Catheter Robot Atlas

Magnetic navigation · msmsr_li2024

CMNS

Multi-section magnetic soft robot with collaborative multirobot navigation system (CMNS)
Xu, University of Macau · 2024
CMNS: main picture1 / 4
Li et al., Cyborg and Bionic Systems (2024), Fig. 1. CC BY 4.0 (Cyborg Bionic Syst.; PMC open-access metadata licence_code CC BY). Changes: native extract from the PMC open-data PDF; no crop, no edits; not upscaled. Source ↗ Picture grade A.
Family
Magnetic navigation (magnetic-navigation)
Devices
own telescoping two-tube soft robot: Tube-I OD 1.8 / ID 1.2 mm, Tube-II OD 3.5 / ID 2.5 mm
Degrees of freedom
UR3e carries a 30 mm NdFeB N42 permanent magnet; second manipulator carries ultrasound probe; seven-joint redundancy handled by QP controller
Push
Magnetic
Twist
Field direction
Tested up to
Phantom
Sim-to-real
Model-based only — Commands computed in a simulator or from a physical model ran on the robot; nothing was learned.

Testing

Highest level
Phantom
What was tested
multi-bifurcation intravascular phantom navigation; counts N.R.
Key numbers
shape error 3.1-3.9 deg, tip error about 1.0-1.4 mm (Table 2); bending range -125 to 125 deg tested

Mechanism

Push (translation)
magnetic (steering) + telescoping tube motion
Twist (rotation)
rotation of external magnet
Device range
N.R.
Target anatomy
intravascular phantoms with multiple bifurcations

Sensing and interface

Force sensing
none
Leader and haptics
no
Imaging
ultrasound tracking of the tip, closed loop
Sterility
N.R.

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
optimised Cosserat rod model with external heterogeneous field (full)
Sim-trained controller
model-based control using the Cosserat model on the hardware; no learned policy

Versions

Lineage
single (Cyborg Bionic Syst. 2024)

Sources

  1. Multi-Section Magnetic Soft Robot with Multirobot Navigation System for Vasculature Intervention · Cyborg and Bionic Systems, 2024 · doi:10.34133/cbsystems.0188 [li2024_multisection]

Notes

Notes
Picture can be cropped from the CC BY PDF once downloaded.

This record

Read level
full — full text from (no PDF)
Picture
li2024_multisection Fig. 1(B) embedded photo (CMNS with two UR arms, magnet, ultrasound probe and phantom; the figure's circles and labels are vector overlays and are not in the raster) ↗ — licence: CC BY 4.0 (Cyborg Bionic Syst.; PMC open-access metadata licence_code CC BY); grade A
Picture changes
native extract from the PMC open-data PDF; no crop, no edits; not upscaled
Found by
actuated-tip: Search of actuated-tip and magnetic devices.
Data
JSON · YAML: records/robots/msmsr_li2024.yaml
Correction