Catheter Robot Atlas

Academic bedside robots · hit_fu2011

HIT catheter robot

HIT master-slave robotic catheter system for a steerable pull-wire catheter
Harbin Institute of Technology, State Key Laboratory of Robotics and System (Yili Fu, Anzhu Gao, Hao Liu, Kai Li, Zhaoguang Liang) with Harbin Medical University · 2011
HIT catheter robot: main picture1 / 3
Not cleared for distribution. fu2011_development Fig. 7 handle manipulation mechanism (native scan raster p. 3). Licence: publisher copyright. Picture grade C.
Family
Academic bedside robots (academic)
Devices
custom 7 Fr single-curve pull-wire steerable catheter with two 5-DOF magnetic tracking sensors, two electrodes and a thermocouple (fu2011_development Sec. III)
Degrees of freedom
3 (push/pull, rotate, bend; fu2011_development Sec. II-C)
Push
Friction rollers
Twist
Rotating drive unit
Tested up to
Phantom
Sim-to-real
None found — No simulation-trained controller found for this robot.

Testing

Highest level
Phantom
What was tested
thorax vascular phantom (Shelley Medical), catheterisation to the brachiocephalic trunk (fu2011_development Sec. VI)
Key numbers
image-sensor registration error mean 0.19 mm, RMS 0.49 mm, SD 0.46 mm over 700 wall points (fu2011_development Sec. VI, Fig. 9)

Mechanism

Push (translation)
friction-roller
Twist (rotation)
other: both stepping motors at equal speed rotate the whole friction-wheel unit about the catheter (fu2011_development Sec. IV); bending by a lead-screw slide unit moving the handle sleeve
Device range
N.R.
Target anatomy
thoracic/cardiac vessels (thorax vascular phantom, brachiocephalic trunk)

Sensing and interface

Force sensing
proximal: two force-sensing elements between the inner and outer wheel of the active friction wheel (fu2011_development Fig. 6)
Leader and haptics
Novint Falcon 3-DOF handle with force feedback (fu2011_development Fig. 2)
Imaging
EM magnetic tracking + MR-derived 3D guiding image (20 slices); intended to reduce X-ray use
Sterility
N.R.

Size

Footprint (mm)
N.R.
Mass (kg)
N.R.

Simulation

Simulation model
none found
Sim-trained controller
none found

Versions

Lineage
2008-2010 active/steerable catheter navigation studies (Fu et al., IJMRCAS 2008-2009, titles); 2011 robotic catheter system (fu2011_development, full); 2011 master-slave catheterisation system (fu2011_masterslave, title)

Sources

  1. Development of a novel robotic catheter system for endovascular minimally invasive surgery · The 2011 IEEE/ICME International Conference on Complex Medical Engineering, 2011 · doi:10.1109/iccme.2011.5876772 [fu2011_development]
  2. The master-slave catheterisation system for positioning the steerable catheter · International Journal of Mechatronics and Automation, 2011 · doi:10.1504/ijma.2011.045271 [fu2011_masterslave]

Notes

Notes
Anzhu Gao later moved to SJTU (row sjtumr_huang2022). The full-system photo (Fig. 1) and insertion mechanism (Fig. 5) carry baked-in labels.

This record

Read level
full — full (fu2011_development, OCR of a scanned PDF); title (fu2011_masterslave)
Picture
fu2011_development Fig. 7 handle manipulation mechanism (native scan raster p. 3) ↗ — licence: publisher copyright; grade C
Picture changes
native extract; cropped below baked-in text labels and above a blue arrow; two thin leader lines on the robot left; not upscaled
Found by
gap: Gap searches for robots the other searches missed.
Data
JSON · YAML: records/robots/hit_fu2011.yaml
Correction
Suggest a correction [repository not created yet]