 "schema_version": 1,
 "id": "rational_denova2023",
 "name": "RATIONAL",
 "full_name": "RATIONAL",
 "group": "Hangzhou Denova (Dinova) Neuro Medtech, 杭州德诺脑神经医疗科技有限公司; co-developed with Beijing Tiantan Hospital (Gao Feng team)",
 "group_short": "Denova Medtech",
 "family": "commercial",
 "category": "clinical-stage",
 "first_year": 2023,
 "latest_year": 2026,
 "versions": "RATIONAL surgical control system (company promo); registered as Cerebrovascular Interventional Procedural Control System (NCT05972252)",
 "target_anatomy": "neurovascular (mechanical thrombectomy, anterior circulation)",
 "devices": [
  "guidewire",
  "microcatheter",
  "catheter",
  "therapeutic-device"
 ],
 "devices_driven": "4: microguidewire, microcatheter, aspiration catheter, stent retriever; guide/aspiration catheters inserted manually",
 "n_dof": null,
 "n_dof_detail": "N.R.",
 "translation_mechanism": "not-reported",
 "translation_detail": "N.R.",
 "rotation_mechanism": "not-reported",
 "rotation_detail": "N.R.",
 "device_range": "N.R.",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "N.R.",
 "haptics_leader": "N.R. (three joysticks)",
 "imaging_compat": "console outside room",
 "sterility": "preloaded single-use cassette (DNNSJ07-IR)",
 "highest_validation": "human",
 "validation_detail": "10 pts first-in-human, Beijing Tiantan, Oct 2023-Jun 2024 (yang2025_use); trial NCT05972252, start 2023-10-15, Beijing Tiantan Hospital (clinicaltrials2023_nct05972252)",
 "key_numbers": "mTICI 3 in 8, 2b in 2; puncture-to-reperfusion 65.10+-22.23 min (robotic 13.70+-5.18, manual 51.40+-20.16); console dose 0.18+-0.11 vs 59.21+-52.57 uSv; 90-day mRS<=2 in 7; technician device exchange = main limitation",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "C",
  "kind": "render",
  "source": "https://medtion-image.medtion.com/pgc/20250409/8911cefa18e30db24504d880554b0cca. (857 x 408 company promo image, 'RATIONAL 手术控制系统' + '配套耗材'), page https://www.brainmed.com/info/detail?id=51164 (脑医汇 2025-04-09, reposted from WeChat 神经介入在线); same pixels as the author's screenshot figures/raw/web/rational_user_promo.png",
  "source_url": "https://medtion-image.medtion.com/pgc/20250409/8911cefa18e30db24504d880554b0cca",
  "licence": "news/company image, publisher copyright",
  "credit": "https://medtion-image.medtion.com/pgc/20250409/8911cefa18e30db24504d880554b0cca. (857 x 408 company promo image, 'RATIONAL 手术控制系统' + '配套耗材'), page https://www.brainmed.com/info/detail?id=51164 (脑医汇 2025-04-09, reposted from WeChat 神经介入在线); same pixels as the author's screenshot figures/raw/web/rational_user_promo.png. Licence: news/company image, publisher copyright.",
  "edits": "RGBA composited on white; 2 orange leader lines on the plain white background (neither touches the robot) painted white by colour mask + 2 px dilation; 2 text blocks on plain white painted white (right: 2 lines of claims; left: end of the consumables text); fragments of the consumables graphic at the left edge (stent-retriever tip, tube end) painted white; cropped to the robot and table (about 22 px margin; the pedestal is cut by the original's bottom border); no inpainting model; not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "rational_denova2023_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "leader",
   "caption": "Whole system: table-mounted arm with drive heads, aspiration pump, console with three joysticks and touch panel, monitors (labelled figure panel)",
   "source_key": "web:brainmed.com/info/detail?id=51164",
   "source": "brainmed.com/info/detail?id=51164, article image (panel 'a', labels A1/A2/D1/D2; original publication not stated)",
   "source_url": "https://medtion-image.medtion.com/pgc/20250409/27f5f82e1f1cb5273e45415a14170b14",
   "where": "article image https://medtion-image.medtion.com/pgc/20250409/27f5f82e1f1cb5273e45415a14170b14. (panel 'a', labels A1/A2/D1/D2; original publication not stated)",
   "licence": "news/company image, publisher copyright"
  },
  {
   "id": "rational_denova2023_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Tiantan team in the angiography suite; the draped robot on the table at the right",
