Magnetic navigation · polymtl_gosselin2011

MRN platform

MRI-based magnetic resonance navigation platform steering bead-tipped catheters and guidewires with added gradient coils
Martel, NanoRobotics Laboratory, Polytechnique Montreal (with Soulez, CHUM, for the animal study) · 2011–2015
MRN platform: main picture1 / 4
Not cleared for distribution. vonthron2011_mri Fig. 2. Licence: publisher copyright. Picture grade C.
Family
Magnetic navigation (magnetic-navigation)
Devices
2.5 Fr catheter (FasTracker 018, Boston Scientific) with two 0.9 mm chrome-steel beads 1.6 mm apart (vonthron2011_mri Sec. VII-A); guidewire with a magnetic bead at its tip (lalande2015_vivo abstract)
Degrees of freedom
3D gradient force on the tip bead (three propulsion gradient axes, vonthron2011_mri Sec. IV)
Push
Magnetic
Twist
Field direction
Tested up to
Animal
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Animal
What was tested
2 rabbits: selective catheterization of right and left renal arteries successful and reproducible, SMA and iliac feasible, no inadvertent side-branch entry when advanced without actuation (lalande2015_vivo abstract); catheter deflection in a water bath for gradients -400 to +400 mT/m in Maxwell coils and in the propulsion insert (vonthron2011_mri Sec. VII-A, Figs. 5-7)
Key numbers
propulsion insert up to 450 mT/m in three directions; minimum coil rise time 20 ms; clinical imaging gradients about 20 mT/m effective (40 mT/m nominal) on a Siemens Sonata (vonthron2011_mri Sec. III-IV)

Mechanism

Push (translation)
magnetic (gradient-force steering of the tip); advance N.R. in the abstract (manual per li2020_magnetic Sec. 4.1: 'blindly controlled' between tracking sequences)
Twist (rotation)
gradient direction of the added coils
Device range
N.R.
Target anatomy
selective catheterization of aortic branches (renal, superior mesenteric and iliac arteries) in rabbits (lalande2015_vivo abstract)

Sensing and interface

Force sensing
none
Leader and haptics
no
Imaging
MRI tracking sequence interleaved with propulsion (imaging and propulsion mutually exclusive, vonthron2011_mri Fig. 3); MRA roadmap and dedicated tracking (lalande2015_vivo abstract)
Sterility
N.R.

Size

Footprint (mm)
N.R. (coil insert inside a clinical 1.5-T bore, vonthron2011_mri Fig. 2)
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
2011: catheter with ferromagnetic spheres steered by MRI gradients, deflection characterised (gosselin2011_characterization, review-only); MRN platform with a propulsion and imaging coil insert in a 1.5-T Siemens Sonata (vonthron2011_mri); 2015: guidewire with a ferromagnetic bead steered in 2 rabbits (lalande2015_vivo, abstract)

Sources

  1. A MRI-based integrated platform for the navigation of micro-devices and microrobots · 2011 IEEE/RSJ International Conference on Intelligent Robots and Systems, 2011 · doi:10.1109/iros.2011.6094721 [vonthron2011_mri]
  2. In vivo demonstration of magnetic guidewire steerability in a MRI system with additional gradient coils · Medical Physics, 2015 · doi:10.1118/1.4906194 [lalande2015_vivo]
  3. Characterization of the deflections of a catheter steered using a magnetic resonance imaging system · Medical Physics, 2011 · doi:10.1118/1.3622599 [gosselin2011_characterization]
  4. A review of magnetic actuation systems and magnetically actuated guidewire- and catheter-based microrobots for vascular interventions · Intelligent Service Robotics, 2020 · doi:10.1007/s11370-020-00311-0 [li2020_magnetic]

Notes

Notes
Family: magnetic. The group's untethered MRN (1.5 mm sphere in a swine carotid) is out of scope. Lalande 2015 and Gosselin 2011 are Med. Phys. (Wiley): manual list; Lalande Fig. (in vivo setup) would be the better picture.

This record

Read level
full — full (vonthron2011_mri); abstract (lalande2015_vivo); review-only (gosselin2011_characterization via li2020_magnetic Sec. 4.2)
Picture
vonthron2011_mri Fig. 2 ↗ — licence: publisher copyright; grade C
Picture changes
native extract from vonthron2011_mri-platform_IROS.pdf p. 3; no crop; none upscaled
Found by
actuated-tip: Search of actuated-tip and magnetic devices.
Data
JSON · YAML: records/robots/polymtl_gosselin2011.yaml