClip delivery catheter, 16 Fr (5.28 mm), 109.5 cm, three pull wires 90 deg apart, inside its sheath (Sec. III-C)
Degrees of freedom
3 (medio-lateral bending, antero-posterior bending, insertion in the sheath)
Push
Linear stage
Twist
None
Tested up to
Phantom
Sim-to-real
Yes — A controller or perception model trained in simulation ran on this robot.

Testing

Highest level
Phantom
What was tested
patient-specific silicone phantom of femoral vein and IVC with rigid fossa ovalis and mitral replicas registered to Unity (Sec. IV-A)
Key numbers
closed-loop tip error 1.12 +- 0.75, 1.09 +- 0.68, 1.66 +- 0.62 mm (X/Y/Z), maximum 2.64 mm; PID reduces mean error by 32.68% vs open loop; planned path 60.63 +- 0.96 mm (BC+GAIL) vs 64.99 +- 6.11 mm (manual) (Sec. V)

Mechanism

Push (translation)
linear-stage+clamp (linear actuator on the motorised stabiliser advances the delivery catheter)
Twist (rotation)
none; two bending DOF actuated through the handle
Device range
one clinical delivery system
Target anatomy
structural heart: transseptal mitral edge-to-edge repair

Sensing and interface

Force sensing
none
Leader and haptics
none (autonomous execution of planned path, PID on EM feedback)
Imaging
NDI Aurora EM: 5-DOF tip sensor (0.7 +- 1.4 mm) for feedback, 6-DOF probe (0.48 +- 0.88 mm) for recording
Sterility
N.R.

Size

Footprint (mm)
N.R.
Mass (kg)
N.R. (3-D-printed PLA)

Simulation

Simulation model
yes: Unity scene of the patient anatomy with a Cosserat-rod-based catheter kinematic model (zhang2022_robotic Sec. III)
Sim-trained controller
yes (phantom): a BC+GAIL inverse-RL path planner trained in the Unity environment generated the path that the robot executed in the physical phantom with EM closed loop (zhang2022_robotic Sec. IV, Fig. 7)
Studies on this robot

Versions

Lineage
2022 IROS (only)

Sources

  1. Robotic Actuation and Control of a Catheter for Structural Intervention Cardiology · 2022 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS), 2022 · doi:10.1109/iros47612.2022.9981676 [zhang2022_robotic]

Notes

Notes
Fills the Politecnico di Milano gap of A3. Only robot in A9 with a simulation-trained planner run on hardware.

This record

Read level
full
Picture
zhang2022_structural_IROS Fig. 4(A) catheter actuation plant (inside the red frame) ↗ — licence: publisher copyright (IEEE); grade C
Picture changes
native extract (961x568, the whole composite is one raster); cropped to panel A interior; labels and dashed callout lines are baked in and left; not upscaled
Found by
gap: Gap searches for robots the other searches missed.
Data
JSON · YAML: records/robots/polimi_zhang2022.yaml
Correction
Suggest a correction [repository not created yet]