 "schema_version": 1,
 "id": "saer_sankaran2018",
 "name": "SAER",
 "full_name": "SAER Surgeon Augmented Endovascular Robotic system",
 "group": "Univ. of Illinois Urbana-Champaign / Univ. at Buffalo (Kesavadas, Siddiqui, Snyder)",
 "group_short": "UIUC / Buffalo",
 "family": "academic",
 "category": "academic",
 "first_year": 2018,
 "latest_year": 2020,
 "versions": "2018 TBME (sankaran2018_design); 2020 IJMRCAS force calibration of the same robot's current-based proximal force estimate by bilevel optimisation (sankaran2020_force, full)",
 "target_anatomy": "branching vascular phantoms (one mimics bovine aortic arch)",
 "devices": [
  "guidewire"
 ],
 "devices_driven": "conventional guidewire (same type on leader and follower)",
 "n_dof": 2,
 "n_dof_detail": "2 (translation, rotation, uncoupled)",
 "translation_mechanism": "friction-roller",
 "translation_detail": "friction-roller (knurled rollers)",
 "rotation_mechanism": "rotating-drive-unit",
 "rotation_detail": "motor rotating the roller box about the tool axis (slip ring)",
 "device_range": "N.R.",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "none; reaction force estimated from motor current, calibrated by bilevel optimisation, checked against a load cell",
 "haptics_leader": "sensing unit with encoder-read rollers on a real tool; vibratory/audio alert above 1.5 N (no force reflection)",
 "imaging_compat": "N.R.",
 "sterility": "N.R.",
 "highest_validation": "phantom",
 "validation_detail": "simple and complex branching phantoms; count of runs N.R.; 2020: load-cell bench with a short phantom section and guidewire navigation in a simple branching phantom with a haptic alert at the block (sankaran2020 Figs. 2, 9)",
 "key_numbers": "stop/haptic threshold 1.5 N; force estimate validated at 30 and 50 rpm (curves, no error number); 2020: guidewire feed 15.7 mm/s at 30 rpm; bias force 0.6716 N at 30 rpm; unbiased estimate vs load cell RMS error 0.22 N (30 rpm) and 0.23 N (50 rpm) (sankaran2020 Results, Fig. 8)",
 "sim_model": "dynamic model of the drive only (no vessel simulator)",
 "has_sim_model": true,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "A",
  "kind": "photo",
  "source": "sankaran2018_design Fig. 4, right panel (robot module photo; embedded 3000x895)",
  "source_url": "https://doi.org/10.1109/tbme.2018.2800639",
  "licence": "publisher copyright (IEEE)",
  "credit": "Sankaran et al., IEEE Transactions on Biomedical Engineering (2018), Fig. 4. publisher copyright (IEEE).",
  "edits": "native extract; cropped off the labelled CAD panel; not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "saer_sankaran2018_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Operator at the sensing unit teleoperating the robot unit into a phantom",
   "source_key": "sankaran2018_design",
   "source": "Sankaran et al., IEEE Transactions on Biomedical Engineering (2018), Fig. 14, p. 8",
   "source_url": "https://doi.org/10.1109/tbme.2018.2800639",
   "where": "Fig. 14, p. 8",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "saer_sankaran2018_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Force validation bench: robot unit, load cell, tube, controllers and power supply",
   "source_key": "sankaran2018_design",
   "source": "Sankaran et al., IEEE Transactions on Biomedical Engineering (2018), Fig. 10, p. 6",
   "source_url": "https://doi.org/10.1109/tbme.2018.2800639",
   "where": "Fig. 10, p. 6",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "saer_sankaran2018_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "CAD of the robot module: gear train, knurled-roller box unit, slip ring",
   "source_key": "sankaran2018_design",
   "source": "Sankaran et al., IEEE Transactions on Biomedical Engineering (2018), Fig. 4, left panel, p. 4",
   "source_url": "https://doi.org/10.1109/tbme.2018.2800639",
   "where": "Fig. 4, left panel, p. 4",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "saer_sankaran2018_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Exploded view of the robot module and its linear drive unit",
   "source_key": "sankaran2018_design",
   "source": "Sankaran et al., IEEE Transactions on Biomedical Engineering (2018), Fig. 5, p. 4",
   "source_url": "https://doi.org/10.1109/tbme.2018.2800639",
   "where": "Fig. 5, p. 4",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "saer_sankaran2018_05",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "leader",
   "caption": "Sensor module CAD and the surgeon's translation and rotation holding postures",
   "source_key": "sankaran2018_design",
   "source": "Sankaran et al., IEEE Transactions on Biomedical Engineering (2018), Fig. 2(a)-(c), p. 3",
   "source_url": "https://doi.org/10.1109/tbme.2018.2800639",
   "where": "Fig. 2(a)-(c), p. 3",
   "licence": "publisher copyright (IEEE)"
  },
  {
   "id": "saer_sankaran2018_06",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "leader",
   "caption": "Sensor module CAD and photo with a hand-held catheter; robot module CAD and photo",
   "source_key": "sankaran2020_force",
   "source": "Sankaran et al., The International Journal of Medical Robotics and Computer Assisted Surgery (2020), Fig. 1(A)-(B), p. 2",
   "source_url": "https://doi.org/10.1002/rcs.2045",
   "where": "Fig. 1(A)-(B), p. 2",
   "licence": "publisher copyright (Wiley)"
  }
 ],
 "sources": [
  {
   "key": "sankaran2018_design",
   "doi": "10.1109/tbme.2018.2800639",
   "url": null,
   "title": "Design and Development of Surgeon Augmented Endovascular Robotic System",
   "year": 2018,
   "venue": "IEEE Transactions on Biomedical Engineering"
  },
  {
   "key": "sankaran2020_force",
   "doi": "10.1002/rcs.2045",
   "url": null,
   "title": "Force calibration for an endovascular robotic system with proximal force measurement",
   "year": 2020,
   "venue": "The International Journal of Medical Robotics and Computer Assisted Surgery"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (partial text) (sankaran2018_design); full (sankaran2020_force, )",
 "notes": "2026-11-08 X3: ashammagari2016_image (IMECE 2016, same Kesavadas group) is NOT this robot: its autonomous aortic-arch run used a 7-DOF Robai Cyton arm feeding a guidewire into the VIST simulator, MDP policy, 30/30 trials, mean 31.23 s (Experiments and Results, Fig. 4)",
 "maintainers_own": false,
 "found_by": "academic"
}