 "schema_version": 1,
 "id": "ryerson_cathnav_ganji2009",
 "name": "CathNav",
 "full_name": "CathNav / Althea I and Althea II: 3-DOF robots for semi-automatic and automatic navigation of steerable EP catheters",
 "group": "Ryerson University (now Toronto Metropolitan University; F. Janabi-Sharifi) with University of Waterloo (Y. Ganji) and St. Michael's Hospital (A. N. Cheema)",
 "group_short": "Ryerson",
 "family": "academic",
 "category": "academic",
 "first_year": 2009,
 "latest_year": 2021,
 "versions": "2009 IJCARS CathNav, called Althea I by the group (norouzighazbi2021 Sec. 1, ref. 14): Velmex linear stage + two rotary stages, stepper motors, 3 swine; 2021 Artif. Organs Althea II (norouzighazbi2021_design, full): three HerkuleX DRS-0602 smart servos for twist, translation and knob bending (Sec. 2.2.1-2.2.2) on an inclined 3D-printed arm (SLS nylon + FDM) riding a Macron MSA-R6S belt-driven linear actuator (325 mm travel, up to 6230 mm/s), customisable handle parts for other catheters, guidewires and sheaths, magnetic stainless-steel shaft holders, gearbox and external bearings removed, portable, assembly 20 min, five times faster than Althea I, two units can cooperate (Sec. 2.2, Figs. 2-3, 11); the 2024 grip-insert-release follower of the same group is robots_A3 toronto_nazari2024",
 "target_anatomy": "right atrium (intracardiac EP navigation)",
 "devices": [
  "guidewire",
  "catheter",
  "sheath"
 ],
 "devices_driven": "one bidirectional steerable ablation catheter SteeroCath-T (Boston Scientific): 7 Fr shaft, 8 Fr tip, 110 cm (ganji2009 Sec. System setup); Althea II: EP catheters with rotational knobs, swappable handle holder parts for other handles, guidewires and sheaths (norouzighazbi2021 Sec. 2.2.1)",
 "n_dof": 3,
 "n_dof_detail": "3 (insertion/retraction, handle twist, knob deflection)",
 "translation_mechanism": "linear-stage+clamp",
 "translation_detail": "linear-stage+clamp (2009: Velmex linear stage moves the handle, telescopic tubing keeps the external shaft straight; 2021: belt-driven linear actuator carries the handle holder, norouzighazbi2021 Sec. 2.2.2)",
 "rotation_mechanism": "handle-rotation",
 "rotation_detail": "two rotary stages: one twists the handle, one turns the steering knob",
 "device_range": "one catheter model",
 "footprint_mm": "N.R.",
 "mass_kg": "N.R.",
 "force_sensing": "none",
 "haptics_leader": "Novint Falcon 3-D haptic device as input (force feedback N.R.); CathNav GUI",
 "imaging_compat": "NDI Aurora EM tracking, two sensors on the distal shaft; fluoroscopy",
 "sterility": "N.R.",
 "highest_validation": "animal",
 "validation_detail": "3 swine (St. Michael's Hospital vivarium); catheter introduced manually into the right atrium then mounted on the robot; automatic navigation to EM-sensor targets (high RA, tricuspid annulus) (ganji2009 Sec. Experiments setup, Results); Althea II 2021: bench accuracy/repeatability in positioning and velocity modes, comparison with Althea I, cooperative two-robot object manipulation, mitral-valve and clip grasping inside an artificial heart model, image-based visual servoing; no in vivo test (norouzighazbi2021 Sec. 3-4)",
 "key_numbers": "mean distance to target 6.53 mm, all trials below 10 mm (ganji2009 Abstract, Results); Althea II: > 90 % accuracy in velocity and positioning modes, about 2.5x more accurate and 6x more repeatable than Althea I (Table 5), cooperative object path error 0.299 mm, YOLO tip detection MSE 0.05 mm (Table 4), assembly 18 / 23 / 17 min by three people (norouzighazbi2021)",
 "sim_model": "none found",
 "has_sim_model": false,
 "sim2real": "none-found",
 "sim2real_detail": "none found",
 "picture": {
  "status": "not-cleared",
  "grade": "B",
  "kind": "photo",
  "source": "norouzighazbi2021_design Fig. 2(A), right photograph (Althea II prototype on its belt-driven linear actuator), p. E174, Artif. Organs 2021 doi:10.1111/aor.13870",
  "source_url": "https://doi.org/10.1111/aor.13870",
  "licence": "publisher copyright (Wiley / ICAOT)",
  "credit": "Norouzi‐Ghazbi et al., Artificial Organs (2021), Fig. 2. publisher copyright (Wiley / ICAOT).",
  "edits": "native extract (1890 x 1217 figure); cropped to the photograph inside panel (A); nothing erased; not upscaled",
