Catheter Robot Atlas

Magnetic navigation · marc_losey2014

MARC catheter

MARC (magnetically assisted remote-controlled) catheter steered by tip-coil current in the MRI field
Hetts, Department of Radiology and Biomedical Imaging, UCSF · 2013–2016
MARC catheter: main picture1 / 3
Not cleared for distribution. li2020_magnetic Fig. 8e(B), reproduced there from losey2014_magnetically Fig. 1B. Licence: publisher copyright. Picture grade C.
Family
Magnetic navigation (magnetic-navigation)
Devices
own MARC microcatheters: 2 mm with solenoid coil at the tip (losey2014_magnetically abstract); 1.6 mm with double-saddle coils (moftakhar2015_new abstract)
Degrees of freedom
tip deflection in 1 axis (2014) or 2 axes (2015); advance and twist by hand
Push
Magnetic
Twist
Field direction
Tested up to
Animal
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Animal
What was tested
3 farm pigs (43 +/- 2 kg): MR-guided MARC renal artery embolization with the contralateral kidney under x-ray; embolization successful in all 6 kidneys (lillaney2016_endovascular abstract); in vitro cryogel abdominal aortic phantom: 192/240 (80%) turns with magnetic assistance vs 144/240 (60%) without (losey2014_magnetically abstract)
Key numbers
tip current 300 mA (losey2014_magnetically); MR-guided catheterization 93 +/- 56 s vs 60 +/- 22 s under x-ray (lillaney2016_endovascular); inexperienced operators 67% (1.5 T) and 75% (3 T) successful turns vs 76% under x-ray (moftakhar2015_new)

Mechanism

Push (translation)
magnetic (Lorentz torque on the current-carrying tip coil in the MRI main field); insertion by hand
Twist (rotation)
tip deflection set by coil current direction and magnitude
Device range
catheter OD 2 mm (2014), 1.6 mm (2015)
Target anatomy
interventional MRI: abdominal aortic branch catheterization, renal artery embolization

Sensing and interface

Force sensing
none
Leader and haptics
no: foot pedal (losey2014_magnetically) or in-room cart (lillaney2016_endovascular)
Imaging
real-time MRI at 1.5 T (bSSFP) and 3 T; compared with x-ray fluoroscopy
Sterility
N.R.

Size

Footprint (mm)
N.R. (clinical MRI scanner)
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
2013: microcatheter tip deflection protocol (hetts2013_magnetically, JoVE); 2014: 2 mm catheter with a solenoid tip coil, foot pedal delivers +/-300 mA (losey2014_magnetically); 2015: 1.6 mm laser-lithographed dual-axis saddle coils, 1.5 T and 3 T (moftakhar2015_new); 2016: intramural copper-braided catheter with a saddle coil, in-room current cart (lillaney2016_endovascular)

Sources

  1. Magnetically Assisted Remote-controlled Endovascular Catheter for Interventional MR Imaging: In Vitro Navigation at 1.5 T versus X-ray Fluoroscopy · Radiology, 2014 · doi:10.1148/radiol.14132041 [losey2014_magnetically]
  2. Endovascular MR-guided Renal Embolization by Using a Magnetically Assisted Remote-controlled Catheter System · Radiology, 2016 · doi:10.1148/radiol.2016152036 [lillaney2016_endovascular]
  3. New-Generation Laser-lithographed Dual-Axis Magnetically Assisted Remote-controlled Endovascular Catheter for Interventional MR Imaging: In Vitro Multiplanar Navigation at 1.5 T and 3 T versus X-ray Fluoroscopy · Radiology, 2015 · doi:10.1148/radiol.2015142648 [moftakhar2015_new]
  4. Magnetically-Assisted Remote Controlled Microcatheter Tip Deflection under Magnetic Resonance Imaging · Journal of Visualized Experiments, 2013 · doi:10.3791/50299 [hetts2013_magnetically]
  5. A review of magnetic actuation systems and magnetically actuated guidewire- and catheter-based microrobots for vascular interventions · Intelligent Service Robotics, 2020 · doi:10.1007/s11370-020-00311-0 [li2020_magnetic]

Notes

Notes
Family: magnetic. Steering by current in the tip coil, insertion by hand; no bedside drive. No photo of the whole set-up (MRI room, current cart) in the four papers' figures; the picture is the catheter itself.

This record

Read level
abstract — abstract + figures (PMC author manuscripts of losey2014, lillaney2016, moftakhar2015; hetts2013)
Picture
li2020_magnetic Fig. 8e(B), reproduced there from losey2014_magnetically Fig. 1B ↗ — licence: publisher copyright; grade C
Picture changes
native extract from li2020_magnetic_ISR.pdf p. 9; cropped to panel B (panel letter removed); '2 mm' and '10 mm' labels kept; none upscaled
Found by
actuated-tip: Search of actuated-tip and magnetic devices.
Data
JSON · YAML: records/robots/marc_losey2014.yaml
Correction
Suggest a correction [repository not created yet]