 "schema_version": 1,
 "id": "kcl_cardioxplorer_xu2024",
 "name": "CardioXplorer",
 "full_name": "CardioXplorer open-source modular teleoperated robotic catheter-ablation system",
 "group": "King's College London, School of Biomedical Engineering & Imaging Sciences (K. S. Rhode, R. J. Housden; Z. Xu, A. M. Zeidan) with Guy's and St Thomas' and St George's hospitals",
 "group_short": "KCL",
 "family": "academic",
 "category": "academic",
 "first_year": 2024,
 "latest_year": 2024,
 "versions": "2024 MDPI Robotics (only); sibling KCL robot for transseptal puncture (kcl_tsp_zeidan2023)",
 "target_anatomy": "atrial-fibrillation catheter ablation (pulmonary vein isolation)",
 "devices": [
  "catheter"
 ],
 "devices_driven": "commercial ablation catheters through a catheter-agnostic handle gripper: Thermocool SmartTouch SF and CS, Webster CS EZ Steer (Biosense Webster), Blazer II XP (Boston Scientific) (Sec. 2.1)",
 "n_dof": 3,
 "n_dof_detail": "3 (insertion/retraction, handle rotation, knob)",
 "translation_mechanism": "linear-stage+clamp",
 "translation_detail": "linear-stage+clamp (linear module: NEMA-17 stepper with GT2 belt drive carries the handle, 220 mm insertion depth)",
 "rotation_mechanism": "handle-rotation",
 "rotation_detail": "NEMA-8 stepper (19:1) rotates the handle without limit through a slip ring; servo (12 kg.cm holding torque) turns the knob +-60 deg",
 "device_range": "four commercial ablation catheter models (open-source CAD handles)",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "none on the robot (knob torque of about 6 kg.cm measured during design)",
 "haptics_leader": "joystick, no haptics",
 "imaging_compat": "NDI Aurora EM tracking in tests; phantom imaged on a Siemens Artis Q biplane",
 "sterility": "detachable disposable single-use catheter; N.R. further",
 "highest_validation": "phantom",
 "validation_detail": "benchtop targeting of left pulmonary veins in an AF-patient cardiac phantom; 3 cardiologists and 2 trainees, 5 trials, manual vs robot (Sec. 3)",
 "key_numbers": "task duration -1.59 s with the robot; robotic motion 35.2% smoother for experts than intermediates; jerk improvement vs manual 42.2% (experts) and 13.6% (intermediates); speeds up to +-25 mm/s, 0.5 deg rotation steps up to 9.375 rps; parts cost about USD 700 (abstract, Sec. 2)",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "cleared",
  "grade": "A",
  "kind": "photo",
  "source": "xu2024_cardioxplorer_Robotics Fig. 4 left photo (SmartTouch SF catheter in the catheter-agnostic gripper); full-size image https://mdpi-res.com/robotics/robotics-13-00080/article_deploy/html/images/robotics-13-00080-g004.png",
  "source_url": "https://doi.org/10.3390/robotics13050080",
  "licence": "CC BY 4.0",
  "credit": "Xu et al., Robotics (2024), Fig. 4. CC BY 4.0.",
  "edits": "cropped from the 3155x1185 MDPI full-size figure (caption text below the photo excluded); no other edits; not upscaled",
  "file": "pictures/kcl_cardioxplorer_xu2024.jpg"
 },
 "pictures": [
  {
   "id": "kcl_cardioxplorer_xu2024_01",
   "status": "cleared",
   "file": "pictures/extra/kcl_cardioxplorer_xu2024_01.jpg",
   "thumb": "pictures/thumbs/kcl_cardioxplorer_xu2024_01.jpg",
   "kind": "in-use",
   "caption": "Robot driving an ablation catheter into the AF cardiac phantom in the cath lab",
   "source_key": "xu2024_cardioxplorer",
   "source": "Xu et al., Robotics (2024), Fig. 12(a), robot trial, p. 10",
   "source_url": "https://doi.org/10.3390/robotics13050080",
   "where": "Fig. 12(a), robot trial, p. 10",
   "licence": "CC BY 4.0"
  },
  {
   "id": "kcl_cardioxplorer_xu2024_02",
   "status": "cleared",
