 "schema_version": 1,
 "id": "siatguidewire_zhang2022",
 "name": "Magnet gantry",
 "full_name": "SIAT magnetically steerable guidewire: two-sided permanent-magnet gantry (2022) and arm-held magnet with a guidewire advancer (2023)",
 "group": "Tiantian Xu group, Shenzhen Institutes of Advanced Technology (CAS), with the SIAT-CUHK joint lab / CUHK (Li Zhang) and the Department of Neurosurgery, South China Hospital of Shenzhen University (xu2023_magnetically)",
 "group_short": "SIAT",
 "family": "magnetic",
 "category": "magnetic-navigation",
 "first_year": 2022,
 "latest_year": 2023,
 "versions": "2022 RA-L 3D steering system, two-sided permanent-magnet gantry, guidewire fed by hand (zhang2022_magnetic); 2022 RCAR paper of Fu et al. (cited as xu2023 ref. 36; not read); 2023 Adv. Intell. Syst. (Fu ... Du, Xu): UR10 arm carrying a D50 x L25 mm N52 magnet with a pole-flip stepper on a mobile cart, 4-DOF guidewire/catheter advancer, new tip, dipole + Cosserat models, trajectory planning, DSA test (xu2023_magnetically, full)",
 "target_anatomy": "2022: peripheral phantoms (silicone thigh vessels ID 1 mm to 1.6 cm); 2023: femoral artery -> aortic arch -> carotid -> intracranial arteries in a 959 mm silicone phantom (xu2023_magnetically Sec. 3.4)",
 "devices": [
  "guidewire",
  "catheter"
 ],
 "devices_driven": "2022: commercial hydrophilic guidewire (LWSTDA35150) joined to an Ecoflex + NdFeB (1:1 mass) soft actuator 2.5 cm long and 1.9 mm in diameter; 2023: 0.035-in (0.89 mm) commercial guidewire with two axially magnetised 0.5 x 2.5 mm N52 NdFeB magnets in a PDMS-filled silicone tube (0.5/1 mm ID/OD); the advancer takes 0.3-5 mm devices (guidewire and catheter), only the guidewire moved in the experiments (xu2023_magnetically Sec. 2.1)",
 "n_dof": null,
 "n_dof_detail": "2022: two symmetric gantries, each three linear motors (x, y, z) carrying a rectangular permanent magnet; 2023: 6-DOF UR10 arm + pole-flip stepper; guidewire and catheter advancers with 4 DOF in total (xu2023_magnetically Sec. 2.1)",
 "translation_mechanism": "magnetic",
 "translation_detail": "magnetic (steering); 2022 forward push by hand; 2023 advancer: stepper-driven friction-wheel pair with a spring/screw clamp (xu2023_magnetically p. 4, Fig. S1)",
 "rotation_mechanism": "magnetic-field",
 "rotation_detail": "field direction; 2023 advancer rotates its linear module with a stepper and gear pair (xu2023_magnetically p. 4)",
 "device_range": "2023: advance speed 0-15 mm/s; magnet distance 40-180 mm (xu2023_magnetically p. 4, 8)",
 "footprint_mm": "2022 system 880 x 620 x 1600; workspace 700 x 500 x 500; 2023 mobile cart, dimensions N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "none",
 "haptics_leader": "no; 2023 teleoperation console with an Xbox joystick through ROS (xu2023_magnetically Sec. 3.4)",
 "imaging_compat": "camera in 2022 and on the 2023 bench; 2023 Philips Azurion 7M20 DSA in a cath room, system operated from outside the room (xu2023_magnetically p. 12-13)",
 "sterility": "N.R.",
 "highest_validation": "phantom",
 "validation_detail": "2022: ring-passing task (8 mm rings) and path following, 3D vascular phantom; 2023: trajectory following in 3 resin phantoms (ID 10-13 mm, 160-180 mm long), teleoperated navigation to 4 intracranial targets in a 959 mm silicone phantom with pulsatile flow, one path under DSA in the partner hospital's cath room; repeat counts N.R.; pig tests named as future work (xu2023_magnetically Sec. 3)",
 "key_numbers": "2022: field 0-20 mT; deflection up to 180 degrees; localisation accuracy 8 mm (ring diameter). 2023: mean tip position error 1.52-1.88 mm and tip angle error 2.36-2.57 degrees over 3 phantoms; deflection up to about 100 degrees; navigation within 1-2 min; advance speeds 1.2-6.06 mm/s (xu2023_magnetically Sec. 3.3)",
 "sim_model": "2022 kinematic model; 2023 dipole field model, constant-curvature beam and Cosserat rod models (MATLAB bvp5c), COMSOL 5.6 field and tip FEM (xu2023_magnetically Sec. 2-3)",
 "has_sim_model": true,
 "sim2real": "yes",
 "sim2real_detail": "yes: luo2025_rlmg (RCAR 2025, full; found by W5): DDPG trained in SOFA+BeamAdapter, UR5 + permanent magnet + this advancer; 4 paths x 3 trials in a femoral-artery phantom; error 2.2-3.1 mm sim -> 2.8-3.8 mm real, time 7-13 s -> 31-67 s (Table I); no success rate",
 "picture": {
  "status": "not-cleared",
  "grade": "B",
  "kind": "photo",
  "source": "zhang2022_magnetic Fig. 1(b) (two-sided permanent-magnet gantry with phantom)",
  "source_url": "https://doi.org/10.1109/lra.2022.3146909",
  "licence": "publisher copyright (IEEE)",
  "credit": "Zhang et al., IEEE Robotics and Automation Letters (2022), Fig. 1(b). publisher copyright (IEEE).",
  "edits": "native extract; crop; 9 red text labels + red arrows and a yellow outline erased (LaMa by colour mask), grey ghosts on the white background flattened, one yellow line inpainted (OpenCV Telea)",
