 "schema_version": 1,
 "id": "gatech_tmvr_qi2023",
 "name": "TMVR delivery",
 "full_name": "Georgia Tech telerobotic transcatheter delivery system for a mitral valve implant",
 "group": "Georgia Institute of Technology, RoboMed Laboratory (J. P. Desai; R. Qi, N. U. Nayar)",
 "group_short": "Georgia Tech",
 "family": "tip",
 "category": "actuated-device",
 "first_year": 2020,
 "latest_year": 2023,
 "versions": "2020 IROS first robotic transcatheter delivery system (title); ICRA 2021, T-MRB 2022 and ICRA 2023 papers of Nayar, Qi and Desai (qi2023 refs. [18]-[20], not read); 2023 RAL telerobotic system: reinforced guide tube in a 24 Fr commercial steerable sheath, stronger actuation, waterproofing, stabiliser, wireless joystick (qi2023, full)",
 "target_anatomy": "mitral valve (transseptal implant delivery)",
 "devices": [
  "catheter",
  "sheath"
 ],
 "devices_driven": "own robotic catheter (5.7 mm OD) on a 1.11 m reinforced guide tube inside a customised steerable outer sheath (VERGO, Duke Extrusion; 6.6 mm ID, 8.3 mm OD, 990.6 mm) (qi2023 Abstract, Sec. IV-A)",
 "n_dof": 4,
 "n_dof_detail": "4 (proximal bending +-90 deg, distal bending +-120 deg, prismatic 70 mm actuatable and restricted to 60 mm, distal torsion +-75 deg; qi2023 Sec. II)",
 "translation_mechanism": "tendon/tip-actuation",
 "translation_detail": "tendon/tip-actuation (distal prismatic joint: the quadlumen tube is advanced by a Pololu 100:1 gearmotor through a lead screw on a linear guide; tendon-driven bending and torsion joints; actuation unit on a stabiliser with a linear guide rail; qi2023 Sec. II)",
 "rotation_mechanism": "tendon-bending",
 "rotation_detail": "tendon-driven distal torsion joint: two antagonistic tendons on a lead-screw holder driven by a DC motor with a 16:1 gearbox (qi2023 Sec. II)",
 "device_range": "own catheter",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "none reported",
 "haptics_leader": "wireless Xbox 360 controller, no haptics",
 "imaging_compat": "N.R.",
 "sterility": "N.R.; waterproofed with two three-way stopcocks for saline introduction (qi2023 Sec. II)",
 "highest_validation": "phantom",
 "validation_detail": "implant delivery in a realistic human cardiovascular phantom (United Biologics) with a ViVitro pulsatile pump (63.1 ml/stroke, 70 beats/min), one demonstration; tube validation with the implant free and against the leaflet; no animal study yet (qi2023 Sec. IV-A, IV-B, V)",
 "key_numbers": "joint tracking RMSE 0.20, 0.74, 0.02 and 0.17 mm (qi2023 Sec. IV-A); Maxon 0.5 W 8 mm DC motors (64:1 bending, 16:1 torsion) and Pololu 100:1 prismatic gearmotor, tendon force up to about 20 N (Sec. II); implant advanced 40 mm under the leaflet (Sec. IV-A); tip deflection about 2 mm during a 15 s hold (Sec. IV-B)",
 "sim_model": "analytical static deflection model of the reinforced quadlumen tube (eqs. 1-10), checked against ANSYS 18.2 FEM and a camera bench test in 12 configurations (qi2023 Sec. III-C, Table I); not a model of the whole robot",
 "has_sim_model": true,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "B",
  "kind": "render",
  "source": "qi2023_telerobotic Fig. 2 (system CAD: actuation system on the stabiliser with the guide tube), RA-L 2023, doi:10.1109/LRA.2023.3265592; native embedded raster p. 2 (3166 x 1136), which carries none of the vector labels",
  "source_url": "https://doi.org/10.1109/LRA.2023.3265592",
  "licence": "publisher copyright (IEEE)",
  "credit": "Qi et al., IEEE Robotics and Automation Letters (2023), Fig. 2. publisher copyright (IEEE).",
  "edits": "native extract; cropped to the actuation system and stabiliser (guide tube cut at the right edge so the detail insets stay out); 40 px of white added at the top as margin (render background is pure white); no other edits; not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "gatech_tmvr_qi2023_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Telerobotic delivery system in a human cardiovascular phantom, side view",
   "source_key": "qi2023_telerobotic",
   "source": "Qi et al., IEEE Robotics and Automation Letters (2023), Fig. 1, p. 1",