   "file": "pictures/extra/kcl_cardioxplorer_xu2024_02.jpg",
   "thumb": "pictures/thumbs/kcl_cardioxplorer_xu2024_02.jpg",
   "kind": "photo",
   "caption": "Manufactured robot from above: rotation module on the linear module, with dimensions",
   "source_key": "xu2024_cardioxplorer",
   "source": "Xu et al., Robotics (2024), Fig. 2(a), lower photo, p. 4",
   "source_url": "https://doi.org/10.3390/robotics13050080",
   "where": "Fig. 2(a), lower photo, p. 4",
   "licence": "CC BY 4.0"
  },
  {
   "id": "kcl_cardioxplorer_xu2024_03",
   "status": "cleared",
   "file": "pictures/extra/kcl_cardioxplorer_xu2024_03.jpg",
   "thumb": "pictures/thumbs/kcl_cardioxplorer_xu2024_03.jpg",
   "kind": "photo",
   "caption": "Side view of the manufactured robot holding a catheter handle, with dimensions",
   "source_key": "xu2024_cardioxplorer",
   "source": "Xu et al., Robotics (2024), Fig. 2(a), upper photo, p. 4",
   "source_url": "https://doi.org/10.3390/robotics13050080",
   "where": "Fig. 2(a), upper photo, p. 4",
   "licence": "CC BY 4.0"
  },
  {
   "id": "kcl_cardioxplorer_xu2024_04",
   "status": "cleared",
   "file": "pictures/extra/kcl_cardioxplorer_xu2024_04.jpg",
   "thumb": "pictures/thumbs/kcl_cardioxplorer_xu2024_04.jpg",
   "kind": "render",
   "caption": "CAD rendering of the linear, rotation and catheter-mount modules",
   "source_key": "xu2024_cardioxplorer",
   "source": "Xu et al., Robotics (2024), Fig. 2(b), p. 4",
   "source_url": "https://doi.org/10.3390/robotics13050080",
   "where": "Fig. 2(b), p. 4",
   "licence": "CC BY 4.0"
  },
  {
   "id": "kcl_cardioxplorer_xu2024_05",
   "status": "cleared",
   "file": "pictures/extra/kcl_cardioxplorer_xu2024_05.jpg",
   "thumb": "pictures/thumbs/kcl_cardioxplorer_xu2024_05.jpg",
   "kind": "close-up",
   "caption": "CAD of the adaptive silicone gripper around a catheter knob",
   "source_key": "xu2024_cardioxplorer",
   "source": "Xu et al., Robotics (2024), Fig. 3(b), p. 4",
   "source_url": "https://doi.org/10.3390/robotics13050080",
   "where": "Fig. 3(b), p. 4",
   "licence": "CC BY 4.0"
  },
  {
   "id": "kcl_cardioxplorer_xu2024_06",
   "status": "cleared",
   "file": "pictures/extra/kcl_cardioxplorer_xu2024_06.jpg",
   "thumb": "pictures/thumbs/kcl_cardioxplorer_xu2024_06.jpg",
   "kind": "leader",
   "caption": "Gamepad controller in the control room and responder robot at the catheterisation suite",
   "source_key": "xu2024_cardioxplorer",
   "source": "Xu et al., Robotics (2024), Fig. 8, p. 7",
   "source_url": "https://doi.org/10.3390/robotics13050080",
   "where": "Fig. 8, p. 7",
   "licence": "CC BY 4.0"
  },
  {
   "id": "kcl_cardioxplorer_xu2024_07",
   "status": "cleared",
   "file": "pictures/extra/kcl_cardioxplorer_xu2024_07.jpg",
   "thumb": "pictures/thumbs/kcl_cardioxplorer_xu2024_07.jpg",
   "kind": "in-use",
   "caption": "Three actuation configurations with the resulting catheter tip articulation",
   "source_key": "xu2024_cardioxplorer",
   "source": "Xu et al., Robotics (2024), Fig. 9, p. 7",
   "source_url": "https://doi.org/10.3390/robotics13050080",
   "where": "Fig. 9, p. 7",
   "licence": "CC BY 4.0"
  }
 ],
 "sources": [
  {
   "key": "xu2024_cardioxplorer",
   "doi": "10.3390/robotics13050080",
   "url": null,
   "title": "CardioXplorer: An Open-Source Modular Teleoperative Robotic Catheter Ablation System",
   "year": 2024,
   "venue": "Robotics"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full",
 "notes": "Second KCL robot family next to the A3 neuro manipulator (kclneuro_sadati2025); open-source designs. CC BY 4.0.",
 "maintainers_own": false,
 "found_by": "gap"
}