 "schema_version": 1,
 "id": "portable_kantu2025",
 "name": "Portable robot",
 "full_name": "Portable and versatile catheter robot",
 "group": "NC State Univ. / NYU (Buckner, Su)",
 "group_short": "NC State / NYU",
 "family": "academic",
 "category": "academic",
 "first_year": 2025,
 "latest_year": 2025,
 "versions": "2025 (single version)",
 "target_anatomy": "coronary (simulated PCI in 3D-printed heart model)",
 "devices": [
  "guidewire",
  "catheter"
 ],
 "devices_driven": "guide catheter 5-9 F (5 F and 8 F tested) and guidewire 0.33-1 mm (0.36 mm and 0.89 mm tested)",
 "n_dof": 4,
 "n_dof_detail": "4 (guide catheter T+R, guidewire T+R)",
 "translation_mechanism": "other",
 "translation_detail": "other: guide catheter on 1.5 mm-pitch lead-screw platform (300 mm range); guidewire by active friction roller plus spring-loaded passive roller, range > 1200 mm",
 "rotation_mechanism": "rotating-clamp",
 "rotation_detail": "guide catheter: BLDC motor with 2:1 spur gears (360 deg); guidewire: commercial torque device on motor gear train (360 deg)",
 "device_range": "catheter 5-9 F; wire 0.33-1 mm (abstract: 1-9 F; text: clamp for 0.33-3 mm)",
 "footprint_mm": "250 x 350 x 250 (abstract; LegoBot Table I writes 360 x 250 x 250)",
 "mass_kg": "N.R. (direct-drive motor 92 g each)",
 "force_sensing": "motor current; external load cell in bench tests only",
 "haptics_leader": "none",
 "imaging_compat": "N.R.",
 "sterility": "N.R. (cartridge for sterilisation proposed in discussion)",
 "highest_validation": "phantom",
 "validation_detail": "benchtop tracking with motion capture; force/torque tests (2 catheters, 2 wires); guidewire advanced to left coronary ostium in a 3D-printed human-heart phantom with EM tracker (one demonstration)",
 "key_numbers": "peak force 5.49 N (8 F), 3.44 N (5 F), wire 0.38/0.35 N; peak torque 10.61/6.55 mN.m (8F/5F), wire 0.42/0.22 mN.m; tracking RMSE 10.3 deg and 0.31 mm; wire deviation ~6.9 deg / 0.9 mm; 90% smaller than CorPath (1780x690x1170 mm)",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "cleared",
  "grade": "A",
  "kind": "photo",
  "source": "BIRO Lab, New York University (own photograph)",
  "source_url": null,
  "licence": "lab own (BIRO Lab, New York University)",
  "credit": "Photo: BIRO Lab, New York University.",
  "edits": "",
  "file": "pictures/portable_kantu2025.jpg"
 },
 "pictures": [
  {
   "id": "portable_kantu2025_01",
   "status": "cleared",
   "file": "pictures/extra/portable_kantu2025_01.jpg",
   "thumb": "pictures/thumbs/portable_kantu2025_01.jpg",
   "kind": "close-up",
   "caption": "Top module with guide-catheter clamp, rotation gear, lead screw and rails",
   "source_key": "kantu2025_portable",
   "source": "Kantu et al., IEEE/ASME Transactions on Mechatronics (2025), lab photo image50.jpg, unannotated original of Fig. 4(b)",
   "source_url": "https://doi.org/10.1109/TMECH.2025.3559911",
   "where": "lab photo image50.jpg, unannotated original of Fig. 4(b)",
   "licence": "lab own (collaborator material)"
  },
  {
   "id": "portable_kantu2025_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "photo",
   "caption": "Assembled 4-DoF prototype from above, top, middle and bottom modules labelled",
   "source_key": "kantu2025_portable",
   "source": "Kantu et al., IEEE/ASME Transactions on Mechatronics (2025), Fig. 4, centre photo, p. 5",
   "source_url": "https://doi.org/10.1109/TMECH.2025.3559911",
   "where": "Fig. 4, centre photo, p. 5",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "portable_kantu2025_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Guidewire module: roller insertion drive and rotation gear",
