 "schema_version": 1,
 "id": "tianjin_song2023",
 "name": "Tianjin MR-fluid",
 "full_name": "Tianjin master-slave interventional robot with MR-fluid passive haptic interface (catheter + guidewire simultaneously)",
 "group": "Tianjin University of Technology, Tianjin Key Laboratory for Control Theory and Applications (Yu Song, Yu Tian, Lin Li) with South China Agricultural University (Xuanchun Yin); Song Yu previously at Kagawa",
 "group_short": "Tianjin U. Tech.",
 "family": "academic",
 "category": "academic",
 "first_year": 2018,
 "latest_year": 2025,
 "versions": "song2018_performance (Biomed Microdevices 2018, Kagawa-era catheter operating system with haptics) -> song2023_masterslave (Sensors 2023) -> song2025_passive (MBEC 2025, passive haptic interface, abstract not read)",
 "target_anatomy": "1:1 in vitro vessel model with circulating pump (song2023_masterslave Sec. 4.3, Figs. 17-18)",
 "devices": [
  "guidewire",
  "catheter"
 ],
 "devices_driven": "catheter and guidewire navigated simultaneously",
 "n_dof": 3,
 "n_dof_detail": "3 (catheter axial + rotation, guidewire axial) (song2023 Sec. 3.2.1)",
 "translation_mechanism": "other",
 "translation_detail": "friction-roller + clamp hybrid: catheter clamped by a solenoid gripper and pushed by a stepper-driven slide (clamp + stage); guidewire between an active and an electromagnet-pressed driven friction wheel on a stepper; a solenoid guidewire-support clamp holds the wire while the catheter advances (Sec. 3.2.1-3.2.2, Figs. 11-12)",
 "rotation_mechanism": "rotating-clamp",
 "rotation_detail": "catheter only: the clamps of the catheter unit grip and a stepper motor rotates them (Sec. 3.2.2); guidewire not rotated",
 "device_range": "N.R.",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "catheter resistance: the gripper floats between four force sensors and touches them under load (Sec. 3.2.1); reflected to the MRF master",
 "haptics_leader": "MRF passive haptic interface with a rigid catheter as handle: axial motion read by a Hall encoder through a removable friction wheel, rotation through a rotating tube with encoder; the force-feedback friction wheel is braked by MR fluid (Sec. 3.1)",
 "imaging_compat": "N.R.",
 "sterility": "N.R.",
 "highest_validation": "phantom",
 "validation_detail": "bench: 1000 mm deliveries at 5-30 mm/s, 10 repeats per speed (Sec. 4.2, Fig. 15); phantom: surgeon at the master in an isolated room driving catheter and guidewire through a 1:1 in vitro model with circulating pump (Sec. 4.3, Figs. 17-18)",
 "key_numbers": "catheter axial delivery error mean 0.95 mm, max < 2.0 mm; guidewire mean 1 mm, max < 2.0 mm over 1000 mm at 5-30 mm/s (Sec. 4.2)",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "cleared",
  "grade": "C",
  "kind": "photo",
  "source": "song2023_masterslave Fig. 11(b) (developed slave equipment), Sensors 2023, image https://mdpi-res.com/sensors/sensors-23-03584/article_deploy/html/images/sensors-23-03584-g011.png",
  "source_url": "https://doi.org/10.3390/s23073584",
  "licence": "CC BY 4.0",
  "credit": "Song et al., Sensors (2023), Fig. 11(b). CC BY 4.0.",
  "edits": "MDPI web image (1657x1196); cropped to panel (b); not upscaled; white dashed module boxes and labels on the robot left in place",
  "file": "pictures/tianjin_song2023.jpg"
 },
 "pictures": [
  {
   "id": "tianjin_song2023_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Master manipulator at the console and slave robot at the 1:1 EVE model",
   "source_key": "song2025_passive",
   "source": "Song et al., Medical &amp; Biological Engineering &amp; Computing (2025), Fig. 23, p. 12",
   "source_url": "https://doi.org/10.1007/s11517-025-03374-2",
   "where": "Fig. 23, p. 12",
   "licence": "publisher copyright"
  },
  {
   "id": "tianjin_song2023_02",
   "status": "cleared",
   "file": "pictures/extra/tianjin_song2023_02.jpg",
   "thumb": "pictures/thumbs/tianjin_song2023_02.jpg",
   "kind": "setup",
   "caption": "Master manipulator, controller boards and slave manipulator with EVE model",
   "source_key": "song2023_masterslave",
   "source": "Song et al., Sensors (2023), Fig. 17 (left part) (MDPI HTML figure g017)",
