Catheter Robot Atlas

Actuated-tip devices · bellows_barnes2025

Bellows guidewire

Soft fluidic bellows guidewire with teleoperated drive and imitation-learning policy
Krieger lab, Johns Hopkins, with Sochol (University of Maryland) · 2025
Bellows guidewire: main picture1 / 6
Not cleared for distribution. barnes2025_toward project page softrobotnavigation.github.io/resources/imgs/hardware_setup.png (middle panel: translation drive with bellows wire). Licence: project page image (authors), licence not stated. Picture grade B.
Family
Actuated-tip devices (actuated-device)
Devices
own 3D-printed fluid-driven soft robot, 5 mm bellows (large-scale analogue of a microguidewire, not a clinical size)
Degrees of freedom
2 (bending velocity, translation velocity)
Push
Tip-actuated
Twist
None
Tested up to
Phantom
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Phantom
What was tested
36 training mazes from 1584 possible, rearranged test set 3 mazes, novel test set 3 mazes; 647 demonstrations (218 normal, 429 recovery incl. DAgger)
Key numbers
83 % success on unseen layouts, 75 % on unseen patient-derived geometry (abstract); ablations 22 % to 61 % success; segmentation tip error 0.6 mm median, 1.9 mm 95th, 8.0 mm max

Mechanism

Push (translation)
tendon/tip-actuation
Twist (rotation)
none
Device range
N.R.
Target anatomy
neuro, large-scale (enlarged) 2D-projected bifurcation mazes

Sensing and interface

Force sensing
force sensor in leader handle only (teleoperation); none on follower
Leader and haptics
no force feedback; operator pushes a force-sensing handle, forces map proportionally to bend and translate velocity
Imaging
overhead camera converted to simulated fluoroscopy; no real X-ray
Sterility
N.R.

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
no physics model; only a fluoroscopy rendering of camera frames (Sec. III-B)
Sim-trained controller
none found: the policy trains on real teleoperated demonstrations with simulated fluoroscopy images; the paper contrasts itself with digital-twin-trained policies

Versions

Lineage
arXiv 2510.09497 v2 (July 2026), submitted to IEEE

Sources

  1. Towards Autonomous Soft Robotic Endovascular Navigation via Imitation Learning · link [barnes2025_toward]

Notes

Notes
Hardware is a large-scale analogue; quote the 5 mm bellows size next to any claim about neurovascular use.

This record

Read level
full
Picture
barnes2025_toward project page softrobotnavigation.github.io/resources/imgs/hardware_setup.png (middle panel: translation drive with bellows wire) ↗ — licence: project page image (authors), licence not stated; grade B
Picture changes
crop; labels "Fluidic tube", "Translation drive" and an arrow erased (LaMa, mask of dark text pixels) then near-white residue set to white
Found by
actuated-tip: Search of actuated-tip and magnetic devices.
Data
JSON · YAML: records/robots/bellows_barnes2025.yaml
Correction
Suggest a correction [repository not created yet]