  "file": null
 },
 "pictures": [
  {
   "id": "ryerson_cathnav_ganji2009_01",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "render",
   "caption": "Althea II motors and motions, inclined arm and rotatory handle holder, with photos",
   "source_key": "norouzighazbi2021_design",
   "source": "Norouzi‐Ghazbi et al., Artificial Organs (2021), Fig. 2(A)-(B), p. E174",
   "source_url": "https://doi.org/10.1111/aor.13870",
   "where": "Fig. 2(A)-(B), p. E174",
   "licence": "publisher copyright (Wiley / ICAOT)"
  },
  {
   "id": "ryerson_cathnav_ganji2009_02",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Two Althea II units cooperating, magnetic shaft holders and mated catheter tips",
   "source_key": "norouzighazbi2021_design",
   "source": "Norouzi‐Ghazbi et al., Artificial Organs (2021), Fig. 11(A)-(C), p. E182",
   "source_url": "https://doi.org/10.1111/aor.13870",
   "where": "Fig. 11(A)-(C), p. E182",
   "licence": "publisher copyright (Wiley / ICAOT)"
  },
  {
   "id": "ryerson_cathnav_ganji2009_03",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Magnetic shaft holders routing the catheter and the EM-compatible test table",
   "source_key": "norouzighazbi2021_design",
   "source": "Norouzi‐Ghazbi et al., Artificial Organs (2021), Fig. 3(A)-(B), p. E175",
   "source_url": "https://doi.org/10.1111/aor.13870",
   "where": "Fig. 3(A)-(B), p. E175",
   "licence": "publisher copyright (Wiley / ICAOT)"
  },
  {
   "id": "ryerson_cathnav_ganji2009_04",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Cooperative object manipulation by two catheters beside the EM field generator",
   "source_key": "norouzighazbi2021_design",
   "source": "Norouzi‐Ghazbi et al., Artificial Organs (2021), Fig. 13, p. E183",
   "source_url": "https://doi.org/10.1111/aor.13870",
   "where": "Fig. 13, p. E183",
   "licence": "publisher copyright (Wiley / ICAOT)"
  },
  {
   "id": "ryerson_cathnav_ganji2009_05",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "in-use",
   "caption": "Catheter grasping the mitral valve and clip inside an artificial heart model",
   "source_key": "norouzighazbi2021_design",
   "source": "Norouzi‐Ghazbi et al., Artificial Organs (2021), Fig. 12, p. E182",
   "source_url": "https://doi.org/10.1111/aor.13870",
   "where": "Fig. 12, p. E182",
   "licence": "publisher copyright (Wiley / ICAOT)"
  },
  {
   "id": "ryerson_cathnav_ganji2009_06",
   "status": "not-cleared",
   "file": null,
   "thumb": null,
   "kind": "setup",
   "caption": "Althea II system overview: smart motors, PC, EM tracking, camera and shaft holders",
   "source_key": "norouzighazbi2021_design",
   "source": "Norouzi‐Ghazbi et al., Artificial Organs (2021), Fig. 1, p. E173",
   "source_url": "https://doi.org/10.1111/aor.13870",
   "where": "Fig. 1, p. E173",
   "licence": "publisher copyright (Wiley / ICAOT)"
  }
 ],
 "sources": [
  {
   "key": "ganji2009_robot",
   "doi": "10.1007/s11548-009-0296-z",
   "url": null,
   "title": "Robot-assisted catheter manipulation for intracardiac navigation",
   "year": 2009,
   "venue": "International Journal of Computer Assisted Radiology and Surgery"
  },
  {
   "key": "norouzighazbi2021_design",
   "doi": "10.1111/aor.13870",
   "url": null,
   "title": "Design and experimental evaluation of an automated catheter operating system",
   "year": 2021,
   "venue": "Artificial Organs"
  }
 ],
 "read_level": "full",
 "read_level_detail": "full (IJCARS 2009; Althea II, norouzighazbi2021_design, )",
 "notes": "Earlier, distinct robot of the A3 Toronto lineage. For A3's 'Althea II year N.R.': Althea II is Norouzi-Ghazbi et al. 2021 (Artif. Organs): reduced weight, assembly 20 min (5x faster), tip detection accuracy 0.05 mm, target accuracy above 90% in velocity and positioning modes (abstract). | 2026-11-08 X3: Althea II paper read in full; it names the 2009 CathNav robot Althea I (ref. 14), so Althea I and II are versions of this row and toronto_nazari2024 is the third generation (its 'Althea II, year N.R.' = 2021). Main picture now Althea II (2021); old 272 px 2009 thumbnail kept as ryerson_cathnav_ganji2009_prev.png",
 "maintainers_own": false,