Catheter Robot Atlas

Magnetic navigation · mpi_tiryaki2023

7-T MRI steering

Ultrahigh-field magnetic guidewire steering inside a 7-T MRI scanner with a robotic MRI stage
Sitti, Physical Intelligence Department, Max Planck Institute for Intelligent Systems · 2023
7-T MRI steering: main picture1 / 3
Tiryaki et al., Science Advances (2023), Fig. 1. CC BY 4.0. Changes: cropped panel A from the 785 px PMC figure; panel letter removed by crop; none upscaled. Source ↗ Picture grade C.
Family
Magnetic navigation (magnetic-navigation)
Devices
own guidewire: MRI-compatible glass fibre-optic rod core with NdFeB tip magnets (1 mm x 1 mm cylinders) in parallel, perpendicular, antiparallel or dual configurations; 900 um Teflon supporting catheter (tiryaki2023_magnetic Results)
Degrees of freedom
robotic MRI stage: x, y and yaw; manual insertion and base twist (tiryaki2023_magnetic Results, Fig. 5A)
Push
Magnetic
Twist
Field direction
Tested up to
Phantom
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Phantom
What was tested
3D glass vascular models with physiological flow (renal artery, aortic arch at 100 mL per beat, CCA, MCAs), access through a Teflon port 40 cm from the target (tiryaki2023_magnetic Fig. 7); 2D obstacle tasks with three targets (Fig. 5); ex vivo porcine kidney renal cavity under real-time MRI, guidewire through the urinary tract, not a vessel (Fig. 8)
Key numbers
7 T; fringe gradients up to 20 T/m at the bore entrance; tip workspace 161 deg at 7 T (Fig. 3C); antiparallel magnet rotates up to 164 deg; platform rotation up to 48 deg needed in 2D tasks (tiryaki2023_magnetic Results)

Mechanism

Push (translation)
magnetic (tip steering by MRI main-field torque and fringe-field gradient); insertion by hand (tiryaki2023_magnetic Fig. 5A, Fig. 7)
Twist (rotation)
motorized stage moves and rotates the phantom relative to the main field; manual base twist (tiryaki2023_magnetic Results)
Device range
N.R.
Target anatomy
endovascular under MRI: renal artery, type 1 aortic arch, common carotid and middle cerebral artery glass models (tiryaki2023_magnetic Fig. 7); ex vivo porcine kidney renal cavity (non-vascular, Fig. 8)

Sensing and interface

Force sensing
none during navigation (MRI-conditioned force sensor used only to calibrate magnet force)
Leader and haptics
no: joystick / remote controller for the stage (tiryaki2023_magnetic Results)
Imaging
MRI (real-time 1-Hz FISP in the kidney test; tip seen as a 37 x 29 mm susceptibility artefact) (tiryaki2023_magnetic Fig. 8)
Sterility
N.R.

Size

Footprint (mm)
N.R. (inside a 7-T preclinical MRI bore)
Mass (kg)
N.R.

Simulation

Simulation model
Cosserat rod model (20 segments) with a UHF magnetization model, used for guidewire design (tiryaki2023_magnetic Methods)
Sim-trained controller
none found (model used for design, no trained controller)

Versions

Lineage
single (Sci. Adv. 2023)

Sources

  1. Magnetic guidewire steering at ultrahigh magnetic fields · Science Advances, 2023 · doi:10.1126/sciadv.adg6438 [tiryaki2023_magnetic]
  2. Robotically steerable guidewires—Current trends and future directions · Science Robotics, 2025 · doi:10.1126/scirobotics.adt7461 [konda2025_robotically]

Notes

Notes
Family: magnetic. konda2025 Table 2 row 4. The guidewire is advanced by hand; motor control is the stage that moves the phantom in the field. No photo of the stage in the paper body (fig. S7 in the supplement, not opened: manual list).

This record

Read level
full — full (PMC full text)
Picture
tiryaki2023_magnetic Fig. 1(A) (PMC10132757 figure file sciadv.adg6438-f1.jpg) ↗ — licence: CC BY 4.0; grade C
Picture changes
cropped panel A from the 785 px PMC figure; panel letter removed by crop; none upscaled
Found by
actuated-tip: Search of actuated-tip and magnetic devices.
Data
JSON · YAML: records/robots/mpi_tiryaki2023.yaml
Correction
Suggest a correction [repository not created yet]