   "source_key": "web:brainmed.com/info/detail?id=51164",
   "source": "brainmed.com/info/detail?id=51164, article image",
   "source_url": "https://medtion-image.medtion.com/pgc/20250409/d2ed8b793d2329e0426a1946720532b7",
   "where": "article image https://medtion-image.medtion.com/pgc/20250409/d2ed8b793d2329e0426a1946720532b7.",
   "licence": "news/company image, publisher copyright"
  },
  {
   "id": "rational_denova2023_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "leader",
   "caption": "Company promo image: the robot on the table with its matched consumables (stent retriever, microcatheter, support catheter, aspiration pump), with labels",
   "source_key": "web:brainmed.com/info/detail?id=51164",
   "source": "brainmed.com/info/detail?id=51164, article image (uncropped source of the main picture)",
   "source_url": "https://medtion-image.medtion.com/pgc/20250409/8911cefa18e30db24504d880554b0cca",
   "where": "article image https://medtion-image.medtion.com/pgc/20250409/8911cefa18e30db24504d880554b0cca. (uncropped source of the main picture)",
   "licence": "news/company image, publisher copyright"
  },
  {
   "id": "rational_denova2023_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "leader",
   "caption": "Patent drawing: catheter controller whose delivery turntable stores and feeds the catheter (Denova patent; tie to RATIONAL inferred)",
   "source_key": "patent:WO2023274259A1",
   "source": "patent:WO2023274259A1, Fig. 1, (applicant 杭州德诺脑神经医疗科技有限公司)",
   "source_url": "https://patentimages.storage.googleapis.com/09/dd/0e/a374aa1ff4b0a5/PCTCN2022102091-appb-300001.png",
   "where": "Fig. 1, https://patentimages.storage.googleapis.com/09/dd/0e/a374aa1ff4b0a5/PCTCN2022102091-appb-300001.png (applicant 杭州德诺脑神经医疗科技有限公司)",
   "licence": "patent drawing (public)"
  }
 ],
 "sources": [
  {
   "key": "yang2025_use",
   "doi": "10.1001/jamaneurol.2025.0418",
   "url": null,
   "title": "Use of a Robot-Assisted Thrombectomy System in Acute Ischemic Stroke",
   "year": 2025,
   "venue": "JAMA Neurology"
  },
  {
   "key": "riosgarcia2026_thrombectomy",
   "doi": "10.1007/s11701-026-03535-x",
   "url": null,
   "title": "Robot-Assisted Mechanical Thrombectomy for Neurointerventional Surgery: An Exploratory Scoping Review",
   "year": 2026,
   "venue": "Journal of Robotic Surgery"
  },
  {
   "key": "brainmed2025_rational",
   "doi": null,
   "url": "https://www.brainmed.com/info/detail?id=51164",
   "title": "JAMA Neurology: Early Clinical Results of the First Use of Robot-Assisted Mechanical Thrombectomy for Acute Ischemic Stroke",
   "year": 2025,
   "venue": null
  },
  {
   "key": "clinicaltrials2023_nct05972252",
   "doi": null,
   "url": "https://clinicaltrials.gov/study/NCT05972252",
   "title": "First-in-man Study of the Cerebrovascular Interventional Procedural Control System (A Feasibility Study to Evaluate the Safety and Efficacy of the Cerebrovascular Interventional Procedural Control System in Treatment of Acute Ischemic Stroke)",
   "year": 2023,
   "venue": null
  }
 ],
 "read_level": "partial",
 "read_level_detail": "partial (PMC page via S8); web (brainmed 2025-04-09; ClinicalTrials.gov API v2)",
 "notes": "Only human thrombectomy robot series with all four devices driven. No regulatory approval found (search for 2026 NMPA approval found nothing). Developer: 杭州德诺脑神经医疗科技有限公司 (trial sponsor name 'Hangzhou Dinova Neuroscience Technology Co., Ltd'; 'Denova Medtech' in yang2025_use), co-developed with Beijing Tiantan Hospital (Gao Feng team) over five years: brainmed 2025-04-09, https://www.brainmed.com/info/detail?id=51164 (reposted from WeChat 神经介入在线), and ClinicalTrials.gov API v2 NCT05972252 (P9_two_robots.md §1). Trial NCT05972252 'First-in-man Study of the Cerebrovascular Interventional Procedural Control System': Beijing Tiantan, contact Feng Gao MD, start 2023-10-15, status UNKNOWN (last verified 2023-10). Consumable: yang2025_use writes DNNSJ07-IR, brainmed DNNS107-IR; the paper's spelling is kept. X5 excluded the row (no picture); X6 2026-11-08 re-included it after P9 found the picture (data/images_P9.csv, grade C).",
 "maintainers_own": false,
 "found_by": "commercial"
}