Catheter Robot Atlas

Commercial and clinical-stage systems · caranx_smits2025

Caranx TAVI robot

Caranx Medical robotic instrument manipulator for transcatheter aortic valve implantation
Caranx Medical (P. Berthet-Rayne, J. Smits, P. Schegg) with Institut Arnault Tzanck, Clinique Pasteur, Emory University, University Hospital Nice · 2023–2025
Caranx TAVI robot: main picture1 / 3
Smits et al., Frontiers in Robotics and AI (2025), Fig. 2. CC BY 4.0. Changes: the PDF splits the photo into 3 raster tiles (705x430, 705x288, 284x715): stitched at native size (no vector annotations included), cropped to the manipulator. Source ↗ Picture grade B.
Family
Commercial and clinical-stage systems (commercial)
Devices
Edwards SAPIEN 3 delivery system (flex catheter and handle), intraventricular guidewire, balloon inflation device (Results, Fig. 2)
Degrees of freedom
N.R. as a number; four drivers: catheter driver, handle driver (flex and balloon fine-positioning wheels), guidewire driver, inflation driver
Push
Friction rollers
Twist
Handle or knob
Tested up to
Animal
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Animal
What was tested
3 in-vivo porcine experiments with 20 navigation and balloon-inflation sequences (no implant) and one robot-assisted TAV implantation in a pig; phantom operator study and level-3 supervised autonomy with camera and fluoroscopy feedback (Results)
Key numbers
fluoroscopy closed-loop positioning 30 runs without failure, final error 0.5 mm median, min-max range 1.03 mm; calibration about 5 s; autonomy up to 2x accuracy and 4x precision vs manual on phantom (Results)

Mechanism

Push (translation)
friction-roller (friction-wheel transmissions for delivery catheter and guidewire, pretensioned by one screw)
Twist (rotation)
handle wheels through geared transmissions; catheter roll N.R.
Device range
one delivery-system type per interface (a reusable base with per-device interfaces envisioned)
Target anatomy
transfemoral TAVI (aortic valve)

Sensing and interface

Force sensing
in-line pressure sensor for pressure-monitored balloon inflation; instrument force sensing N.R.
Leader and haptics
single wireless handheld user-input device; teleoperation from a shielded room
Imaging
fluoroscopy (wireless pedal) and camera test bench; neural-network feature tracking for closed-loop positioning
Sterility
plastic transmission covers on handle wheels; quick-release thumbscrews and spring-loaded sliders; N.R. further

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found (network trained on real fluoroscopy data)

Versions

Lineage
Oct 2023 hybrid-room phantom tests; Dec 2023 - Feb 2024 three in-vivo porcine experiments (CERIMED Marseille); 15 Mar 2024 robot-assisted TAV implantation in a pig; 2025 FRAI paper

Sources

  1. Towards autonomous robot-assisted transcatheter heart valve implantation: in vivo teleoperation and phantom validation of AI-guided positioning · Frontiers in Robotics and AI, 2025 · doi:10.3389/frobt.2025.1650228 [smits2025_towards]

Notes

Notes
Pre-market company prototype (not in robots_A4). CC BY 4.0 paper.

This record

Read level
full
Picture
smits2025_tavi_FRAI Fig. 2 (left panel, in-vivo set-up 15 Mar 2024), robotic instrument manipulator region ↗ — licence: CC BY 4.0; grade B
Picture changes
the PDF splits the photo into 3 raster tiles (705x430, 705x288, 284x715): stitched at native size (no vector annotations included), cropped to the manipulator