Commercial and clinical-stage systems · phocus_aeon2015

Aeon Phocus

Aeon Phocus electromagnetic catheter steering system
Aeon Scientific AG (ETH Zurich spin-off) · 2015–2017
Aeon Phocus: main picture1 / 3
Not cleared for distribution. chautems2017_variable Fig. 1(B) (same photo as li2020_magnetic Fig. 6a). Licence: publisher copyright. Picture grade B.
Family
Commercial and clinical-stage systems (magnetic-navigation)
Devices
magnetic ablation catheters (model N.R.); in ETH tests a tethered magnet (NdFeB, 29 mm long, 5 mm radius) and variable-stiffness magnetic catheters with LMPA segments (chautems2017_tethered Fig. 7; chautems2017_variable)
Degrees of freedom
3 (alekseeva2025_design Table 14); field and field gradient from 7 electromagnets (chautems2017_tethered Sec. II)
Push
Magnetic
Twist
Field direction
Tested up to
Bench
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Bench
What was tested
bench only in the sources read: tethered-magnet trajectories and force measurements inside the Aeon Phocus at Aeon's lab (chautems2017_tethered Sec. III); variable-stiffness catheter configurations in the Aeon field (chautems2017_variable Fig. 3). Clinical use is stated ('a medically certified system currently in clinical use', chautems2017_tethered abstract) but no patient study was found; alekseeva2025_design Table 14 gives 'ex vivo phantom studies (TRL3)'
Key numbers
7 electromagnets; model vs measured force error below 0.01 N, nine points at 0.071 +/- 0.005 N (chautems2017_tethered Sec. III-C); field controlled closed-loop at 1 Hz in the ETH tests (chautems2017_tethered Sec. II); 100 mT used to load a VS catheter with 0.003 Nm (chautems2017_variable Sec. III)

Mechanism

Push (translation)
magnetic (field gradient on the tip magnet); catheter advance mechanism N.R.
Twist (rotation)
tip direction by field of 7 electromagnets arranged around the patient's torso (li2020_magnetic Sec. 3.4); static electromagnets (chautems2017_tethered Sec. I) vs 'movable electromagnets' (mandhumula2025_robotic): conflict
Device range
N.R.
Target anatomy
cardiac arrhythmias (catheter ablation) in all four heart chambers (chautems2017_tethered Sec. III-A)

Sensing and interface

Force sensing
none in the system; ETH added Hall sensors around the workspace for field correction (chautems2017_tethered Sec. II)
Leader and haptics
no: joystick in a separate control room (li2020_magnetic Sec. 3.4; chautems2017_tethered Sec. III-A)
Imaging
integrated with a Siemens Artis Zee angiography system (chautems2017_tethered Fig. 6; li2020_magnetic Sec. 3.4)
Sterility
N.R.

Size

Footprint (mm)
N.R. (encloses the patient's upper body, li2020_magnetic Fig. 6b)
Mass (kg)
N.R.

Simulation

Simulation model
none found (linear field model with saturation correction, no simulator; chautems2017_tethered Sec. II)
Sim-trained controller
none found

Versions

Lineage
single system: CE mark announced 2015-07-30 (startupticker2015_aeon); host of the ETH magnetic-catheter tests of 2017 (chautems2017_tethered, chautems2017_variable); company no longer in business (li2020_magnetic Sec. 3.4); 'discontinued despite regulatory approval' (mandhumula2025_robotic)

Sources

  1. The tethered magnet: Force and 5-DOF pose control for cardiac ablation · 2017 IEEE International Conference on Robotics and Automation (ICRA), 2017 · doi:10.1109/icra.2017.7989562 [chautems2017_tethered]
  2. A variable stiffness catheter controlled with an external magnetic field · 2017 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS), 2017 · doi:10.1109/iros.2017.8202155 [chautems2017_variable]
  3. Green light to Aeon Scientific to launch its Aeon Phocus · no DOI on file [startupticker2015_aeon]
  4. A review of magnetic actuation systems and magnetically actuated guidewire- and catheter-based microrobots for vascular interventions · Intelligent Service Robotics, 2020 · doi:10.1007/s11370-020-00311-0 [li2020_magnetic]
  5. Robotic Systems Design in Endovascular Treatment · IEEE Transactions on Medical Robotics and Bionics, 2024 · doi:10.1109/tmrb.2024.3377374 [li2024_design]
  6. Robotic Systems in Cardiovascular Interventions: Evolving Platforms and the Path Forward · Methodist DeBakey Cardiovascular Journal, 2025 · doi:10.14797/mdcvj.1652 [mandhumula2025_robotic]
  7. The Design of Endovascular Robots for Cardiac Interventions: A Survey · Journal of Field Robotics, 2024 · doi:10.1002/rob.22480 [alekseeva2025_design]

Notes

Notes
Family: commercial (A4 scope; CE-marked magnetic navigation). A4 had set it aside as EP-only; the author put it in (round 8). Nanoflex (navion_nanoflex2025) is a later, different ETH spin-off. Same ETH group tested the tethered magnet in it (review_coverage rows 'ETH_CHAUTEMS', not added here).

This record

Read level
full — full (chautems2017_tethered, chautems2017_variable); web (startupticker2015_aeon)
Picture
chautems2017_variable Fig. 1(B) (same photo as li2020_magnetic Fig. 6a) ↗ — licence: publisher copyright; grade B