Catheter Robot Atlas

Commercial and clinical-stage systems · cgci_magnetecs2011

CGCI

CGCI (Catheter Guidance Control and Imaging) remote magnetic navigation system
Magnetecs Corp. · 2011–2013
CGCI: main picture1 / 3
Not cleared for distribution. li2020_magnetic Fig. 5 (Hwang 2020; reproduced there from Moya 2013 Rev Esp Cardiol). Licence: publisher copyright. Picture grade B.
Family
Commercial and clinical-stage systems (magnetic-navigation)
Devices
magnetized irrigated gold-tip EP ablation catheter (MedFact) with three permanent-magnet pellets at the tip, through an Agilis ES sheath (filgueirasrama2013_remote Protocol 1, Fig. 1); 4-mm-tip ablation catheter in pigs (gang2011_dynamically abstract); catheter 7 Fr / 2.2 mm (jeong2017_feasibility Introduction)
Degrees of freedom
3 (mandhumula2025_robotic Table 1; alekseeva2025_design Table 14); field shaped by 8 coil-core electromagnets, push-pull by field gradient (filgueirasrama2013_remote Representative Results) plus a motorized linear catheter advancer
Push
Magnetic
Twist
Field direction
Tested up to
Human
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Human
What was tested
human: 5 atrial flutter ablations (3 left, 2 counterclockwise right), acute success 5/5, no procedure-derived complications, no recurrence at 3-11 months (filgueirasrama2013_remote Table 1); animal: 10 pigs, left-atrial mapping, linear and circumferential RF lesions, selective PV isolation in a subset (gang2011_dynamically abstract)
Key numbers
field up to 0.16 T (filgueirasrama2013_remote Protocol 1); 0.14 T and 0.7 T/m gradients (mandhumula2025_robotic text); pigs, automated mode: 96 +/- 8.4% reproducible, 1.9 +/- 0.4 mm from target, 11.6 +/- 3.5 s to target, lesion depth 78.5 +/- 12.1% of atrial wall (gang2011_dynamically abstract); five initial experimental procedures: 95.7% reproducible, 1.9 +/- 0.9 mm, 23.28 +/- 14.8 s (filgueirasrama2013_remote Representative Results)

Mechanism

Push (translation)
magnetic (field-gradient push-pull at the tip); motorized linear catheter advancement mechanism with a sterilizable wheel-drive gearbox on a motor base, joystick-controlled slack/advance (filgueirasrama2013_remote Protocol 1, Fig. 2A, 2C)
Twist (rotation)
tip direction by dynamically shaped field of 8 fixed electromagnets, 4 above and 4 below the patient in a semi-spherical chamber (li2020_magnetic Sec. 3.3; gang2011_dynamically abstract)
Device range
N.R.
Target anatomy
cardiac electrophysiology: right and left atrial flutter ablation (filgueirasrama2013_remote Table 1); pig left atrium mapping and RF lesions (gang2011_dynamically abstract)

Sensing and interface

Force sensing
none reported
Leader and haptics
no: right-hand 3D controller and left-hand 3-axis joystick at a remote operation console (filgueirasrama2013_remote Fig. 2B-C)
Imaging
X-ray fluoroscopy, EnSite NavX electroanatomic mapping, EP recording and ICE unified at the operation console (filgueirasrama2013_remote Protocol 1); recording and fluoroscopic equipment unaffected by the fields (gang2011_dynamically abstract)
Sterility
sterilizable wheel-drive gearbox, disposable sheath clip and leg mount on the advancer (filgueirasrama2013_remote Fig. 2A)

Size

Footprint (mm)
N.R. (semi-spherical chamber of 8 electromagnets around the torso, li2020_magnetic Fig. 5)
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
single system: animal validation in 10 pigs (gang2011_dynamically, abstract); initial human safety and feasibility protocol at La Paz University Hospital, Madrid, 5 flutter ablations (filgueirasrama2013_remote, Table 1); later discontinued (mandhumula2025_robotic: 'now seems to have been discontinued', 'not adopted commercially')

Sources

  1. Dynamically Shaped Magnetic Fields: Initial Animal Validation of a New Remote Electrophysiology Catheter Guidance and Control System · Circulation: Arrhythmia and Electrophysiology, 2011 · doi:10.1161/circep.110.959692 [gang2011_dynamically]
  2. Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures · Journal of Visualized Experiments, 2013 · doi:10.3791/3658 [filgueirasrama2013_remote]
  3. A review of magnetic actuation systems and magnetically actuated guidewire- and catheter-based microrobots for vascular interventions · Intelligent Service Robotics, 2020 · doi:10.1007/s11370-020-00311-0 [li2020_magnetic]
  4. Robotic Systems in Cardiovascular Interventions: Evolving Platforms and the Path Forward · Methodist DeBakey Cardiovascular Journal, 2025 · doi:10.14797/mdcvj.1652 [mandhumula2025_robotic]
  5. Robotic Systems Design in Endovascular Treatment · IEEE Transactions on Medical Robotics and Bionics, 2024 · doi:10.1109/tmrb.2024.3377374 [li2024_design]
  6. Current and Emerging Robot-Assisted Endovascular Catheterization Technologies: A Review · Annals of Biomedical Engineering, 2014 · doi:10.1007/s10439-013-0946-8 [rafiitari2014_current]
  7. The Design of Endovascular Robots for Cardiac Interventions: A Survey · Journal of Field Robotics, 2024 · doi:10.1002/rob.22480 [alekseeva2025_design]
  8. Feasibility study on magnetically steerable guidewire device for percutaneous coronary intervention · International Journal of Control, Automation and Systems, 2016 · doi:10.1007/s12555-015-0269-7 [jeong2017_feasibility]

Notes

Notes
Family: commercial (A4 scope; clinical-stage magnetic navigation). A4 had set it aside as EP-only and discontinued; the author put it in (round 8). data/snowball_candidates.csv files Gang 2011 under 'Niobe': it is the CGCI paper. Regulatory clearance N.R. in the sources read. Conflict: field maximum 0.16 T (filgueirasrama2013_remote) vs 0.14 T (mandhumula2025_robotic).

This record

Read level
full — full (filgueirasrama2013_remote, Europe PMC full text); abstract (gang2011_dynamically); review-only for status (li2020_magnetic, mandhumula2025_robotic)
Picture
li2020_magnetic Fig. 5 (Hwang 2020; reproduced there from Moya 2013 Rev Esp Cardiol) ↗ — licence: publisher copyright; grade B
Picture changes
native extract from p. 5; no crop; none upscaled
Found by
commercial: Search of commercial and clinical-stage systems.
Data
JSON · YAML: records/robots/cgci_magnetecs2011.yaml
Correction