   "source_url": "https://doi.org/10.1109/lra.2023.3265592",
   "where": "Fig. 1, p. 1",
   "licence": "publisher copyright"
  },
  {
   "id": "gatech_tmvr_qi2023_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Implant delivery setup: robot, pulsatile pump, phantom and gamepad",
   "source_key": "qi2023_telerobotic",
   "source": "Qi et al., IEEE Robotics and Automation Letters (2023), Fig. 7(a), p. 6",
   "source_url": "https://doi.org/10.1109/lra.2023.3265592",
   "where": "Fig. 7(a), p. 6",
   "licence": "publisher copyright"
  },
  {
   "id": "gatech_tmvr_qi2023_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Robotic catheter with bending section beside a one-cent coin",
   "source_key": "qi2023_telerobotic",
   "source": "Qi et al., IEEE Robotics and Automation Letters (2023), Fig. 7(b), p. 6",
   "source_url": "https://doi.org/10.1109/lra.2023.3265592",
   "where": "Fig. 7(b), p. 6",
   "licence": "publisher copyright"
  },
  {
   "id": "gatech_tmvr_qi2023_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Guide tube and implant inside the heart phantom, two views",
   "source_key": "qi2023_telerobotic",
   "source": "Qi et al., IEEE Robotics and Automation Letters (2023), Fig. 7(c-d), p. 6",
   "source_url": "https://doi.org/10.1109/lra.2023.3265592",
   "where": "Fig. 7(c-d), p. 6",
   "licence": "publisher copyright"
  },
  {
   "id": "gatech_tmvr_qi2023_05",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Earlier steerable catheter delivering the mitral clip inside a phantom heart",
   "source_key": "nayar2020_towards",
   "source": "Nayar et al., 2020 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (2020), Fig. 9(a), p. 6 (IROS 2020 predecessor)",
   "source_url": "https://doi.org/10.1109/iros45743.2020.9341551",
   "where": "Fig. 9(a), p. 6 (IROS 2020 predecessor)",
   "licence": "publisher copyright"
  },
  {
   "id": "gatech_tmvr_qi2023_06",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Tendon actuation stages on a perforated plate with the catheter, earlier prototype",
   "source_key": "nayar2020_towards",
   "source": "Nayar et al., 2020 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (2020), Fig. 7, p. 4",
   "source_url": "https://doi.org/10.1109/iros45743.2020.9341551",
   "where": "Fig. 7, p. 4",
   "licence": "publisher copyright"
  },
  {
   "id": "gatech_tmvr_qi2023_07",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Assembled catheter with bending and torsion joints and implant module",
   "source_key": "nayar2020_towards",
   "source": "Nayar et al., 2020 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (2020), Fig. 6(a), p. 4",
   "source_url": "https://doi.org/10.1109/iros45743.2020.9341551",
   "where": "Fig. 6(a), p. 4",
   "licence": "publisher copyright"
  },
  {
   "id": "gatech_tmvr_qi2023_08",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Renderings of the implant delivery module with clip closed and opened",
   "source_key": "nayar2020_towards",
   "source": "Nayar et al., 2020 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (2020), Fig. 5, p. 3",
   "source_url": "https://doi.org/10.1109/iros45743.2020.9341551",
   "where": "Fig. 5, p. 3",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "qi2023_telerobotic",
   "doi": "10.1109/lra.2023.3265592",
   "url": null,
   "title": "Telerobotic Transcatheter Delivery System for Mitral Valve Implant",
   "year": 2023,
   "venue": "IEEE Robotics and Automation Letters"
  },
  {
   "key": "nayar2020_towards",
   "doi": "10.1109/iros45743.2020.9341551",
   "url": null,
   "title": "Towards the Development of a Robotic Transcatheter Delivery System for Mitral Valve Implant",
   "year": 2020,
   "venue": "2020 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS)"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (IEEE PDF, 2026-10-14)",
 "notes": "Tip-actuated device, so category actuated-device; Desai's COAST and balloon guidewires are in robots_A5. F4 2026-10-14: no clean photo of the whole robot in the paper (Figs. 1, 7(a), 8 show the unit small or with pointers).",
 "maintainers_own": false,
 "found_by": "gap"
}