 "schema_version": 1,
 "id": "marc_losey2014",
 "name": "MARC catheter",
 "full_name": "MARC (magnetically assisted remote-controlled) catheter steered by tip-coil current in the MRI field",
 "group": "Hetts, Department of Radiology and Biomedical Imaging, UCSF",
 "group_short": "UCSF",
 "family": "magnetic",
 "category": "magnetic-navigation",
 "first_year": 2013,
 "latest_year": 2016,
 "versions": "2013: microcatheter tip deflection protocol (hetts2013_magnetically, JoVE); 2014: 2 mm catheter with a solenoid tip coil, foot pedal delivers +/-300 mA (losey2014_magnetically); 2015: 1.6 mm laser-lithographed dual-axis saddle coils, 1.5 T and 3 T (moftakhar2015_new); 2016: intramural copper-braided catheter with a saddle coil, in-room current cart (lillaney2016_endovascular)",
 "target_anatomy": "interventional MRI: abdominal aortic branch catheterization, renal artery embolization",
 "devices": [
  "microcatheter"
 ],
 "devices_driven": "own MARC microcatheters: 2 mm with solenoid coil at the tip (losey2014_magnetically abstract); 1.6 mm with double-saddle coils (moftakhar2015_new abstract)",
 "n_dof": null,
 "n_dof_detail": "tip deflection in 1 axis (2014) or 2 axes (2015); advance and twist by hand",
 "translation_mechanism": "magnetic",
 "translation_detail": "magnetic (Lorentz torque on the current-carrying tip coil in the MRI main field); insertion by hand",
 "rotation_mechanism": "magnetic-field",
 "rotation_detail": "tip deflection set by coil current direction and magnitude",
 "device_range": "catheter OD 2 mm (2014), 1.6 mm (2015)",
 "footprint_mm": "N.R. (clinical MRI scanner)",
 "mass_kg": "N.R.",
 "force_sensing": "none",
 "haptics_leader": "no: foot pedal (losey2014_magnetically) or in-room cart (lillaney2016_endovascular)",
 "imaging_compat": "real-time MRI at 1.5 T (bSSFP) and 3 T; compared with x-ray fluoroscopy",
 "sterility": "N.R.",
 "highest_validation": "animal",
 "validation_detail": "3 farm pigs (43 +/- 2 kg): MR-guided MARC renal artery embolization with the contralateral kidney under x-ray; embolization successful in all 6 kidneys (lillaney2016_endovascular abstract); in vitro cryogel abdominal aortic phantom: 192/240 (80%) turns with magnetic assistance vs 144/240 (60%) without (losey2014_magnetically abstract)",
 "key_numbers": "tip current 300 mA (losey2014_magnetically); MR-guided catheterization 93 +/- 56 s vs 60 +/- 22 s under x-ray (lillaney2016_endovascular); inexperienced operators 67% (1.5 T) and 75% (3 T) successful turns vs 76% under x-ray (moftakhar2015_new)",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "C",
  "kind": "photo",
  "source": "li2020_magnetic Fig. 8e(B), reproduced there from losey2014_magnetically Fig. 1B",
  "source_url": "https://doi.org/10.1007/s11370-020-00311-0",
  "licence": "publisher copyright",
  "credit": "Hwang et al., Intelligent Service Robotics (2020), Fig. 8e. publisher copyright.",
  "edits": "native extract from li2020_magnetic_ISR.pdf p. 9; cropped to panel B (panel letter removed); '2 mm' and '10 mm' labels kept; none upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "marc_losey2014_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "MARC tip schematic, prototype and MR images of deflection",
   "source_key": "losey2014_magnetically",
   "source": "Losey et al., Radiology (2014), Fig. 1, PMC4065602",
   "source_url": "https://doi.org/10.1148/radiol.14132041",
   "where": "Fig. 1, PMC4065602",
   "licence": "publisher copyright"
  },
  {
   "id": "marc_losey2014_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Laser-lithographed coil on the microcatheter tip",
   "source_key": "hetts2013_magnetically",
   "source": "Hetts et al., Journal of Visualized Experiments (2013), Fig. 5, PMC3643177",
   "source_url": "https://doi.org/10.3791/50299",
   "where": "Fig. 5, PMC3643177",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "losey2014_magnetically",
   "doi": "10.1148/radiol.14132041",
   "url": null,
   "title": "Magnetically Assisted Remote-controlled Endovascular Catheter for Interventional MR Imaging: In Vitro Navigation at 1.5 T versus X-ray Fluoroscopy",
   "year": 2014,
   "venue": "Radiology"
  },
  {
   "key": "lillaney2016_endovascular",
   "doi": "10.1148/radiol.2016152036",
   "url": null,
   "title": "Endovascular MR-guided Renal Embolization by Using a Magnetically Assisted Remote-controlled Catheter System",
   "year": 2016,
   "venue": "Radiology"
  },
  {
   "key": "moftakhar2015_new",
   "doi": "10.1148/radiol.2015142648",
   "url": null,
   "title": "New-Generation Laser-lithographed Dual-Axis Magnetically Assisted Remote-controlled Endovascular Catheter for Interventional MR Imaging: In Vitro Multiplanar Navigation at 1.5 T and 3 T versus X-ray Fluoroscopy",
   "year": 2015,
   "venue": "Radiology"
  },
  {
   "key": "hetts2013_magnetically",
   "doi": "10.3791/50299",
   "url": null,
   "title": "Magnetically-Assisted Remote Controlled Microcatheter Tip Deflection under Magnetic Resonance Imaging",
   "year": 2013,
   "venue": "Journal of Visualized Experiments"
  },
  {
   "key": "li2020_magnetic",
   "doi": "10.1007/s11370-020-00311-0",
   "url": null,
   "title": "A review of magnetic actuation systems and magnetically actuated guidewire- and catheter-based microrobots for vascular interventions",
   "year": 2020,
   "venue": "Intelligent Service Robotics"
  }
 ],
 "read_level": "abstract",
 "read_level_detail": "abstract + figures (PMC author manuscripts of losey2014, lillaney2016, moftakhar2015; hetts2013)",
 "notes": "Family: magnetic. Steering by current in the tip coil, insertion by hand; no bedside drive. No photo of the whole set-up (MRI room, current cart) in the four papers' figures; the picture is the catheter itself.",
 "maintainers_own": false,
 "found_by": "actuated-tip"
}