RATIONAL

Hangzhou Denova (Dinova) Neuro Medtech, 杭州德诺脑神经医疗科技有限公司; co-developed with Beijing Tiantan Hospital (Gao Feng team) · 2023–2026
RATIONAL: main picture1 / 5
Not cleared for distribution. https://medtion-image.medtion.com/pgc/20250409/8911cefa18e30db24504d880554b0cca. (857 x 408 company promo image, 'RATIONAL 手术控制系统' + '配套耗材'), page https://www.brainmed.com/info/detail?id=51164 (脑医汇 2025-04-09, reposted from WeChat 神经介入在线); same pixels as the author's screenshot figures/raw/web/rational_user_promo.png. Licence: news/company image, publisher copyright. Picture grade C.
Family
Commercial and clinical-stage systems (clinical-stage)
Devices
4: microguidewire, microcatheter, aspiration catheter, stent retriever; guide/aspiration catheters inserted manually
Degrees of freedom
N.R.
Push
Not reported
Twist
Not reported
Tested up to
Human
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Human
What was tested
10 pts first-in-human, Beijing Tiantan, Oct 2023-Jun 2024 (yang2025_use); trial NCT05972252, start 2023-10-15, Beijing Tiantan Hospital (clinicaltrials2023_nct05972252)
Key numbers
mTICI 3 in 8, 2b in 2; puncture-to-reperfusion 65.10+-22.23 min (robotic 13.70+-5.18, manual 51.40+-20.16); console dose 0.18+-0.11 vs 59.21+-52.57 uSv; 90-day mRS<=2 in 7; technician device exchange = main limitation

Mechanism

Push (translation)
N.R.
Twist (rotation)
N.R.
Device range
N.R.
Target anatomy
neurovascular (mechanical thrombectomy, anterior circulation)

Sensing and interface

Force sensing
N.R.
Leader and haptics
N.R. (three joysticks)
Imaging
console outside room
Sterility
preloaded single-use cassette (DNNSJ07-IR)

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
RATIONAL surgical control system (company promo); registered as Cerebrovascular Interventional Procedural Control System (NCT05972252)

Sources

  1. Use of a Robot-Assisted Thrombectomy System in Acute Ischemic Stroke · JAMA Neurology, 2025 · doi:10.1001/jamaneurol.2025.0418 [yang2025_use]
  2. Robot-Assisted Mechanical Thrombectomy for Neurointerventional Surgery: An Exploratory Scoping Review · Journal of Robotic Surgery, 2026 · doi:10.1007/s11701-026-03535-x [riosgarcia2026_thrombectomy]
  3. JAMA Neurology: Early Clinical Results of the First Use of Robot-Assisted Mechanical Thrombectomy for Acute Ischemic Stroke · link [brainmed2025_rational]
  4. First-in-man Study of the Cerebrovascular Interventional Procedural Control System (A Feasibility Study to Evaluate the Safety and Efficacy of the Cerebrovascular Interventional Procedural Control System in Treatment of Acute Ischemic Stroke) · link [clinicaltrials2023_nct05972252]

Notes

Notes
Only human thrombectomy robot series with all four devices driven. No regulatory approval found (search for 2026 NMPA approval found nothing). Developer: 杭州德诺脑神经医疗科技有限公司 (trial sponsor name 'Hangzhou Dinova Neuroscience Technology Co., Ltd'; 'Denova Medtech' in yang2025_use), co-developed with Beijing Tiantan Hospital (Gao Feng team) over five years: brainmed 2025-04-09, https://www.brainmed.com/info/detail?id=51164 (reposted from WeChat 神经介入在线), and ClinicalTrials.gov API v2 NCT05972252 (P9_two_robots.md §1). Trial NCT05972252 'First-in-man Study of the Cerebrovascular Interventional Procedural Control System': Beijing Tiantan, contact Feng Gao MD, start 2023-10-15, status UNKNOWN (last verified 2023-10). Consumable: yang2025_use writes DNNSJ07-IR, brainmed DNNS107-IR; the paper's spelling is kept. X5 excluded the row (no picture); X6 2026-11-08 re-included it after P9 found the picture (data/images_P9.csv, grade C).

This record

Read level
partial — partial (PMC page via S8); web (brainmed 2025-04-09; ClinicalTrials.gov API v2)
Picture
https://medtion-image.medtion.com/pgc/20250409/8911cefa18e30db24504d880554b0cca. (857 x 408 company promo image, 'RATIONAL 手术控制系统' + '配套耗材'), page https://www.brainmed.com/info/detail?id=51164 (脑医汇 2025-04-09, reposted from WeChat 神经介入在线); same pixels as the author's screenshot figures/raw/web/rational_user_promo.png ↗ — licence: news/company image, publisher copyright; grade C
Picture changes
RGBA composited on white; 2 orange leader lines on the plain white background (neither touches the robot) painted white by colour mask + 2 px dilation; 2 text blocks on plain white painted white (right: 2 lines of claims; left: end of the consumables text); fragments of the consumables graphic at the left edge (stent-retriever tip, tube end) painted white; cropped to the robot and table (about 22 px margin; the pedestal is cut by the original's bottom border); no inpainting model; not upscaled
Found by
commercial: Search of commercial and clinical-stage systems.
Data
JSON · YAML: records/robots/rational_denova2023.yaml