SIAT belt grasper

SIAT second-generation robotic catheter system for cardiovascular intervention (belt-driven graspers)
Shenzhen Institutes of Advanced Technology, CAS (Lei Wang; Olatunji Mumini Omisore) · 2018
SIAT belt grasper: main picture1 / 3
Not cleared for distribution. omisore2018_towards Fig. 13 (experimental setup, robot outlined in red), TBCAS 2018, doi:10.1109/TBCAS.2018.2825359. Licence: publisher copyright. Picture grade C.
Family
Academic bedside robots (academic)
Devices
one 9 Fr, 130 cm unactuated catheter sheath (no steerable catheters)
Degrees of freedom
2 (axial translation, rotation)
Push
Belt pinch
Twist
Rotating clamp
Tested up to
Phantom
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Phantom
What was tested
Tubular PVC phantoms (10 mm) from the radial artery to the aortic arch; one insertion with compensation and none without (S8 reading)
Key numbers
Backlash gaps 3.94-9.30 mm between motor encoder and tip EM sensor; adaptive neuro-fuzzy compensation final error 0.4 mm (duan2023_progress, review statement); sensor accuracy 0.9 mm / 0.8 deg

Mechanism

Push (translation)
belt/track
Twist (rotation)
Rotary drive grasper on a Maxon EC16 motor; belt carries four axial drive graspers for continuous push (Maxon EC32)
Device range
9 Fr catheter sheath; two guiding sheaths of 35 mm length, 10 mm inner diameter at the belt ends
Target anatomy
radial artery through axillary and subclavian arteries to the aortic arch (tubular phantoms)

Sensing and interface

Force sensing
Strain gauges on the 9 Fr sheath, 5 mm and 25 mm from the tip (gauge factor 2.08, 120 ohm), Coulomb friction term subtracted; EM sensor (NDI Aurora) in the lumen for tip motion
Leader and haptics
none in this paper
Imaging
N.R.
Sterility
N.R.

Size

Footprint (mm)
400 x 180 x 150
Mass (kg)
about 9.5 (S8 evidence reading of Sec. II-C); battery 10,000 mAh, 220 +/- 27 min

Simulation

Simulation model
none found for the robot
Sim-trained controller
none found for this generation. The 2024 DRL PID-tuning study (omisore2024_sample, full) ran on the current-generation slave (Sec. II-A: "The current generation is an isomorphic master-slave RCS"; 57 x 22 x 16 cm, 4 DoF), not on this belt-driven robot: see row siat_haptic_duan2025.

Versions

Lineage
omisore2018_towards (TBCAS 2018, second generation); the same lab's third-generation isomorphic primary-secondary system (du2024_analyzing, omisore2024_sample, duan2025_development) is row siat_haptic_duan2025; 2011-14: SIAT first-generation master-slave robot (P. Zhang 2011 ICMA; Mi 2014 ICMA) (Alekseeva 2025 Table 2 row 5; Belikov 2024 Fig. 6 MIs; Zhao 2022 Table 1 row 15b) (alekseeva2025_design; belikov2024_evolution; zhao2022_remote)

Sources

  1. Towards Characterization and Adaptive Compensation of Backlash in a Novel Robotic Catheter System for Cardiovascular Interventions · IEEE Transactions on Biomedical Circuits and Systems, 2018 · doi:10.1109/tbcas.2018.2825359 [omisore2018_towards]
  2. Technical and Clinical Progress on Robot-Assisted Endovascular Interventions: A Review · Micromachines, 2023 · doi:10.3390/mi14010197 [duan2023_progress]
  3. The Design of Endovascular Robots for Cardiac Interventions: A Survey · Journal of Field Robotics, 2024 · doi:10.1002/rob.22480 [alekseeva2025_design]
  4. Evolution and current state of robotic catheters for endovascular surgery: A comprehensive review · Engineering Science and Technology, an International Journal, 2024 · doi:10.1016/j.jestch.2024.101789 [belikov2024_evolution]
  5. Remote vascular interventional surgery robotics: a literature review · Quantitative Imaging in Medicine and Surgery, 2022 · doi:10.21037/qims-21-792 [zhao2022_remote]

Notes

Notes
Fig. 1 of the paper is a CAD model; the setup photo was chosen as the prototype image. F4 2026-10-14: omisore2024_sample removed from this row (it describes the current, isomorphic generation).

This record

Read level
full — full (omisore2018 via S8 evidence + figures); omisore2024_sample full text read 2026-10-14 and moved to row siat_haptic_duan2025
Picture
omisore2018_towards Fig. 13 (experimental setup, robot outlined in red), TBCAS 2018, doi:10.1109/TBCAS.2018.2825359 ↗ — licence: publisher copyright; grade C
Picture changes
native extract (2050x1275 figure, soft source photo); cropped; red outline, red arrow, label box and a yellow leader erased (LaMa, colour mask + boxes); not upscaled
Found by
academic: Search of academic bedside robots.
Data
JSON · YAML: records/robots/siat_omisore2018.yaml
Correction
Suggest a correction [repository not created yet]