 "schema_version": 1,
 "id": "tut_ma2021",
 "name": "TUT catheter system",
 "full_name": "TUT catheter operating system with real-catheter master and clamp-relay slave",
 "group": "Tianjin University of Technology, Tianjin Key Laboratory for Control Theory & Applications in Complicated Industry Systems (Xu Ma), with Tianjin University of Technology and Education",
 "group_short": "Tianjin U. Tech.",
 "family": "academic",
 "category": "academic",
 "first_year": 2020,
 "latest_year": 2021,
 "versions": "2020 robotic catheter manipulation system with BP-neural-network PID (ma2020_development, title); 2021 new catheter operating system (ma2021_design)",
 "target_anatomy": "N.R.",
 "devices": [
  "catheter"
 ],
 "devices_driven": "catheter",
 "n_dof": 2,
 "n_dof_detail": "2 (axial + rotation frames)",
 "translation_mechanism": "clamp-relay",
 "translation_detail": "clamp-relay",
 "rotation_mechanism": "rotating-drive-unit",
 "rotation_detail": "rotating frame (drive N.R.)",
 "device_range": "N.R.",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "proximal thrust measured during insertion; force sensor UNCSR and torque sensor OPT-563B listed (ma2021_design, PMC HTML)",
 "haptics_leader": "real catheter used as the input handle on a pulley; DC motor with encoder on the pulley gives force/torque feedback",
 "imaging_compat": "N.R.",
 "sterility": "N.R.",
 "highest_validation": "bench",
 "validation_detail": "simulation and axial-displacement bench tests only; no phantom or animal (ma2021_design)",
 "key_numbers": "axial stroke 300 mm, push/withdraw 30 mm/s; BP-PID max axial error < 1.5 mm, mean 0.48 mm; step-response overshoot 20% (PID) vs 12% (BP-PID) in simulation; Maxon EC32 axial motor; DSP TMS320F28335, serial 19200 baud (ma2021_design)",
 "sim_model": "none found (controller simulated as a mass-spring-damper only)",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "cleared",
  "grade": "C",
  "kind": "photo",
  "source": "ma2021_design Fig. 3 slave manipulator, https://cdn.ncbi.nlm.nih.gov/pmc/blobs/69b3/7861939/0d63d0cf83e2/ABB2021-8898311.003.jpg (page https://pmc.ncbi.nlm.nih.gov/articles/PMC7861939/)",
  "source_url": "https://doi.org/10.1155/2021/8898311",
  "licence": "CC BY (Creative Commons Attribution License, as stated)",
  "credit": "Ma et al., Applied Bionics and Biomechanics (2021), Fig. 3. CC BY (Creative Commons Attribution License, as stated).",
  "edits": "PMC image as served; no edits; not upscaled",
  "file": "pictures/tut_ma2021.jpg"
 },
 "pictures": [
  {
   "id": "tut_ma2021_01",
   "status": "cleared",
   "file": "pictures/extra/tut_ma2021_01.jpg",
   "thumb": "pictures/thumbs/tut_ma2021_01.jpg",
   "kind": "leader",
   "caption": "Master end manipulator: encoders on the catheter held by the surgeon",
   "source_key": "ma2021_design",
   "source": "Ma et al., Applied Bionics and Biomechanics (2021), Fig. 4 (PMC7861939 figure 004)",
   "source_url": "https://doi.org/10.1155/2021/8898311",
   "where": "Fig. 4 (PMC7861939 figure 004)",
   "licence": "CC BY (Creative Commons Attribution License, as stated)"
  },
  {
   "id": "tut_ma2021_02",
   "status": "cleared",
   "file": "pictures/extra/tut_ma2021_02.jpg",
   "thumb": "pictures/thumbs/tut_ma2021_02.jpg",
   "kind": "render",
   "caption": "Slave end console with two graspers, load cell and torque sensor",
   "source_key": "ma2021_design",
   "source": "Ma et al., Applied Bionics and Biomechanics (2021), Fig. 7 (PMC7861939 figure 007)",
   "source_url": "https://doi.org/10.1155/2021/8898311",
   "where": "Fig. 7 (PMC7861939 figure 007)",
   "licence": "CC BY (Creative Commons Attribution License, as stated)"
  },
  {
   "id": "tut_ma2021_03",
   "status": "cleared",
   "file": "pictures/extra/tut_ma2021_03.jpg",
   "thumb": "pictures/thumbs/tut_ma2021_03.jpg",
   "kind": "render",
   "caption": "Surgeon's console: hollow encoder, couplings and shaft holders",
   "source_key": "ma2021_design",
   "source": "Ma et al., Applied Bionics and Biomechanics (2021), Fig. 6 (PMC7861939 figure 006)",
   "source_url": "https://doi.org/10.1155/2021/8898311",
   "where": "Fig. 6 (PMC7861939 figure 006)",
   "licence": "CC BY (Creative Commons Attribution License, as stated)"
  },
  {
   "id": "tut_ma2021_04",
   "status": "cleared",
   "file": "pictures/extra/tut_ma2021_04.jpg",
   "thumb": "pictures/thumbs/tut_ma2021_04.jpg",
   "kind": "render",
   "caption": "2020 system: main end handle stage and slave end on its inclined stand",
   "source_key": "ma2020_development",
   "source": "Ma et al., Applied Bionics and Biomechanics (2020), Fig. 1 (PMC7775165 figure 001)",
   "source_url": "https://doi.org/10.1155/2020/8870106",
   "where": "Fig. 1 (PMC7775165 figure 001)",
   "licence": "CC BY (Creative Commons Attribution License, as stated)"
  },
  {
   "id": "tut_ma2021_05",
   "status": "cleared",
   "file": "pictures/extra/tut_ma2021_05.jpg",
   "thumb": "pictures/thumbs/tut_ma2021_05.jpg",
   "kind": "render",
   "caption": "Axial force measurement mechanism: catheter frame, clamp plate, load cell",
   "source_key": "ma2020_development",
   "source": "Ma et al., Applied Bionics and Biomechanics (2020), Fig. 4 (PMC7775165 figure 004)",
   "source_url": "https://doi.org/10.1155/2020/8870106",
   "where": "Fig. 4 (PMC7775165 figure 004)",
   "licence": "CC BY (Creative Commons Attribution License, as stated)"
  }
 ],
 "sources": [
  {
   "key": "ma2021_design",
   "doi": "10.1155/2021/8898311",
   "url": null,
   "title": "Design of a New Catheter Operating System for the Surgical Robot",
   "year": 2021,
   "venue": "Applied Bionics and Biomechanics"
  },
  {
   "key": "ma2020_development",
   "doi": "10.1155/2020/8870106",
   "url": null,
   "title": "Development of a Robotic Catheter Manipulation System Based on BP Neural Network PID Controller",
   "year": 2020,
   "venue": "Applied Bionics and Biomechanics"
  }
 ],
 "read_level": "partial",
 "read_level_detail": "partial (PMC full-text HTML read through WebFetch)",
 "notes": "Same Tianjin Key Lab (TUT) as A2's tianjin_song2023 but a different robot; lineage of Shuxiang Guo's Kagawa/BIT group.",
 "maintainers_own": false,
 "found_by": "gap"
}