 "schema_version": 1,
 "id": "cstar_ablation_khoshnam2014",
 "name": "CSTAR ablation robot",
 "full_name": "CSTAR robotics-assisted ablation-catheter manipulation system (pull-wire drive with tension sensing and model-based contact-force estimation)",
 "group": "Western University / CSTAR (R. V. Patel; M. Khoshnam, I. Khalaji)",
 "group_short": "Western / CSTAR",
 "family": "academic",
 "category": "academic",
 "first_year": 2014,
 "latest_year": 2015,
 "versions": "2014 BioRob robotics-assisted catheter manipulation for ablation efficiency (title); 2015 IROS system with real-time force estimation (full); 3-DOF manipulator named as ongoing work (IROS 2015 Sec. V)",
 "target_anatomy": "cardiac ablation (atria)",
 "devices": [
  "catheter"
 ],
 "devices_driven": "one 7 Fr unidirectional steerable ablation catheter (Biosense Webster); handle cover removed, prismatic knob fixed to the stage slider (IROS 2015 Sec. II)",
 "n_dof": 1,
 "n_dof_detail": "1 actuated (pull-wire flexion); catheter body free in the sheath",
 "translation_mechanism": "other",
 "translation_detail": "other: body insertion not actuated; pull-wire driven by a Zaber T-LSR300B motorised linear stage",
 "rotation_mechanism": "none",
 "rotation_detail": "none (not actuated)",
 "device_range": "one catheter model",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "ATI Nano43 measures pull-wire tension; tip-tissue force estimated from tension plus tip pose (no tip sensor); ATI Nano17 under the tissue as ground truth (IROS 2015 Sec. II)",
 "haptics_leader": "none; GUI displays real-time contact-force estimate",
 "imaging_compat": "camera tracking of tip markers (640 x 480 px, 30 fps) as fluoroscopy stand-in",
 "sterility": "N.R.",
 "highest_validation": "bench",
 "validation_detail": "artificial tissue on a Nano17 sensor moved by a Zaber T-LSM050A stage, including 8.5 mm sinusoidal wall motion; three cases in Table II (IROS 2015 Sec. IV)",
 "key_numbers": "tip placed within +-1 mm of target; contact-force estimation error below 3 gf, maximum 2.64 gf (IROS 2015 abstract, Sec. V)",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "B",
  "kind": "render",
  "source": "khoshnam2015_ablation_IROS Fig. 1 (CAD of set-up), left part: motorized linear stage, handle clamp, pull-wire force sensor",
  "source_url": "https://doi.org/10.1109/iros.2015.7353821",
  "licence": "publisher copyright (IEEE)",
  "credit": "Khoshnam et al., 2015 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (2015), Fig. 1. publisher copyright (IEEE).",
  "edits": "native extract (3828x1769); cropped to the pull-wire drive; 3 leader arrows on white background erased (LaMa, imgprep erase --box/--line), robot untouched; not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "cstar_ablation_khoshnam2014_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Proximal actuator: motorised stage pulling the catheter knob through a force sensor",
   "source_key": "khoshnam2015_robotics",
   "source": "Khoshnam et al., 2015 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (2015), Fig. 2, p. 2",
   "source_url": "https://doi.org/10.1109/iros.2015.7353821",
   "where": "Fig. 2, p. 2",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "cstar_ablation_khoshnam2014_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Catheter tip against a force sensor under a side camera, linear stage behind",
   "source_key": "khoshnam2015_robotics",
   "source": "Khoshnam et al., 2015 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (2015), Fig. 3, p. 2",
   "source_url": "https://doi.org/10.1109/iros.2015.7353821",
   "where": "Fig. 3, p. 2",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "cstar_ablation_khoshnam2014_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "CAD of the 2014 set-up: handle actuation stage, camera and tip force sensor",
   "source_key": "khoshnam2014_robotics",
   "source": "Khoshnam et al., 5th IEEE RAS/EMBS International Conference on Biomedical Robotics and Biomechatronics (2014), Fig. 1, p. 2",
   "source_url": "https://doi.org/10.1109/biorob.2014.6913794",
   "where": "Fig. 1, p. 2",
   "licence": "publisher copyright (IEEE)"
  }
 ],
 "sources": [
  {
   "key": "khoshnam2015_robotics",
   "doi": "10.1109/iros.2015.7353821",
   "url": null,
   "title": "A robotics-assisted catheter manipulation system for cardiac ablation with real-time force estimation",
   "year": 2015,
   "venue": "2015 IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS)"
  },
  {
   "key": "khoshnam2014_robotics",
   "doi": "10.1109/biorob.2014.6913794",
   "url": null,
   "title": "Robotics-assisted catheter manipulation for improving cardiac ablation efficiency",
   "year": 2014,
   "venue": "5th IEEE RAS/EMBS International Conference on Biomedical Robotics and Biomechatronics"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (IROS 2015); title (BioRob 2014)",
 "notes": "Boundary case: actuates only tip flexion of a conventional ablation catheter (compare robots_A3 toronto_nazari2024 EP-handle robot).",
 "maintainers_own": false,
 "found_by": "gap"
}