Catheter Robot Atlas

Commercial and clinical-stage systems · panvisa_abrobo2024

PANVIS-A

Shenzhen Institute of Advanced Biomedical Robot (Abrobo, Aibo Hechuang) · 2023–2026
PANVIS-A: main picture1 / 7
Not cleared for distribution. zhang2026_safety Fig. 1(A), PMC image https://cdn.ncbi.nlm.nih.gov/pmc/blobs/cc91/13217118/7172c200f078/jnis-18-5-g001.jpg (article page https://pmc.ncbi.nlm.nih.gov/articles/PMC13217118/). Licence: CC BY-NC 4.0. Picture grade C.
Family
Commercial and clinical-stage systems (commercial)
Devices
one 0.035-in guidewire + one 5F angiographic / 6F guiding catheter
Degrees of freedom
2 per device
Push
Clamp relay
Twist
Rolling between rollers
Tested up to
Human
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Human
What was tested
RCT 64 robot vs 66 manual, 4 centres, NCT05778214 (zhang2026_safety)
Key numbers
success 100% both arms; operator dose 1.67+-3.49 vs 43.63+-38.95 uSv (zhang2026_safety abstract)

Mechanism

Push (translation)
clamp-relay
Twist (rotation)
clamps twist guidewire (thumb-index mimic)
Device range
N.R.
Target anatomy
neurovascular (cerebral angiography)

Sensing and interface

Force sensing
N.R.
Leader and haptics
no (master-slave manipulator)
Imaging
slave on DSA table
Sterility
N.R.

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
third generation per press reports

Sources

  1. Safety and efficacy of cerebral robot assisted angiography: randomized comparison of robotic versus manual procedures · Journal of NeuroInterventional Surgery, 2026 · doi:10.1136/jnis-2025-023412 [zhang2026_safety]
  2. A Reciprocating Delivery Device-Based Endovascular Intervention Robot With Multimanipulators Collaboration · IEEE Transactions on Instrumentation and Measurement, 2024 · doi:10.1109/tim.2023.3342853 [cao2024_reciprocating]

Notes

Notes
NMPA Class III 2024-08-26 (company release; yicai says Aibo Hechuang was the only interventional robot to get a certificate in 2024, indication angiography only, not treatment).

This record

Read level
full — full text on disk
Picture
zhang2026_safety Fig. 1(A), PMC image https://cdn.ncbi.nlm.nih.gov/pmc/blobs/cc91/13217118/7172c200f078/jnis-18-5-g001.jpg (article page https://pmc.ncbi.nlm.nih.gov/articles/PMC13217118/) ↗ — licence: CC BY-NC 4.0; grade C
Picture changes
cropped (0,0)-(305,192) of 684x390 (panel A, above the panel letter); not upscaled; no tone change; no erasing
Found by
commercial: Search of commercial and clinical-stage systems.
Data
JSON · YAML: records/robots/panvisa_abrobo2024.yaml
Correction