Catheter Robot Atlas

Magnetic navigation · magflow_pancaldi2025

MagFlow

MagFlow flow-driven magnetic microcatheter + OmniMag steering platform
Sakar, EPFL (with Mosimann, Toronto Western Hospital) · 2025
MagFlow: main picture1 / 7
Not cleared for distribution. EPFL news "Microcatheter delivers therapies to the tiniest blood vessels" (actu.epfl.ch/news/microcatheter-delivers-therapies-to-the-tiniest-bl), image 158504 original 7678x5121 (OmniMag on a UR arm over the light pad, stylus, pump). Licence: CC BY-SA 4.0 (EPFL/Alain Herzog). Picture grade A.
Family
Magnetic navigation (magnetic-navigation)
Devices
inflatable flat-tube microcatheter, head width 150-250 um; fed through a 2.7 Fr (0.7 mm lumen) delivery microcatheter inside a 5 Fr guide catheter
Degrees of freedom
OmniMag: spherical rotor with permanent magnet oriented by three orthogonal coils, mounted on a collaborative robot arm
Push
Other drives
Twist
Field direction
Tested up to
Animal
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Animal
What was tested
porcine in vivo under X-ray; six branches repeatedly steered in vertebral region; navigated 180 um arteries and a 500 um hairpin with 0.69 mm radius; superselective contrast and embolic infusion
Key numbers
OmniMag 3.7 mT at 2 A, 1.7 N m torque on the magnet; average field during steering 5.7 +/- 1.4 mT; arteries as small as 180 um; min curvature radius 0.69 mm

Mechanism

Push (translation)
other: blood-flow carried; no pushed shaft
Twist (rotation)
field direction
Device range
N.R.
Target anatomy
distal cerebral / spinal arteries, porcine model

Sensing and interface

Force sensing
none
Leader and haptics
no
Imaging
bi-plane fluoroscopy compatible
Sterility
N.R.

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
arXiv 2508.05312

Sources

  1. Flow-driven magnetic microcatheter for superselective arterial embolization · link [pancaldi2025_flowdriven]

Notes

Notes
Smallest device in my scope; there is no bedside push drive.

This record

Read level
full — full (abstract, platform and in vivo sections)
Picture
EPFL news "Microcatheter delivers therapies to the tiniest blood vessels" (actu.epfl.ch/news/microcatheter-delivers-therapies-to-the-tiniest-bl), image 158504 original 7678x5121 (OmniMag on a UR arm over the light pad, stylus, pump) — licence: CC BY-SA 4.0 (EPFL/Alain Herzog); grade A
Picture changes
crop removing the operators faces, downscaled 50 %
Found by
actuated-tip: Search of actuated-tip and magnetic devices.
Data
JSON · YAML: records/robots/magflow_pancaldi2025.yaml
Correction
Suggest a correction [repository not created yet]