Catheter Robot Atlas

Academic bedside robots · dupont_teer_pistorius2025

TEER robot

Boston Children's robotic drive for transcatheter edge-to-edge mitral repair (MitraClip cartridges on a universal catheter drive)
Boston Children's Hospital / Harvard Medical School (P. E. Dupont; L. Pistorius, N. U. Nayar, P. Tran) with UCSF (S. Elmariah) · 2025–2026
TEER robot: main picture1 / 5
Not cleared for distribution. pistorius2026_teer_RAL Fig. 4(b) robotic clip delivery system (drive system with three cartridges). Licence: publisher copyright (IEEE; NIH author manuscript PMC13271658). Picture grade C.
Family
Academic bedside robots (academic)
Devices
Abbott MitraClip NTR through an Edwards PASCAL Precision transseptal sheath; manual handles replaced by transmission cartridges (Sec. II)
Degrees of freedom
8 used (transseptal sheath 3, intermediate sheath 3, device sheath 2); drive supports up to four telescoping sheaths with up to four motorised inputs each
Push
Linear stage
Twist
Handle or knob
Tested up to
Phantom
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Phantom
What was tested
silicone mitral model (United Biologics); one operator; 5 manual and 5 robotic trials per step test; 30 manual and 30 robotic clip deliveries, 10 per segment A1P1, A2P2, A3P3 (Sec. IV)
Key numbers
segment 2 robotic time 52% of manual; mean clip placement error from 3 mm to below 1 mm for segments 2 and 3; inadvertent clip rotation 28 +- 4 deg manual vs 6 +- 7 deg robotic; speed limits 5.46 deg/s flexure, 14.56 deg/s roll, 6 mm/s translation

Mechanism

Push (translation)
linear-stage+clamp (device-agnostic carts translate each cartridge)
Twist (rotation)
motor-driven worm gears for bending and rotation; clamp on the handle for device-sheath roll
Device range
procedure-specific cartridges on device-agnostic carts
Target anatomy
mitral valve (TEER)

Sensing and interface

Force sensing
none reported
Leader and haptics
game controller (haptic device input supported by the ROS 2 software)
Imaging
N.R.
Sterility
N.R.

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
2025 arXiv; 2026 RAL; the universal drive itself is announced for a separate letter

Sources

  1. Using Robotics to Improve Transcatheter Edge-to-Edge Repair of the Mitral Valve · IEEE Robotics and Automation Letters, 2026 · doi:10.1109/lra.2026.3698357 [pistorius2026_using]

Notes

Notes
Distinct from the Dupont autonomous intracardiac catheter (robots_A3 dupont_fagogenis2019).

This record

Read level
full
Picture
pistorius2026_teer_RAL Fig. 4(b) robotic clip delivery system (drive system with three cartridges) ↗ — licence: publisher copyright (IEEE; NIH author manuscript PMC13271658); grade C
Picture changes
native extract (2050x1926); cropped above the label arrow, so the wheels are cut at the bottom; not upscaled
Found by
gap: Gap searches for robots the other searches missed.
Data
JSON · YAML: records/robots/dupont_teer_pistorius2025.yaml
Correction
Suggest a correction [repository not created yet]