 "schema_version": 1,
 "id": "srmc_abah2024",
 "name": "Robotic microcatheter",
 "full_name": "Self-steering robotic microcatheter (series-elastic, two bending segments)",
 "group": "Simaan, Vanderbilt (ARMA lab) with Chitale, VUMC",
 "group_short": "Vanderbilt",
 "family": "tip",
 "category": "actuated-device",
 "first_year": 2024,
 "latest_year": 2024,
 "versions": "single (2024 T-MRB); path planner in companion paper (ref [45] in abah2024, not read)",
 "target_anatomy": "neuroendovascular (ICA path, 2D vasculature models)",
 "devices": [
  "catheter"
 ],
 "devices_driven": "own notched-nitinol catheter: outer tube OD 1.2 mm / ID 0.95 mm, intermediate 0.89/0.725, inner 0.66 OD with 0.475 mm bore (S8 check of abah2024 Sec. IV-A); roughly 3.6 Fr at maximum OD (1.2 mm / 0.333, my conversion)",
 "n_dof": 4,
 "n_dof_detail": "4 (insertion, roll, proximal bend, distal bend; bending in two separate planes)",
 "translation_mechanism": "tendon/tip-actuation",
 "translation_detail": "tendon/tip-actuation",
 "rotation_mechanism": "rotating-drive-unit",
 "rotation_detail": "custom actuation unit roll DoF",
 "device_range": "N.R.",
 "footprint_mm": "N.R. (benchtop actuation unit on aluminium extrusion frame, Fig. 3)",
 "mass_kg": "N.R.",
 "force_sensing": "joint-level load sensing via series-elastic actuation; active compliance triggers at 0.01 N tip force (about 1 g)",
 "haptics_leader": "no",
 "imaging_compat": "bi-plane fluoroscopy for tracking (Fig. 11); catheter tracking and pose filtering algorithms",
 "sterility": "N.R.",
 "highest_validation": "phantom",
 "validation_detail": "2D vasculature model: branch selection left and right, insertion into unknown channel under active compliance, one autonomous deployment along 2D ICA path; no success counts",
 "key_numbers": "distal segment OD about 0.88-1.2 mm (text differs, see S8 note); compliance trigger 0.01 N; autonomous insertion errors larger than simulated because passive extension floats between actuation unit and phantom entry",
 "sim_model": "kinematic model with calibration and simulated path following along 2D ICA (Fig. 8) (abah2024 Sec. V, full)",
 "has_sim_model": true,
 "sim2real": "model-based",
 "sim2real_detail": "model-based: planner configurations computed in simulation were run on the hardware; hardware position errors exceeded simulated errors (abah2024 Sec. V-C, full); no learned policy",
 "picture": {
  "status": "not-cleared",
  "grade": "A",
  "kind": "photo",
  "source": "abah2024_self Fig. 1 (photo of the 4-DOF actuation unit)",
  "source_url": "https://doi.org/10.1109/tmrb.2024.3464123",
  "licence": "CC BY-NC-ND 4.0 (PMC12245187 author manuscript; edits mean internal use only)",
  "credit": "Abah et al., IEEE Transactions on Medical Robotics and Bionics (2024), Fig. 1. CC BY-NC-ND 4.0 (PMC12245187 author manuscript; edits mean internal use only).",
  "edits": "native extract (4110 px figure); cropped to the photo; one brown arrow on the wooden base and a 6 px arrow tip erased (LaMa)",
  "file": null
 },
 "pictures": [
  {
   "id": "srmc_abah2024_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Actuation unit photo with CAD of the push-pull tube carriages",
   "source_key": "abah2024_self",
   "source": "Abah et al., IEEE Transactions on Medical Robotics and Bionics (2024), Fig. 1, p. 3",
   "source_url": "https://doi.org/10.1109/tmrb.2024.3464123",
   "where": "Fig. 1, p. 3",
   "licence": "CC BY-NC-ND 4.0 (PMC author manuscript)"
  },
  {
   "id": "srmc_abah2024_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Active compliance demonstration: microcatheter in a curved channel and beside a Phenom 21",
   "source_key": "abah2024_self",
   "source": "Abah et al., IEEE Transactions on Medical Robotics and Bionics (2024), Fig. 7, p. 6",
   "source_url": "https://doi.org/10.1109/tmrb.2024.3464123",
   "where": "Fig. 7, p. 6",
   "licence": "CC BY-NC-ND 4.0 (PMC author manuscript)"
  },
  {
   "id": "srmc_abah2024_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Video snapshots of autonomous insertion and steering in a 2D ICA model",
   "source_key": "abah2024_self",
   "source": "Abah et al., IEEE Transactions on Medical Robotics and Bionics (2024), Fig. 10, p. 7",
   "source_url": "https://doi.org/10.1109/tmrb.2024.3464123",
   "where": "Fig. 10, p. 7",
   "licence": "CC BY-NC-ND 4.0 (PMC author manuscript)"
  },
  {
   "id": "srmc_abah2024_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Microcatheter selecting the left and right branch of a Y phantom",
   "source_key": "abah2024_self",
   "source": "Abah et al., IEEE Transactions on Medical Robotics and Bionics (2024), Fig. 6, p. 6",
   "source_url": "https://doi.org/10.1109/tmrb.2024.3464123",
   "where": "Fig. 6, p. 6",
   "licence": "CC BY-NC-ND 4.0 (PMC author manuscript)"
  }
 ],
 "sources": [
  {
   "key": "abah2024_self",
   "doi": "10.1109/tmrb.2024.3464123",
   "url": null,
   "title": "Self-Steering Catheters for Neuroendovascular Interventions",
   "year": 2024,
   "venue": "IEEE Transactions on Medical Robotics and Bionics"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full",
 "notes": "Sized for microcatheter range, the smallest actuated-tip multi-DOF catheter among the read rows. Uses the CC BY-NC-ND statement printed on the author manuscript.",
 "maintainers_own": false,
 "found_by": "actuated-tip"
}