CSTAR ablation robot

CSTAR robotics-assisted ablation-catheter manipulation system (pull-wire drive with tension sensing and model-based contact-force estimation)
Western University / CSTAR (R. V. Patel; M. Khoshnam, I. Khalaji) · 2014–2015
CSTAR ablation robot: main picture1 / 4
Not cleared for distribution. khoshnam2015_ablation_IROS Fig. 1 (CAD of set-up), left part: motorized linear stage, handle clamp, pull-wire force sensor. Licence: publisher copyright (IEEE). Picture grade B.
Family
Academic bedside robots (academic)
Devices
one 7 Fr unidirectional steerable ablation catheter (Biosense Webster); handle cover removed, prismatic knob fixed to the stage slider (IROS 2015 Sec. II)
Degrees of freedom
1 actuated (pull-wire flexion); catheter body free in the sheath
Push
Other drives
Twist
None
Tested up to
Bench
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Bench
What was tested
artificial tissue on a Nano17 sensor moved by a Zaber T-LSM050A stage, including 8.5 mm sinusoidal wall motion; three cases in Table II (IROS 2015 Sec. IV)
Key numbers
tip placed within +-1 mm of target; contact-force estimation error below 3 gf, maximum 2.64 gf (IROS 2015 abstract, Sec. V)

Mechanism

Push (translation)
other: body insertion not actuated; pull-wire driven by a Zaber T-LSR300B motorised linear stage
Twist (rotation)
none (not actuated)
Device range
one catheter model
Target anatomy
cardiac ablation (atria)

Sensing and interface

Force sensing
ATI Nano43 measures pull-wire tension; tip-tissue force estimated from tension plus tip pose (no tip sensor); ATI Nano17 under the tissue as ground truth (IROS 2015 Sec. II)
Leader and haptics
none; GUI displays real-time contact-force estimate
Imaging
camera tracking of tip markers (640 x 480 px, 30 fps) as fluoroscopy stand-in
Sterility
N.R.

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
2014 BioRob robotics-assisted catheter manipulation for ablation efficiency (title); 2015 IROS system with real-time force estimation (full); 3-DOF manipulator named as ongoing work (IROS 2015 Sec. V)

Sources

  1. A robotics-assisted catheter manipulation system for cardiac ablation with real-time force estimation · 2015 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS), 2015 · doi:10.1109/iros.2015.7353821 [khoshnam2015_robotics]
  2. Robotics-assisted catheter manipulation for improving cardiac ablation efficiency · 5th IEEE RAS/EMBS International Conference on Biomedical Robotics and Biomechatronics, 2014 · doi:10.1109/biorob.2014.6913794 [khoshnam2014_robotics]

Notes

Notes
Boundary case: actuates only tip flexion of a conventional ablation catheter (compare robots_A3 toronto_nazari2024 EP-handle robot).

This record

Read level
full — full (IROS 2015); title (BioRob 2014)
Picture
khoshnam2015_ablation_IROS Fig. 1 (CAD of set-up), left part: motorized linear stage, handle clamp, pull-wire force sensor ↗ — licence: publisher copyright (IEEE); grade B
Picture changes
native extract (3828x1769); cropped to the pull-wire drive; 3 leader arrows on white background erased (LaMa, imgprep erase --box/--line), robot untouched; not upscaled
Found by
gap: Gap searches for robots the other searches missed.
Data
JSON · YAML: records/robots/cstar_ablation_khoshnam2014.yaml
Correction
Suggest a correction [repository not created yet]