Catheter Robot Atlas

Academic bedside robots · portable_kantu2025

Portable robot

Portable and versatile catheter robot
NC State Univ. / NYU (Buckner, Su) · 2025
Portable robot: main picture1 / 9
Photo: BIRO Lab, New York University. Picture grade A.
Family
Academic bedside robots (academic)
Devices
guide catheter 5-9 F (5 F and 8 F tested) and guidewire 0.33-1 mm (0.36 mm and 0.89 mm tested)
Degrees of freedom
4 (guide catheter T+R, guidewire T+R)
Push
Other drives
Twist
Rotating clamp
Tested up to
Phantom
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Phantom
What was tested
benchtop tracking with motion capture; force/torque tests (2 catheters, 2 wires); guidewire advanced to left coronary ostium in a 3D-printed human-heart phantom with EM tracker (one demonstration)
Key numbers
peak force 5.49 N (8 F), 3.44 N (5 F), wire 0.38/0.35 N; peak torque 10.61/6.55 mN.m (8F/5F), wire 0.42/0.22 mN.m; tracking RMSE 10.3 deg and 0.31 mm; wire deviation ~6.9 deg / 0.9 mm; 90% smaller than CorPath (1780x690x1170 mm)

Mechanism

Push (translation)
other: guide catheter on 1.5 mm-pitch lead-screw platform (300 mm range); guidewire by active friction roller plus spring-loaded passive roller, range > 1200 mm
Twist (rotation)
guide catheter: BLDC motor with 2:1 spur gears (360 deg); guidewire: commercial torque device on motor gear train (360 deg)
Device range
catheter 5-9 F; wire 0.33-1 mm (abstract: 1-9 F; text: clamp for 0.33-3 mm)
Target anatomy
coronary (simulated PCI in 3D-printed heart model)

Sensing and interface

Force sensing
motor current; external load cell in bench tests only
Leader and haptics
none
Imaging
N.R.
Sterility
N.R. (cartridge for sterilisation proposed in discussion)

Size

Footprint (mm)
250 x 350 x 250 (abstract; LegoBot Table I writes 360 x 250 x 250)
Mass (kg)
N.R. (direct-drive motor 92 g each)

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
2025 (single version)

Sources

  1. Portable and Versatile Catheter Robot for Image-Guided Cardiovascular Interventions · IEEE/ASME Transactions on Mechatronics, 2025 · doi:10.1109/TMECH.2025.3559911 [kantu2025_portable]
  2. LegoBot: Modular, compact, device-agnostic robot for endovascular catheter and guidewire manipulation in stroke thrombectomy · IEEE/ASME Transactions on Mechatronics, 2026 · doi:10.1109/TMECH.2026.3695773 [gao2026_legobot]

Notes

Notes
Coupled guidewire translation-rotation per LegoBot Table I; direct-drive 40x28 mm motors 0.15 N.m nominal / 0.33 N.m peak.

This record

Read level
full
Picture
BIRO Lab, New York University (own photograph) — licence: lab own (BIRO Lab, New York University); grade A
Found by
academic: Search of academic bedside robots.
Data
JSON · YAML: records/robots/portable_kantu2025.yaml
Correction