  "file": null
 },
 "pictures": [
  {
   "id": "siatguidewire_zhang2022_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Guidewire passing through a set of rings steered by the magnet gantry",
   "source_key": "zhang2022_magnetic",
   "source": "Zhang et al., IEEE Robotics and Automation Letters (2022), Fig. 8, p. 5",
   "source_url": "https://doi.org/10.1109/lra.2022.3146909",
   "where": "Fig. 8, p. 5",
   "licence": "publisher copyright"
  },
  {
   "id": "siatguidewire_zhang2022_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Guidewire following three planned paths in a vascular phantom, time series",
   "source_key": "zhang2022_magnetic",
   "source": "Zhang et al., IEEE Robotics and Automation Letters (2022), Fig. 10, p. 6",
   "source_url": "https://doi.org/10.1109/lra.2022.3146909",
   "where": "Fig. 10, p. 6",
   "licence": "publisher copyright"
  },
  {
   "id": "siatguidewire_zhang2022_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "close-up",
   "caption": "Soft magnetic actuator at the guidewire tip bending under 1 to 20 mT",
   "source_key": "zhang2022_magnetic",
   "source": "Zhang et al., IEEE Robotics and Automation Letters (2022), Fig. 6, p. 5",
   "source_url": "https://doi.org/10.1109/lra.2022.3146909",
   "where": "Fig. 6, p. 5",
   "licence": "publisher copyright"
  },
  {
   "id": "siatguidewire_zhang2022_04",
   "status": "cleared",
   "file": "pictures/extra/siatguidewire_zhang2022_04.jpg",
   "thumb": "pictures/thumbs/siatguidewire_zhang2022_04.jpg",
   "kind": "render",
   "caption": "2023 system concept: arm-held magnet on a mobile cart, wire/catheter advancer, console (labelled)",
   "source_key": "xu2023_magnetically",
   "source": "Fu et al., Advanced Intelligent Systems (2023), Fig. 1(b), p. 3",
   "source_url": "https://doi.org/10.1002/aisy.202300267",
   "where": "Fig. 1(b), p. 3",
   "licence": "CC BY (Creative Commons Attribution License, version not printed)"
  },
  {
   "id": "siatguidewire_zhang2022_05",
   "status": "cleared",
   "file": "pictures/extra/siatguidewire_zhang2022_05.jpg",
   "thumb": "pictures/thumbs/siatguidewire_zhang2022_05.jpg",
   "kind": "setup",
   "caption": "In-vitro setup: arm-held magnet, pulsatile pumps, 3D phantom and guidewire advancer (labelled)",
   "source_key": "xu2023_magnetically",
   "source": "Fu et al., Advanced Intelligent Systems (2023), Fig. 5(a), p. 12",
   "source_url": "https://doi.org/10.1002/aisy.202300267",
   "where": "Fig. 5(a), p. 12",
   "licence": "CC BY (Creative Commons Attribution License, version not printed)"
  },
  {
   "id": "siatguidewire_zhang2022_06",
   "status": "cleared",
   "file": "pictures/extra/siatguidewire_zhang2022_06.jpg",
   "thumb": "pictures/thumbs/siatguidewire_zhang2022_06.jpg",
   "kind": "in-use",
   "caption": "System on a mobile cart beside the bed under DSA in a catheterisation room",
   "source_key": "xu2023_magnetically",
   "source": "Fu et al., Advanced Intelligent Systems (2023), Fig. 6(a), p. 13",
   "source_url": "https://doi.org/10.1002/aisy.202300267",
   "where": "Fig. 6(a), p. 13",
   "licence": "CC BY (Creative Commons Attribution License, version not printed)"
  },
  {
   "id": "siatguidewire_zhang2022_07",
   "status": "cleared",
   "file": "pictures/extra/siatguidewire_zhang2022_07.jpg",
   "thumb": "pictures/thumbs/siatguidewire_zhang2022_07.jpg",
   "kind": "setup",
   "caption": "Trajectory-following platform: arm-held magnet, phantom, cameras and guidewire advancer (labelled)",
   "source_key": "xu2023_magnetically",
   "source": "Fu et al., Advanced Intelligent Systems (2023), Fig. 4(a), p. 10",
   "source_url": "https://doi.org/10.1002/aisy.202300267",
   "where": "Fig. 4(a), p. 10",
   "licence": "CC BY (Creative Commons Attribution License, version not printed)"
  }
 ],
 "sources": [
  {
   "key": "zhang2022_magnetic",
   "doi": "10.1109/lra.2022.3146909",
   "url": null,
   "title": "Design and Characteristics of 3D Magnetically Steerable Guidewire System for Minimally Invasive Surgery",
   "year": 2022,
   "venue": "IEEE Robotics and Automation Letters"
  },
  {
   "key": "xu2023_magnetically",
   "doi": "10.1002/aisy.202300267",
   "url": null,
   "title": "A Magnetically Controlled Guidewire Robot System with Steering and Propulsion Capabilities for Vascular Interventional Surgery",
   "year": 2023,
   "venue": "Advanced Intelligent Systems"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (zhang2022_magnetic, xu2023_magnetically)",
 "notes": "2026-10-13 coordinator: sim2real corrected from 'none found' (W5_notes.md item 1). | X2 2026-11-08: xu2023 (Wiley, CC BY) received by hand and read in full; first author is Fu (key xu2023_magnetically kept); validation stays phantom; luo2025_rlmg used a UR5, xu2023 a UR10 (same group, different arm).",
 "maintainers_own": false,
 "found_by": "actuated-tip"
}