 "schema_version": 1,
 "id": "sjtu_coronary_meng2025",
 "name": "SJTU PCI robot",
 "full_name": "SJTU master-slave PCI robot with magnetic-coupling modular drive and thin-film force sensing",
 "group": "Shanghai Jiao Tong University, School of Mechanical Engineering (Meng, Fang, Jing, Fu et al.)",
 "group_short": "SJTU",
 "family": "academic",
 "category": "academic",
 "first_year": 2025,
 "latest_year": 2025,
 "versions": "meng2025_design (ROBIO 2025)",
 "target_anatomy": "coronary PCI (complex vascular geometries in simulation and bench)",
 "devices": [
  "guidewire",
  "catheter",
  "therapeutic-device"
 ],
 "devices_driven": "guidewire (0.014-0.035 in), guide catheter, balloon/stent catheter",
 "n_dof": 4,
 "n_dof_detail": "4 (guidewire advance and rotation; catheter linear insertion; balloon/stent linear transport)",
 "translation_mechanism": "friction-roller",
 "translation_detail": "friction-roller (guidewire: gear-driven dual friction wheels; balloon/stent: symmetric friction wheels; catheter: gear-rack clamping with linear drive)",
 "rotation_mechanism": "other",
 "rotation_detail": "Guidewire rotated continuously through drive wheel 1 and bevel gears (360 deg)",
 "device_range": "guidewire 0.014 in to 0.035 in",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "Thin-film pressure sensors between worm and housing give axial guidewire force",
 "haptics_leader": "N.R.",
 "imaging_compat": "N.R.",
 "sterility": "Electromechanical decoupling by magnetic coupling torque transmission so the distal module can be a hygienic, separable part",
 "highest_validation": "bench",
 "validation_detail": "Simulink control simulation plus hardware experiments; number of trials N.R.",
 "key_numbers": "Displacement error within +/-1 mm and angular error within +/-15 deg; Smith predictor + RBF self-tuning PID",
 "sim_model": "Simulink control-loop model only",
 "has_sim_model": true,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "C",
  "kind": "render",
  "source": "meng2025_design Fig. 1 (structure of the robotic system), ROBIO 2025, doi:10.1109/ROBIO66223.2025.11378419",
  "source_url": "https://doi.org/10.1109/ROBIO66223.2025.11378419",
  "licence": "publisher copyright",
  "credit": "Meng et al., 2025 IEEE International Conference on Robotics and Biomimetics (ROBIO) (2025), Fig. 1. publisher copyright.",
  "edits": "figure rebuilt from its 19 embedded raster tiles at native scale (labels and leader lines are inside the tiles); main panel cropped; fragments of neighbouring panels whitened; not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "sjtu_coronary_meng2025_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "CAD of the guidewire drive module with an inset of the catheter module",
   "source_key": "meng2025_design",
   "source": "Meng et al., 2025 IEEE International Conference on Robotics and Biomimetics (ROBIO) (2025), Fig. 1 (right part), p. 2",
   "source_url": "https://doi.org/10.1109/robio66223.2025.11378419",
   "where": "Fig. 1 (right part), p. 2",
   "licence": "publisher copyright"
  }
 ],
 "sources": [
  {
   "key": "meng2025_design",
   "doi": "10.1109/robio66223.2025.11378419",
   "url": null,
   "title": "Design and Control of a Robotic System for Coronary Interventions",
   "year": 2025,
   "venue": "2025 IEEE International Conference on Robotics and Biomimetics (ROBIO)"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (meng2025 text)",
 "notes": "CAD render only; no prototype photo in the paper.",
 "maintainers_own": false,
 "found_by": "academic"
}