 "schema_version": 1,
 "id": "nebraska_mohammadi2026",
 "name": "Haptic wire robot",
 "full_name": "Haptic portable guidewire robot",
 "group": "Univ. of Nebraska Omaha / UNMC (Kamenskiy, MacTaggart)",
 "group_short": "U. Nebraska",
 "family": "academic",
 "category": "academic",
 "first_year": 2026,
 "latest_year": 2026,
 "versions": "medRxiv Feb 2026; Ann Biomed Eng 2026 (DOI 10.1007/s10439-026-04315-2)",
 "target_anatomy": "peripheral / prehospital use; bench stenosis tube",
 "devices": [
  "guidewire",
  "catheter"
 ],
 "devices_driven": "guidewire or catheter (standard tools)",
 "n_dof": 2,
 "n_dof_detail": "2 (advance/retract, axial twist)",
 "translation_mechanism": "clamp-relay",
 "translation_detail": "clamp-relay (servo-driven timing belt; gripper engaged on advance, released on retract)",
 "rotation_mechanism": "rotating-clamp",
 "rotation_detail": "front servo plus twisting gripper (rack-and-pinion)",
 "device_range": "N.R.",
 "footprint_mm": "160 x 75 x 90",
 "mass_kg": "0.4",
 "force_sensing": "servo current only",
 "haptics_leader": "current-based impedance control with virtual mass-spring-damper; autonomous mode",
 "imaging_compat": "N.R.",
 "sterility": "N.R.",
 "highest_validation": "bench",
 "validation_detail": "flexible tubing with 20-80% stenosis inserts, 160 experiments (two current thresholds, rigid and flexible inserts)",
 "key_numbers": "400 g; lesion stiffness F-score 0.83 / 0.77 / 0.95 (soft / medium / stiff); retractions rise with stenosis",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "cleared",
  "grade": "A",
  "kind": "photo",
  "source": "mohammadi2026_haptic Fig. 8(a) (Ann Biomed Eng version 10.1007/s10439-026-04315-2; embedded 1500x1917)",
  "source_url": "https://doi.org/10.1007/s10439-026-04315-2",
  "licence": "CC BY 4.0",
  "credit": "Mohammadi et al., Annals of Biomedical Engineering (2026), Fig. 8(a). CC BY 4.0.",
  "edits": "native extract; cropped inside the panel border; panel letter 'a' erased (LaMa, on wood, not touching the robot); 12 px trimmed at top-left; not upscaled",
  "file": "pictures/nebraska_mohammadi2026.jpg"
 },
 "pictures": [
  {
   "id": "nebraska_mohammadi2026_01",
   "status": "cleared",
   "file": "pictures/extra/nebraska_mohammadi2026_01.jpg",
   "thumb": "pictures/thumbs/nebraska_mohammadi2026_01.jpg",
   "kind": "in-use",
   "caption": "Robot feeding a guidewire into the modular artificial-artery testbed with stenosis",
   "source_key": "mohammadi2026_haptic",
   "source": "Mohammadi et al., Annals of Biomedical Engineering (2026), Fig. 9(b), p. 11",
   "source_url": "https://doi.org/10.1007/s10439-026-04315-2",
   "where": "Fig. 9(b), p. 11",
   "licence": "CC BY 4.0"
  },
  {
   "id": "nebraska_mohammadi2026_02",
   "status": "cleared",
   "file": "pictures/extra/nebraska_mohammadi2026_02.jpg",
   "thumb": "pictures/thumbs/nebraska_mohammadi2026_02.jpg",
   "kind": "render",
   "caption": "CAD overview: main Dynamixel servo, three miniature servos, belt and wire guide",
   "source_key": "mohammadi2026_haptic",
   "source": "Mohammadi et al., Annals of Biomedical Engineering (2026), Fig. 3(a), p. 5",
   "source_url": "https://doi.org/10.1007/s10439-026-04315-2",
   "where": "Fig. 3(a), p. 5",
   "licence": "CC BY 4.0"
  },
  {
   "id": "nebraska_mohammadi2026_03",
   "status": "cleared",
   "file": "pictures/extra/nebraska_mohammadi2026_03.jpg",
   "thumb": "pictures/thumbs/nebraska_mohammadi2026_03.jpg",
   "kind": "close-up",
   "caption": "CAD of the gripper-belt advance cycle and rack-and-pinion rotation",
   "source_key": "mohammadi2026_haptic",
   "source": "Mohammadi et al., Annals of Biomedical Engineering (2026), Fig. 2(a)-(d), p. 4",
   "source_url": "https://doi.org/10.1007/s10439-026-04315-2",
   "where": "Fig. 2(a)-(d), p. 4",
   "licence": "CC BY 4.0"
  },
  {
   "id": "nebraska_mohammadi2026_04",
   "status": "cleared",
   "file": "pictures/extra/nebraska_mohammadi2026_04.jpg",
   "thumb": "pictures/thumbs/nebraska_mohammadi2026_04.jpg",
   "kind": "render",
   "caption": "CAD of the assembled testbed rack with the robot installed",
   "source_key": "mohammadi2026_haptic",
   "source": "Mohammadi et al., Annals of Biomedical Engineering (2026), Fig. 7(c), p. 9",
   "source_url": "https://doi.org/10.1007/s10439-026-04315-2",
   "where": "Fig. 7(c), p. 9",
   "licence": "CC BY 4.0"
  }
 ],
 "sources": [
  {
   "key": "mohammadi2026_haptic",
   "doi": "10.1007/s10439-026-04315-2",
   "url": null,
   "title": "Haptic Portable Robotic Device for Automated Guidewire or Catheter Navigation in Endovascular Procedures",
   "year": 2026,
   "venue": "Annals of Biomedical Engineering"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full",
 "notes": "Autonomous rule-based controller, not learned.",