Catheter Robot Atlas

Actuated-tip devices · growing_mangan2025

Growing robot

Hybrid tendon-notched continuum robot + hydraulic everting (growing) robot
Morimoto (UCSD) with Hawkes (UCSB), Heit (Stanford) · 2025
Growing robot: main picture1 / 5
Not cleared for distribution. mangan2025_serially Fig. 5 (CAD of the watertight actuation subsystem). Licence: publisher copyright (IEEE). Picture grade C.
Family
Actuated-tip devices (actuated-device)
Devices
own growing robot (2.67 mm OD, about 8 Fr) that pulls a commercial 5 Fr aspiration catheter (SOFIA); notched base 7.6 mm OD
Degrees of freedom
notched base two segments, tendon-driven by 2 capstans; growth by fluid pressure; system roll and insertion
Push
Other drives
Twist
Rotating drive unit
Tested up to
Phantom
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Phantom
What was tested
silicone arch and ICA models with two 360-degree loops of 7.3 mm radius; one full deployment in under 10 min with mock clot (Sec. VI-D, S8); no repeated-trial success rate found
Key numbers
growth pressure 35-40 kPa; housing withstands 85 kPa; steering error 9.9 +/- 1.2 % and 17.6 +/- 0.7 % of segment length (S8 citing Sec. VI-A); retraction buckles

Mechanism

Push (translation)
other: eversion by hydraulic pressure (35-40 kPa) plus linear actuator and friction drive for the aspiration catheter
Twist (rotation)
system roll stage
Device range
growing robot length 445 mm (S8 citing mangan Sec. III-C)
Target anatomy
neuro: type III aortic arch and extremely tortuous ICA benchtop models (mechanical thrombectomy)

Sensing and interface

Force sensing
pressure sensor on growth fluid; N.R. for tip force
Leader and haptics
no
Imaging
N.R.
Sterility
N.R.

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
none found in text scanned (kinematics of notched robot and design optimisation reported; no simulator)
Sim-trained controller
none found

Versions

Lineage
single (T-MRB 2025); 2021: VINE everting (growing) catheter, UCSB/UCSD/Stanford, same team (Li 2021 T-MRB) (Alekseeva 2025 Table 12 row 9, Fig. 19g) (alekseeva2025_design)

Sources

  1. Serially-Connected Soft Continuum Robots for Endovascular Emergencies · IEEE Transactions on Medical Robotics and Bionics, 2025 · doi:10.1109/tmrb.2025.3583160 [mangan2025_serially]
  2. The Design of Endovascular Robots for Cardiac Interventions: A Survey · Journal of Field Robotics, 2024 · doi:10.1002/rob.22480 [alekseeva2025_design]

Notes

Notes
Affiliations are UCSD/UCSB/Stanford; brief mentioned UCSD/Stanford.

This record

Read level
full — full text downloaded; abstract, intro, hardware sections read, results via S8
Picture
mangan2025_serially Fig. 5 (CAD of the watertight actuation subsystem) ↗ — licence: publisher copyright (IEEE); grade C
Picture changes
native extract, no edits (labels and leader lines kept: they cross the motors, capstans and housing)
Found by
actuated-tip: Search of actuated-tip and magnetic devices.
Data
JSON · YAML: records/robots/growing_mangan2025.yaml
Correction