Catheter Robot Atlas

Magnetic navigation · cardiomag_fischer2022

CardioMag

CardioMag eMNS with magnetic catheter and magnetic localisation (ETH)
Nelson, ETH Zurich (Fischer, Boehler) · 2022
CardioMag: main picture1 / 3
Not cleared for distribution. fischer2022_using Fig. 3(a) (setup inside CardioMag: advancer, catheter, phantom between the coils). Licence: publisher copyright (IEEE). Picture grade A.
Family
Magnetic navigation (magnetic-navigation)
Devices
own catheter: OD 2.8 mm / ID 2 mm Flexan tube with 3 mm OD x 6 mm permanent magnet at tip, Hall sensor 12 mm behind magnet
Degrees of freedom
field direction and gradient from 8 electromagnets; advance by Maxon-motor catheter advancer at 10 mm/s
Push
Magnetic
Twist
Field direction
Tested up to
Phantom
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Phantom
What was tested
2D vascular models a and b; catheter navigated through all branches; localisation tested over 40 trials on path 1-3 of model a
Key numbers
eMNS fields 20 mT, gradient 0.15 mT/mm; mean localisation error 3.64 mm (sd 1.61) model b, 2.74 mm with extra field change; model a mean 7.94 mm (sd 2.94), 5.65 mm with extra gradient; position estimate 38.3 ms

Mechanism

Push (translation)
magnetic
Twist (rotation)
field direction
Device range
N.R.
Target anatomy
endoluminal, 2D vascular models (24 mm and 11 mm channels)

Sensing and interface

Force sensing
Hall-effect sensor for localisation (range +/- 36 mT per axis), no force sensing
Leader and haptics
no
Imaging
localisation from magnetic field; fluoroscopy minimised; overhead camera as ground truth
Sterility
N.R.

Size

Footprint (mm)
eMNS about 0.3 m scale bar in Fig. 1 (size N.R. in text)
Mass (kg)
N.R.

Simulation

Simulation model
linear model of the eMNS (coil currents to field), calibrated; no physics simulator of the catheter
Sim-trained controller
none found

Versions

Lineage
single (RA-L 2022); same eMNS platform as Navion's academic origin; 2017-19: ETH (Nelson) magnetic catheters in Aeon Phocus: tethered magnet catheter with force and 5-DOF pose control (Chautems 2017 ICRA) and variable-stiffness 2- and 3-segment catheters (Chautems 2017 IROS; 2019 Adv. Intell. Syst.) (Hwang 2020 Figs. 8a, 8g) (hwang2020_review)

Sources

  1. Using Magnetic Fields to Navigate and Simultaneously Localize Catheters in Endoluminal Environments · IEEE Robotics and Automation Letters, 2022 · doi:10.1109/lra.2022.3181420 [fischer2022_using]
  2. A review of magnetic actuation systems and magnetically actuated guidewire- and catheter-based microrobots for vascular interventions · Intelligent Service Robotics, 2020 · doi:10.1007/s11370-020-00311-0 [hwang2020_review]

Notes

Notes
Centre-line automation plus localisation from field only. Commercial Navion is A4's.

This record

Read level
full
Picture
fischer2022_using Fig. 3(a) (setup inside CardioMag: advancer, catheter, phantom between the coils) ↗ — licence: publisher copyright (IEEE); grade A
Picture changes
native extract, crop left panel only
Found by
actuated-tip: Search of actuated-tip and magnetic devices.
Data
JSON · YAML: records/robots/cardiomag_fischer2022.yaml
Correction