Catheter Robot Atlas

Academic bedside robots · nebraska_mohammadi2026

Haptic wire robot

Haptic portable guidewire robot
Univ. of Nebraska Omaha / UNMC (Kamenskiy, MacTaggart) · 2026
Haptic wire robot: main picture1 / 5
Mohammadi et al., Annals of Biomedical Engineering (2026), Fig. 8(a). CC BY 4.0. Changes: native extract; cropped inside the panel border; panel letter 'a' erased (LaMa, on wood, not touching the robot); 12 px trimmed at top-left; not upscaled. Source ↗ Picture grade A.
Family
Academic bedside robots (academic)
Devices
guidewire or catheter (standard tools)
Degrees of freedom
2 (advance/retract, axial twist)
Push
Clamp relay
Twist
Rotating clamp
Tested up to
Bench
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Bench
What was tested
flexible tubing with 20-80% stenosis inserts, 160 experiments (two current thresholds, rigid and flexible inserts)
Key numbers
400 g; lesion stiffness F-score 0.83 / 0.77 / 0.95 (soft / medium / stiff); retractions rise with stenosis

Mechanism

Push (translation)
clamp-relay (servo-driven timing belt; gripper engaged on advance, released on retract)
Twist (rotation)
front servo plus twisting gripper (rack-and-pinion)
Device range
N.R.
Target anatomy
peripheral / prehospital use; bench stenosis tube

Sensing and interface

Force sensing
servo current only
Leader and haptics
current-based impedance control with virtual mass-spring-damper; autonomous mode
Imaging
N.R.
Sterility
N.R.

Size

Footprint (mm)
160 x 75 x 90
Mass (kg)
0.4

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
medRxiv Feb 2026; Ann Biomed Eng 2026 (DOI 10.1007/s10439-026-04315-2)

Sources

  1. Haptic Portable Robotic Device for Automated Guidewire or Catheter Navigation in Endovascular Procedures · Annals of Biomedical Engineering, 2026 · doi:10.1007/s10439-026-04315-2 [mohammadi2026_haptic]

Notes

Notes
Autonomous rule-based controller, not learned.

This record

Read level
full
Picture
mohammadi2026_haptic Fig. 8(a) (Ann Biomed Eng version 10.1007/s10439-026-04315-2; embedded 1500x1917) ↗ — licence: CC BY 4.0; grade A
Picture changes
native extract; cropped inside the panel border; panel letter 'a' erased (LaMa, on wood, not touching the robot); 12 px trimmed at top-left; not upscaled
Found by
academic: Search of academic bedside robots.
Data
JSON · YAML: records/robots/nebraska_mohammadi2026.yaml
Correction
Suggest a correction [repository not created yet]