Catheter Robot Atlas

Academic bedside robots · kcl_cardioxplorer_xu2024

CardioXplorer

CardioXplorer open-source modular teleoperated robotic catheter-ablation system
King's College London, School of Biomedical Engineering & Imaging Sciences (K. S. Rhode, R. J. Housden; Z. Xu, A. M. Zeidan) with Guy's and St Thomas' and St George's hospitals · 2024
CardioXplorer: main picture1 / 8
Xu et al., Robotics (2024), Fig. 4. CC BY 4.0. Changes: cropped from the 3155x1185 MDPI full-size figure (caption text below the photo excluded); no other edits; not upscaled. Source ↗ Picture grade A.
Family
Academic bedside robots (academic)
Devices
commercial ablation catheters through a catheter-agnostic handle gripper: Thermocool SmartTouch SF and CS, Webster CS EZ Steer (Biosense Webster), Blazer II XP (Boston Scientific) (Sec. 2.1)
Degrees of freedom
3 (insertion/retraction, handle rotation, knob)
Push
Linear stage
Twist
Handle or knob
Tested up to
Phantom
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Phantom
What was tested
benchtop targeting of left pulmonary veins in an AF-patient cardiac phantom; 3 cardiologists and 2 trainees, 5 trials, manual vs robot (Sec. 3)
Key numbers
task duration -1.59 s with the robot; robotic motion 35.2% smoother for experts than intermediates; jerk improvement vs manual 42.2% (experts) and 13.6% (intermediates); speeds up to +-25 mm/s, 0.5 deg rotation steps up to 9.375 rps; parts cost about USD 700 (abstract, Sec. 2)

Mechanism

Push (translation)
linear-stage+clamp (linear module: NEMA-17 stepper with GT2 belt drive carries the handle, 220 mm insertion depth)
Twist (rotation)
NEMA-8 stepper (19:1) rotates the handle without limit through a slip ring; servo (12 kg.cm holding torque) turns the knob +-60 deg
Device range
four commercial ablation catheter models (open-source CAD handles)
Target anatomy
atrial-fibrillation catheter ablation (pulmonary vein isolation)

Sensing and interface

Force sensing
none on the robot (knob torque of about 6 kg.cm measured during design)
Leader and haptics
joystick, no haptics
Imaging
NDI Aurora EM tracking in tests; phantom imaged on a Siemens Artis Q biplane
Sterility
detachable disposable single-use catheter; N.R. further

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
2024 MDPI Robotics (only); sibling KCL robot for transseptal puncture (kcl_tsp_zeidan2023)

Sources

  1. CardioXplorer: An Open-Source Modular Teleoperative Robotic Catheter Ablation System · Robotics, 2024 · doi:10.3390/robotics13050080 [xu2024_cardioxplorer]

Notes

Notes
Second KCL robot family next to the A3 neuro manipulator (kclneuro_sadati2025); open-source designs. CC BY 4.0.

This record

Read level
full
Picture
xu2024_cardioxplorer_Robotics Fig. 4 left photo (SmartTouch SF catheter in the catheter-agnostic gripper); full-size image https://mdpi-res.com/robotics/robotics-13-00080/article_deploy/html/images/robotics-13-00080-g004.png ↗ — licence: CC BY 4.0; grade A
Picture changes
cropped from the 3155x1185 MDPI full-size figure (caption text below the photo excluded); no other edits; not upscaled
Found by
gap: Gap searches for robots the other searches missed.
Data
JSON · YAML: records/robots/kcl_cardioxplorer_xu2024.yaml
Correction