   "source_key": "kantu2025_portable",
   "source": "Kantu et al., IEEE/ASME Transactions on Mechatronics (2025), Fig. 4(c), p. 5",
   "source_url": "https://doi.org/10.1109/TMECH.2025.3559911",
   "where": "Fig. 4(c), p. 5",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "portable_kantu2025_04",
   "status": "cleared",
   "file": "pictures/extra/portable_kantu2025_04.jpg",
   "thumb": "pictures/thumbs/portable_kantu2025_04.jpg",
   "kind": "in-use",
   "caption": "Robot with top covers on feeding a guide catheter toward a heart model",
   "source_key": "kantu2025_portable",
   "source": "Kantu et al., IEEE/ASME Transactions on Mechatronics (2025), lab video poster frame image18.png, PCI lecture deck",
   "source_url": "https://doi.org/10.1109/TMECH.2025.3559911",
   "where": "lab video poster frame image18.png, PCI lecture deck",
   "licence": "lab own (collaborator material)"
  },
  {
   "id": "portable_kantu2025_05",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Robot deploying a guidewire into a human heart phantom with EM tracking",
   "source_key": "kantu2025_portable",
   "source": "Kantu et al., IEEE/ASME Transactions on Mechatronics (2025), Fig. 9 (top), p. 8",
   "source_url": "https://doi.org/10.1109/TMECH.2025.3559911",
   "where": "Fig. 9 (top), p. 8",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "portable_kantu2025_06",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Transparent CAD of the three-module robot with instrument paths and dimensions",
   "source_key": "kantu2025_portable",
   "source": "Kantu et al., IEEE/ASME Transactions on Mechatronics (2025), Fig. 2(a), p. 3",
   "source_url": "https://doi.org/10.1109/TMECH.2025.3559911",
   "where": "Fig. 2(a), p. 3",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "portable_kantu2025_07",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "CAD of the guidewire drive: active and passive rollers, torquer, gears",
   "source_key": "kantu2025_portable",
   "source": "Kantu et al., IEEE/ASME Transactions on Mechatronics (2025), Fig. 2(c), p. 3",
   "source_url": "https://doi.org/10.1109/TMECH.2025.3559911",
   "where": "Fig. 2(c), p. 3",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "portable_kantu2025_08",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Guide-catheter clamp holding 5 F and 8 F catheters",
   "source_key": "kantu2025_portable",
   "source": "Kantu et al., IEEE/ASME Transactions on Mechatronics (2025), Fig. 5(a)-(b), p. 6",
   "source_url": "https://doi.org/10.1109/TMECH.2025.3559911",
   "where": "Fig. 5(a)-(b), p. 6",
   "licence": "publisher copyright (IEEE)"
  }
 ],
 "sources": [
  {
   "key": "kantu2025_portable",
   "doi": "10.1109/TMECH.2025.3559911",
   "url": null,
   "title": "Portable and Versatile Catheter Robot for Image-Guided Cardiovascular Interventions",
   "year": 2025,
   "venue": "IEEE/ASME Transactions on Mechatronics"
  },
  {
   "key": "gao2026_legobot",
   "doi": "10.1109/TMECH.2026.3695773",
   "url": null,
   "title": "LegoBot: Modular, compact, device-agnostic robot for endovascular catheter and guidewire manipulation in stroke thrombectomy",
   "year": 2026,
   "venue": "IEEE/ASME Transactions on Mechatronics"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full",
 "notes": "Coupled guidewire translation-rotation per LegoBot Table I; direct-drive 40x28 mm motors 0.15 N.m nominal / 0.33 N.m peak.",
 "maintainers_own": true,
 "found_by": "academic"
}