   "source_url": "https://doi.org/10.3390/s23073584",
   "where": "Fig. 17 (left part) (MDPI HTML figure g017)",
   "licence": "CC BY 4.0"
  },
  {
   "id": "tianjin_song2023_03",
   "status": "cleared",
   "file": "pictures/extra/tianjin_song2023_03.jpg",
   "thumb": "pictures/thumbs/tianjin_song2023_03.jpg",
   "kind": "leader",
   "caption": "Master manipulator coupled to force sensors for calibration",
   "source_key": "song2023_masterslave",
   "source": "Song et al., Sensors (2023), Fig. 13 (MDPI HTML figure g013)",
   "source_url": "https://doi.org/10.3390/s23073584",
   "where": "Fig. 13 (MDPI HTML figure g013)",
   "licence": "CC BY 4.0"
  },
  {
   "id": "tianjin_song2023_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "leader",
   "caption": "Passive haptic master: rigid catheter, axial and radial haptic devices",
   "source_key": "song2025_passive",
   "source": "Song et al., Medical &amp; Biological Engineering &amp; Computing (2025), Fig. 3, p. 4",
   "source_url": "https://doi.org/10.1007/s11517-025-03374-2",
   "where": "Fig. 3, p. 4",
   "licence": "publisher copyright"
  },
  {
   "id": "tianjin_song2023_05",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "CAD of the slave robot with catheter and guidewire manipulators",
   "source_key": "song2025_passive",
   "source": "Song et al., Medical &amp; Biological Engineering &amp; Computing (2025), Fig. 19, p. 11",
   "source_url": "https://doi.org/10.1007/s11517-025-03374-2",
   "where": "Fig. 19, p. 11",
   "licence": "publisher copyright"
  },
  {
   "id": "tianjin_song2023_06",
   "status": "cleared",
   "file": "pictures/extra/tianjin_song2023_06.jpg",
   "thumb": "pictures/thumbs/tianjin_song2023_06.jpg",
   "kind": "leader",
   "caption": "Prototype of the MR-fluid master manipulator",
   "source_key": "song2023_masterslave",
   "source": "Song et al., Sensors (2023), Fig. 3(b) (MDPI HTML figure g003)",
   "source_url": "https://doi.org/10.3390/s23073584",
   "where": "Fig. 3(b) (MDPI HTML figure g003)",
   "licence": "CC BY 4.0"
  },
  {
   "id": "tianjin_song2023_07",
   "status": "cleared",
   "file": "pictures/extra/tianjin_song2023_07.jpg",
   "thumb": "pictures/thumbs/tianjin_song2023_07.jpg",
   "kind": "render",
   "caption": "CAD of the slave: friction wheels, electromagnet and guidewire clamp",
   "source_key": "song2023_masterslave",
   "source": "Song et al., Sensors (2023), Fig. 11(a) (MDPI HTML figure g011)",
   "source_url": "https://doi.org/10.3390/s23073584",
   "where": "Fig. 11(a) (MDPI HTML figure g011)",
   "licence": "CC BY 4.0"
  },
  {
   "id": "tianjin_song2023_08",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Axial haptic device with force sensor and rigid catheters on the bench",
   "source_key": "song2025_passive",
   "source": "Song et al., Medical &amp; Biological Engineering &amp; Computing (2025), Fig. 15, p. 9",
   "source_url": "https://doi.org/10.1007/s11517-025-03374-2",
   "where": "Fig. 15, p. 9",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "song2023_masterslave",
   "doi": "10.3390/s23073584",
   "url": null,
   "title": "A Novel Master–Slave Interventional Surgery Robot with Force Feedback and Collaborative Operation",
   "year": 2023,
   "venue": "Sensors"
  },
  {
   "key": "song2018_performance",
   "doi": "10.1007/s10544-018-0294-4",
   "url": null,
   "title": "Performance evaluation of a robot-assisted catheter operating system with haptic feedback",
   "year": 2018,
   "venue": "Biomedical Microdevices"
  },
  {
   "key": "song2025_passive",
   "doi": "10.1007/s11517-025-03374-2",
   "url": null,
   "title": "Passive haptic interface for robot-assisted endovascular catheterization",
   "year": 2025,
   "venue": "Medical &amp; Biological Engineering &amp; Computing"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (song2023_masterslave, ); abstract (song2018_performance, song2025_passive)",
 "notes": "Mechanism details need the Sensors 2023 PDF (manual-download list). | 2026-11-08 X3: Sensors 2023 read in full; translation N.R. -> friction-roller + clamp hybrid; gallery already 8 (G2 used this paper's MDPI figures)",
 "maintainers_own": false,
 "found_by": "academic"
}