Catheter Robot Atlas

Academic bedside robots · siemens_ice_kim2020

ICE manipulator

Siemens Healthineers robotic ICE catheter manipulator (proximal and distal units)
Siemens Healthineers Digital Technology & Innovation (Y.-H. Kim, A. Kapoor, T. Mansi) with Houston Methodist (C. H. Lin, Z. Li) and Texas A&M (R. Langari, zero-slack design) · 2020–2022
ICE manipulator: main picture1 / 4
Not cleared for distribution. kim2022_ice_IJCARS Fig. 1(b) robotic ICE manipulator (distal and proximal units) on the pig. Licence: publisher copyright (Springer, (c) CARS 2022). Picture grade C.
Family
Academic bedside robots (academic)
Devices
ACUSON AcuNav Volume ICE catheters (Siemens Healthineers)
Degrees of freedom
4 (two knob angles, bulk rotation, translation)
Push
Other drives
Twist
Handle or knob
Tested up to
Animal
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Animal
What was tested
4 animal experiments (pig, Houston Methodist MITIE) including 2 fluoro-less robot-guided transseptal punctures; beating-heart phantom (Archetype Medical) (kim2022 Sec. 3)
Key numbers
view recovery 2.09 +- 0.90 mm and 3.93 +- 2.07 deg in animals, 0.67 +- 0.79 mm and 0.37 +- 0.19 deg in phantom (kim2022 abstract); bench compensation at 60 deg bend 14.8 mm / 31.0 deg -> 5.0 mm / 10.3 deg RMSE (kim2020 arXiv); zero-slack bench, 3 sinusoid trials: tip MAE 4.80 mm / 6.11 deg, 3.31 mm / 3.26 deg with +/-10 deg hysteresis compensation (debuy2022 Table 1)

Mechanism

Push (translation)
other: proximal unit at the introducer sheath gives linear and rotational motion of the shaft (mechanism N.R.)
Twist (rotation)
distal unit holds the handle, turns both knobs and gives bulk rotation synchronised with the proximal unit
Device range
one ICE catheter family
Target anatomy
intracardiac echocardiography during EP and structural-heart procedures (incl. transseptal puncture)

Sensing and interface

Force sensing
none
Leader and haptics
6-DOF analog joystick, no force feedback
Imaging
ICE ultrasound; fluoroscopy-free robot-guided transseptal punctures in animals
Sterility
zero-slack version: reusable actuator back portion, disposable front portion that holds the catheter tendons (debuy2022 Introduction, Fig. 1); 2020/2022 manipulator N.R.

Size

Footprint (mm)
zero-slack version: 298 mm long, 66 mm diameter (debuy2022 Fig. 1(a) dimension labels); 2020/2022 manipulator N.R.
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
2020 arXiv (kim2020_towards: 'Back' unit on the handle turns both knobs and gives bulk rotation, 'Front' unit at the introducer gives translation and rotation; 3 pig experiments) / 2022 IJCARS 4-DOF manipulator with automated view recovery (distal and proximal units; 4 pig experiments); 2022 arXiv zero-slack separable ICE robot (debuy2022_design: reusable back portion with four Faulhaber LM 1483-080-11-C linear motors pulling the four tendons directly through 3:1 spools, disposable front portion clipped on with a twist lock and magnetic coupling; 2 bending DOF only, translation/rotation not handled; Materials and Methods, Fig. 1); 2025 arXiv tip-visibility guidance (title)

Sources

  1. Automated catheter tip repositioning for intra-cardiac echocardiography · International Journal of Computer Assisted Radiology and Surgery, 2022 · doi:10.1007/s11548-022-02631-1 [kim2022_automated]
  2. Towards Automatic Manipulation of Intra-cardiac Echocardiography Catheter · arXiv, 2020 · doi:10.48550/ARXIV.2009.05859 [kim2020_towards]
  3. Design and validation of zero-slack separable manipulator for Intracardiac Echocardiography · arXiv, 2022 · doi:10.48550/ARXIV.2204.00645 [debuy2022_design]

Notes

Notes
Industrial research prototype, not a product; category academic for that reason. Covers the 'Houston' gap through Houston Methodist. | 2026-11-08 X3: zero-slack paper read in full (it is a 2-page abstract); kim2020 PDF confirms 3 animal experiments in the 2020 version, the row keeps kim2022's 4

This record

Read level
full — full (IJCARS 2022; arXiv 2009.05859, PDF ); full (zero-slack, 2-page arXiv 2204.00645, )
Picture
kim2022_ice_IJCARS Fig. 1(b) robotic ICE manipulator (distal and proximal units) on the pig ↗ — licence: publisher copyright (Springer, (c) CARS 2022); grade C
Picture changes
native extract (1500x961); cropped to panel (b) without its panel caption; label boxes and arrows on the robot left as is; not upscaled
Found by
gap: Gap searches for robots the other searches missed.
Data
JSON · YAML: records/robots/siemens_ice_kim2020.yaml
Correction
Suggest a correction [repository not created yet]